Presentation is loading. Please wait.

Presentation is loading. Please wait.

Ch 08 Measurement of Flow and volume of Blood

Similar presentations


Presentation on theme: "Ch 08 Measurement of Flow and volume of Blood"— Presentation transcript:

1 Ch 08 Measurement of Flow and volume of Blood

2 Blood flow; blood volume change
ECG 退求其次 Blood pressure 退求其次 Blood flow; blood volume change 退求其次 O2 or nutrients concentration in the cells

3 8.1 Indicator-dilution method that uses continuous infusion
Cardiac output : blood flow from the heart (liter/min) Ci * F dm/dt = Co * F liters/liters * liters/min + liters/min = liters/liters * liters/min O2/Blood * Blood/min + O2/min = O2/Blood * Blood/min F m Ci Co Indicator F = (dm/dt)/(Co – Ci)

4 Fick Technique (cont.) Ca Cv dm/dt heart Rotational displacement
sensor Other signal processing Strip-chart recorder Counterweight Kymograph Bell Water seal Blood flow One-way valves Mouthpiece Soda-lime canister dm/dt Ca O2 Spirometer system Pulmonary system heart Cv

5 Figure 8.1 Several methods of measuring cardiac output In the Fick method, the indicator is O2; consumption is measured by a spirometer. The arterial-venous concentration difference is measure by drawing simples through catheters placed in an artery and in the pulmonary artery. In the dye-dilution method, dye is injected into the pulmonary artery and samples are taken from an artery. In the thermodilution method, cold saline is injected into the right atrium and temperature is measured in the pulmonary artery.

6 Thermodilution ( continuous infusion)
Heat injection T (continuous) t T1 heart T2 Methods for heat injetion: Heated or cooled saline Electric heater

7 8.2 Indicator-dilution method that uses rapid injection
Convenience: rapid injection > continuous infusion

8 Figure 8.2 Rapid-injection indicator-dilution curve After the bolus is injected at time A, there is a transportation delay before the concentration begins rising at time B. After the peak is passed, the curve enters an exponential decay region between C and D, which would continue decaying alone the dotted curve to t1 if there were no recirculation. However, recirculation causes a second peak at E before the indicator becomes thoroughly mixed in the blood at F. The dashed curve indicates the rapid recirculation that occurs when there is a hole between the left and right sides of the heart.

9 Dye dilution heart Dye injection Colorimeter

10 Thermodilution ( rapid injection)
heart T2 Baloon Thermistor Heat injection

11 Thermodilution ( rapid injection)
Advantages: The catheter can be left in place for 24 h No need to puncture an artery Sources of errors: Inadequate mixing - Heat exchange between the blood and the walls of the heart chambers - Heat exchange through the catheter walls

12 8.3 Electromagnetic flowmeters
Applicability: conductive liquid, such as blood and saline Measures: Instantaneous pulsatile blood flow * * Note: Indicator-dilution methods measure only average flow.

13 Faraday’s law of induction
弗萊明右手定律(又稱發電機定律): 將右手之大姆指.食指及中指伸直並相互垂直,若以大拇指表示導體運動之方向, 食指表示磁場由N至S的方向,則中指之指之指向就是感應電勢的方向。 Figure 8.3 Electromagnetic flowmeter When blood flows in the vessel with velocity u and passes through the magnetic field B, the induced emf e is measured at the electrodes shown. When an ac magnetic field is used, any flux lines cutting the shaded loop induce an undesired transformer voltage.

14 Figure 8.4 Solid lines show the weighting function that represents relative velocity contributions (indicated by numbers) to the total induced voltage for electrodes at the top and bottom of the circular cross section. If the vessel wall extends from the outside circle to the dashed line, the range of the weighting function is reduced. (Adapted from J. A. Shercliff, The Theory of Electromagnetic Flow Measurement, © 1962, Cambridge University Press.)

15 DC flowmeters were not satisfactory because:
Electrode-tissue interface impedance ECG interference Large 1/f noise in the 0-30 Hz range Figure 8.3 Electromagnetic flowmeter When blood flows in the vessel with velocity u and passes through the magnetic field B, the induced emf e is measured at the electrodes shown. When an ac magnetic field is used, any flux lines cutting the shaded loop induce an undesired transformer voltage.

16 Figure 8.5 Electromagnetic flowmeter waveforms The transformer voltage is 90º out of phase with the magnet current. Other waveforms are shown solid for forward flow and dashed for reverse flow. The gated signal from the gated-sine-wave flowmeter includes less area than the in-phase signal from the quadrature-suppression flowmeter.

17 Figure 8.6 The quadrature-suppression flowmeter detects the amplifier quadrature voltage. The quadrature generator feeds back a voltage to balance out the probe-generated transformer voltage.

18 Toroidal-type cuff probe
Figure 8.7 The toroidal-type cuff probe has two oppositely wound windings on each half of the core. The magnetic flux thus leaves the top of both sides, flows down in the center of the cuff, enters the base of the toroid, and flows up through both sides.

19 Figure 3.16 Functional operation of a phase-sensitive demodulator (a) Switching function. (b) Switch. (c), (e), (g), (i) Several input voltages. (d), (f), (h), (j) Corresponding output voltages.

20 Figure 3.17 A ring demodulator This phase-sensitive detector produces a full-wave-rectified output uo that is positive when the input voltage ui is in phase with the carrier voltage uc and negative when ui is 180º out of phase with uc .

21 8.4 Ultrasonic Flowmeters
* Can measure instantaneous flow * Transcutaneous * Can measure flow profile

22 Transducer: electric to acoustic
Transducers Transducer: electric to acoustic Near field: dnf = D2/(4) Far field: sin = 1.2 /D Selection: higher freq. and larger transducer Absorption coefficient (by tissue)  f (The lower the better) Back-scattered power (by blood cells)  f4 (The higher the better) Compromise: MHz Figure 8.8 Near and far fields for various transducer diameters and frequencies. Beams are drawn to scale, passing through a 10-mm-diameter vessel. Transducer diameters are 5, 2, and 1 mm. Solid lines are for 1.5 MHz, dashed lines for 7.5 MHz.

23 Transit-time flowmeter
Figure 8.9 Ultrasonic transducer configurations (a) A transit-time probe requires two transducers facing each other alone a path of length D inclined from the vessel axis at an angle . The hatched region represents a single acoustic pulse traveling between the two transducers. (b) In a transcutaneous probe, both transducers are placed on the same side of the vessel, so the probe can be placed on the skin. Beam intersection is shown hatched. (c) Any transducer may contain a plastic lens that focuses and narrows the beam. (d) For pulsed operation, the transducer is loaded by backing it with a mixture of tungsten powder in epoxy. This increases losses and lowers Q. Shaded region is shown for a single time of range gating. (e) A shaped piece of Lucite on the front loads the transducer and also refracts the beam. (f) A transducers placed on the end of a catheter beams ultrasound down the vessel. (g) For pulsed operation, the transducer is placed at an angle.

24 Continuous-wave Doppler flowmeter
Doppler effect Fixed source Moving source

25 Continuous-wave Doppler flowmeter (cont.)
As an ambulance approaches, the siren's sound waves become more frequent (compressed together). As the ambulance drives away, the siren's sound waves become less frequent (stretched apart). (Source:

26 Continuous-wave Doppler flowmeter (cont.)
For example, let c = 1500 m/s and u = 1.5 m/s. If a 1-kHz sound is emitted from the source, the frequency shift of the sound at the moving target will be The frequency of sound perceived at the moving target will be

27 Continuous-wave Doppler flowmeter (cont.)
Figure 8.10 Doppler ultrasonic blood flowmeter. In the simplest instrument, ultrasound is beamed through the vessel walls, back-scattered by the red blood cells, and received by a piezoelectric crystal.

28 Continuous-wave Doppler flowmeter (cont.)

29 Continuous-wave Doppler flowmeter (cont.)

30 Directional Doppler Figure 8.11 Directional Doppler block diagram (a) Quadrature-phase detector. Sine and cosine signals at the carrier frequency are summed with the RF output before detection. The output C from the cosine channel then leads (or lags) the output S from the sine channel if the flow is away from (or toward) the transducer. (b) Logic circuits route one-shot pulses through the top (or bottom) AND gate when the flow is away from (or toward) the transducer. The differential amplifier provides bi-directional output pulses that are then filtered.

31 Figure 8. 12 Directional Doppler signal waveforms (a) Vector diagram
Figure 8.12 Directional Doppler signal waveforms (a) Vector diagram. The sine wave at the carrier frequency lags the cosine wave by 90º. If flow is away from the transducer, the Doppler frequency is lower than the carrier. The short vector represents the Doppler signal and rotates clockwise, as shown by the numbers 1, 2, 3, and 4. (b) Timing diagram. The top two waves represent the single-peak envelope of the carrier plus the Doppler before detection Comparator outputs respond to the cosine channel audio signal after detection. One-shot pulses are derived from the sine channel and are gated through the correct AND gate by comparator outputs. The dashed lines indicate flow toward the transducer.

32 T obtain a velocity profile across the vessel.
Pulsed Doppler T obtain a velocity profile across the vessel.

33 Laser beams light through fiber opticals into the skin
Laser Doppler A laser power source; Laser beams light through fiber opticals into the skin

34 8.5 Thermal-convection velocity sensors
To measure local velocity

35 Figure 8.13 Thermal velocity probes (a) Velocity-sensitive thermistor Ru is exposed to the velocity stream. Temperature-compensating thermistor Rt is placed within the probe. (b) Thermistors placed down- and upstream from Ru are heated or not heated by Ru, thus indicating velocity direction. (c) Thermistors exposed to and shielded from flow can also indicate velocity direction.

36

37 Constant-temperature sensor
ub R1 R2 - Linearizer + uo Ru Rt Figure 8.14 Thermal velocity meter circuit. A velocity increase cools Ru, the velocity-measuring thermistor. This increases voltage to the noninverting op-amp input, which increases bridge voltage ub and heats Ru. Rt provides temperature compensation.

38

39 8.6 Chamber plethysmography
(a device for measuring and recording changes in the volume of the body or of a body part or organ) chamber 腔、室

40 Venous-occlusion plethysmography
To measure venous flow: pressurize the venous-occlusion cuff to 50 mmHg. To measure arterial flow: pressurize the arterial-occlusion cuff to 180 mmHg. Figure 8.15 In chamber plethysmography, the venous-occlusion cuff is inflated to 50 mm Hg (6.7 kPa), stopping venous return. Arterial flow causes an increase in volume of the leg segment, which the chamber measures. The text explains the purpose of the arterial-occlusion cuff.

41 thrombosis 血栓形成 Figure 8.16 After venous-occlusion cuff pressure is turned on, the initial volume-versus-time slope is caused by arterial inflow. After the cuff is released, segment volume rapidly returns to normal (A). If a venous thrombosis blocks the vein, return to normal is slower (B).

42 Venous-occlusion plethysmography
To measure venous flow: pressurize the venous-occlusion cuff to 50 mmHg. To measure arterial flow: pressurize the arterial-occlusion cuff to 180 mmHg. Figure 8.15 In chamber plethysmography, the venous-occlusion cuff is inflated to 50 mm Hg (6.7 kPa), stopping venous return. Arterial flow causes an increase in volume of the leg segment, which the chamber measures. The text explains the purpose of the arterial-occlusion cuff.

43 8.7 Electric-impedance plethysmography

44 Figure 8. 17 (a) A model for impedance plethysmography
Figure 8.17 (a) A model for impedance plethysmography. A cylindrical limb has length L and cross-sectional area A. With each pressure pulse, A increases by the shaded area D A. (b) This causes impedance of the blood, Zb, to be added in parallel to Z. (c) Usually DZ is measured instead of Zb.

45 Figure In two-electrode impedance plethysmography, switches are in the position shown, resulting in a high current density (solid lines) under voltage-sensing electrodes. In four-electrode impedance plethysmography, switches are thrown to the other position, resulting in a more uniform current density (dashed lines) under voltage-sensing electrodes.

46 Z1 Z2 =Z + DZ i DZ Zi =DZ KDZ Amp Zv Amp uo + - Z ub= Z Z3 Comparator
Demod. Amp uo + - Z ub= Z Z3 Comparator KDZ > +10 V or KDZ < -10 V? Sample and hold Z4 0.1 s one shot Sample Figure 8.19 In four-electrode impedance plethysmography, current is injected through two outer electrodes, and voltage is sensed between two inner electrodes. Amplification and demodulation yield Z + DZ. Normally, a balancing voltage ub is applied to produce the desired DZ. In the automatic-reset system, when saturation of uo occurs, the comparator commands the sample and hold to sample Z + DZ and hold it as ub. This resets the input to the final amplifier and uo zero. Further changes in DZ cause change in ub without saturation.

47 1. Legs – whether pulsations of volume are normal
Applications 1. Legs – whether pulsations of volume are normal 2. Legs --- detecting venous 3. Thorax – rate of ventilation 4. Neck and waist -- baet-by-beat changes in cardiac output 5. Limbs – flow of blood 6. Ventricle – changes in ventricular volume – cardiac output 7. Limbs – body water and body fat 8. EIT – pneumonia, stomach empltying, ventilation

48 8.8 Photoplethysmography

49 Figure 8.20 (a) Light transmitted into the finger pad is reflected off bone and detected by a photosensor. (b) Light transmitted through the aural pinna is detected by a photosensor.

50 Figure 8.21 In this photoplethysmograph, the output of a light-emitting diode is altered by tissue absorption to modulate the phototransistor. The dc level is blocked by the capacitor, and switch S restores the trace. A noninverting amplifier can drive low impedance loads, and it provides a gain of 100.


Download ppt "Ch 08 Measurement of Flow and volume of Blood"

Similar presentations


Ads by Google