1 Chapter 8 RF/Microwave Measurements
2
Introduction At low frequencies parameters such as voltage, current, etc. can be measured. from these impedance, power factor and phase angle can be calculated. At microwave frequencies It is more convenient to measure power instead of V and I. Properties of devices and circuits at microwave frequencies are characterized by S-parameters, power, frequency and VSWR and noise figure. 2
3
Power Measurement Power is defined as the quantity of energy dissipated or stored per unit time. Microwave power is divided into three categories: low power (less than 10mW), medium power (from 10mW to 10W) and high power (greater than 10W). Average power concept is used in microwaves P Avg = P Peak X Duty cycle 3
4
Power Measurement The general measurement technique for average power is to attach a properly calibrated sensor to the transmission line port at which the unknown power is to be measured. The output from the sensor is connected to an appropriate power meter. The RF power to the sensor is then turned off and the power meter zeroed. This operation is often referred to as “zero setting” or “zeroing.” Power is then turned on. The sensor, reacting to the new input level, sends a signal to the power meter and the new meter reading is observed. 4
5
Power Measurement Sensors for the measurement of microwave power can be divided into two categories: Devices whose resistance changes with applied power such as Schottky diode detectors, bolometer, thermocouple, etc. (used for low power measurements). Devices whose temperature changes with the applied power like calorimeter (used for medium to high power measurement). 5
6
Power Measurement Schottky Barrier Diode Detectors These are used as square law detector whose output is proportional to the input power. These are able to detect and measure power as low as −70 dBm (100 pW) at frequencies up to 18 GHz. The RF input signal is applied to R1, it passes through R2. The diode detects the input power and converts into heat energy. The corresponding temperature rise provides a change in electrical parameters which outputs current in low frequency circuitry. 6
7
Calorimeter Method Calorimetric method is used for high power microwave measurements which involves conversion of microwave energy into heat. The heat is absorbed by a fluid (usually water) and then temperature of fluid is measured to calculate power. 7
8
Calorimeter Method There are two methods to measure the heat of the fluid: Direct heating method: The rate of production of heat is measured by observing the rise in temperature of dissipating medium. Indirect heating method: In this method heat is transferred to another medium before measurement. In both the methods static calorimeter and circular calorimeter are used. 8
9
Static Calorimeter 9
10
Circular Calorimeter 10
11
Calorimeter Wattmeter/Powermeter The unknown RF power is checked against a 1200-cps (Hz/cycles per second) comparison power in the bridge circuit. Two temperature-sensitive resistors serve as gauges. In operation, the unknown RF heats an input load resistor. This resistor and one gauge are in close thermal proximity so that heat generated in the input load heats the gauge and unbalances the bridge. The unbalanced signal is amplified and applied to the comparison load resistor which is in close proximity to the second gauge, and rebalances the bridge. 11
12
Calorimeter Wattmeter/Powermeter The meter measures the power supplied to the comparison load to rebalance the bridge. Efficient heat transfer from the loads to the temperature gauges is accomplished by immersing the components in an oil stream. 12
13
Power Measurement Bolometer Bridge Method Bolometers are power sensors that operate by changing resistance due to a change in temperature. The change in temperature results from converting RF or microwave energy into heat within the bolometric element. There are two principle types of bolometers, barretters and thermistors. 13
14
Bolometer A barretter is a thin wire (like a fuse made of platinum or tungsten) that has a positive temperature coefficient of resistance. Thermistors are semiconductors with a negative temperature coefficient. Kobid Karkee, KEC Dhapakhel14 Barretter Thermistor
15
Power Measurement Bolometers are usually operated in standard Wheatstone bridge circuit. A bolometer mounting is placed on one of the arms of the bridge. The microwave power incident on the bolometer changes its resistance which imbalances the bridge. The change in the galvanometer current measures the incident power. Proportionate calibration of galvanometer can be done to read the power. 15
16
Single Bridge Bolometer Initially the bridge is at its balanced condition under zero incident power. The microwave power applied to bolometer arm will change its resistance causing an unbalance. The non-zero power is recorded in voltmeter which is calibrated to read the level of input microwave power. Suppose under balanced condition, the dc bias voltage of bolometer is E 1 and E 2 is the dc bias voltage of bolometer after microwave input is applied. 16
17
Single Bridge Bolometer The change in dc bias voltage (E 1 – E 2 ) is directly proportional to the microwave power. Disadvantage of using single bridge: The change of resistance due to mismatch at the microwave input part results in incorrect reading. The thermistor is sensitive to changes in ambient temperature resulting in false reading. These disadvantages can be overcome by using microwave double bridge. 17
18
Double Bridge Bolometer The upper bridge measures the microwave power. The lower bridge compensates the effects of ambient temperature variation(V 1 =V 2 ). The added microwave power due to mismatch is compensated by the negative dc feedback. The initial zero setting of the bridge is done by adjusting E 1 = E 2 = E 0 with no input signal applied. In absence of input signal E 1 /2 is the dc biasing voltage across the sensor at balance. In presence of input signal E 2 /2 is the dc biasing voltage across the sensor at balance. 18
19
Double Bridge Bolometer The average input Pav is equal to the change in dc power: For any change in temperature if the voltage change by ΔE, the change in RF power is given by: Since V1+V2 >> ΔV, ΔP=0, so the second equation can be used directly to calculate the average power. 19
20
Thermocouple Sensors A thermocouple is a junction of two dissimilar metals or semiconductors. The semiconductor used in thermocouple is n-type Si. A thin film of titanium-nitride resistive load is deposited on a Si substrate which forms one electrode of thermocouple. The thermocouple generates an emf when two ends are heated up differently by absorption of microwaves in resistive loads. The emf is proportional to the incident microwave power to be measured. 20
21
Thermocouple Sensors As shown in figure, C 2 is the RF bypass capacitor and C 1 is the input coupling capacitor or dc block. The emf generated in the parallel thermocouples are added to appear across C 2. The output leads going to the dc voltmeter are at RF ground so that the output meter reads pure dc voltage proportional to the input microwave power. For square wave modulated microwave signal peak power can be calculated from average power as P peak = (P avg X T)/τ where T is time period τ is pulse width 21
22
Slotted Line Carriages A slotted line carriage is a microwave instrument which is used to measure: Wavelength Voltage Standing Wave Ratio (VSWR) and standing wave pattern Impedance, reflection coefficient and return loss measurement It has a coaxial E-field probe which penetrates inside a rectangular waveguides slotted in sections from the outer wall. The probe is able to transverse a longitudinal narrow slot and locate the standing waves maxima(V max ) and minima(V min ) along the line giving VSWR. 22
23
VSWR Meter VSWR meter is a highly sensitive, high gain, low noise voltage amplifier tuned normally at fixed frequency of 1KHZ square wave of which microwave signals modulated. The modulated signal is then amplified and detected which then measured with a calibrated voltmeter. This meter indicates calibrated VSWR reading for any loads. 23
24
Spectrum Analyser Spectrum analyser is a microwave instrument which provides signal spectrums, i.e. the plot of amplitude against the frequencies. The simplified block diagram is shown below: 24
25
Spectrum Analyser The microwave signal to be measured is superheterodyned with sweep voltage produced by a sweep generator and oscillates with local oscillator. The mixed signal is then amplified by narrow bandwidth intermediate frequency amplifier. The signal is then detected and video amplified for display in terms of amplitude and frequency. The sweep voltage is sawtooth type signal. The zero flyback time of sweep voltage moves the spot on display horizontally in synchronization with frequency sweep. This makes the horizontal position function of frequency and amplitude of signal the vertical deflection of the signal. 25
26
Vector Network Analyser (VNA) VNA measures both amplitude and phase over a wide range of frequencies. When an RF signal is applied to a network, such as a filter, amplifier, or transmission line, that signal is altered in magnitude and phase. If the magnitude and phase of the altered signal can be compared to the magnitude and phase of the originating RF signal, the characteristics of that network can be evaluated. 26
27
Vector Network Analyser (VNA) The network, or component, being tested is called the Device Under Test (DUT). The RF signal that is input to the DUT is the Reference signal. The DUT will alter the Reference signal's two components, Magnitude and Phase. The DUT will change the magnitude component, due to it's resistive natures. It will alter the phase component due to it's reactive natures. These two altered components of the Reference signal are measured by the magnitude and phase comparators within the VNA with reference signal. 27
28
Vector Network Analyser (VNA) The output of the Magnitude Comparator is some value that represents the difference between the voltage, or power, of it's two input signals. This value of differential magnitude is called the Magnitude Vector. The output of the Phase Comparator is some value that represents the difference between the phase of it's two input signals. This value of differential phase is called the Phase Vector. 28