Download presentation
1
Nuclear Medicine Physics
Single Photon Emission Computed Tomography (SPECT) Jerry Allison, Ph.D. Department of Radiology Medical College of Georgia
2
Medical College of Georgia And Sameer Tipnis, Ph.D.
A note of thanks to Z. J. Cao, Ph.D. Medical College of Georgia And Sameer Tipnis, Ph.D. G. Donald Frey, Ph.D. Medical University of South Carolina for Sharing nuclear medicine presentation content
3
Tomographic NM imaging (SPECT)
Single Photon Emission Computed Tomography Nuclear Medicine Physics for Radiology Residents Sameer Tipnis, PhD, DABR
4
Scintillation Camera © Physics in Nuclear Medicine: Cherry, Sorenson and Phelps, 4th edition, 2012
5
Scintillation Camera PMT Event Positioning Network
© Physics in Nuclear Medicine: Cherry, Sorenson and Phelps, 4th edition, 2012
6
Tomographic imaging (SPECT)
Produce tomographic images by acquiring conventional gamma camera projection data at several angles around the patient Similar to CT
7
SPECT Provide 3-D images to eliminate overlaying and underlying activity of a slice better contrast more accurate lesion localization more demanding technically and longer data acquisition more severe image noise
8
SPECT data acquisition
Generally, two detectors mounted at 180 or 90 on a rotation gantry
9
SPECT data acquisition
a sequence of 2-D static images at different angular positions (views) detector rotation range 180º with 2 perpendicular detectors or 360º with 2 opposite detectors 45º RAO In principle, rotation of the gamma camera about 180° allows for the acquisition of sufficient projections for tomographic reconstruction. However, in practice, opposing views acquired 180° apart differ due to various factors (photon attenuation, depth dependent collimator response) and SPECT data are commonly acquired over 360°. 45º LPO
10
SPECT data acquisition
Circular or elliptical orbit, which is better? closer to the patient better spatial resolution
11
SPECT Image Acquisition
(Improves spatial resolution) Nuclear Medicine Physics for Radiology Residents Sameer Tipnis, PhD, DABR
12
Data Collection: Configuration
Non Cardiac Cardiac Delete??????
13
SPECT Image Acquisition
Typically 2 camera heads, rotating around patient Projection images every 3 – 6 degrees ~ 30 s / projection, ~ 15 minutes total Matrix - 64 x 64 or 128 x 128 Nuclear Medicine Physics for Radiology Residents Sameer Tipnis, PhD, DABR
14
Data Collection: Angular Stops
3 to 6 degrees is common a lesser number causes streaking a larger number does not improve image quality Step-shoot or continuous acquisition, which is better?????? Step & Shoot Characteristics Some loss of time Less Blur
15
View number for 360º SPECT number of views = matrix size 128 views
16
Each projection has 128 data points
An image with 128 x 128 matrix Contains 128 projections Each projection has 128 data points Equivalent to 128 slice CT (i.e., 128 tomographic slices per rotation) Nuclear Medicine Physics for Radiology Residents Sameer Tipnis, PhD, DABR
17
Sinogram (for one of many slices)
There are many (e.g. 64 or 128) sinograms per rotation. © Physics in Nuclear Medicine: Cherry, Sorenson and Phelps
18
© Physics in Nuclear Medicine: Cherry, Sorenson and Phelps
19
Back Projection Leads to blurring in image (streaks and star-like artifacts) Deleted Kuhl’s Brain Imaging
20
Filtered Backprojection
Suppress blurring through filtering the projections A high-pass filter (ramp filter) can be used to suppress blurring A problem that is immediately apparent is the blurring (star-like artifacts) that occur. One would expect that a high-pass filter could be used to eliminate blurring, and that is the case. The optimal way to eliminate these patterns in the noiseless case is through a ramp filter. The combination of back projection and ramp filtering is known as filtered back projection.
21
Filtered Back Projection (of noiseless data)
© Physics in Nuclear Medicine: Cherry, Sorenson and Phelps © Physics in Nuclear Medicine: Cherry, Sorenson and Phelps
22
Filter Amplitude Frequency
Ramp filter Amplitude Ramp w/ rolloff filter Frequency Define K(L) & L In the absence of noise, a ramp filter works well For noisy images, a ramp filter with roll-off is required
23
Filter Ramp Ramp with some roll-off filter
Suppresses blurring but enhances noise Ramp with some roll-off filter Smoothes the image and suppresses noise Trade off noise vs resolution Roll-off filter characteristics are adjustable
24
Filter Types © Physics in Nuclear Medicine: Cherry, Sorenson and Phelps
25
Filter Types (A) Butterworth: (less noise, more smooth)
(B) Butterworth: (more noise, less smooth) Nuclear Medicine Physics for Radiology Residents Sameer Tipnis, PhD, DABR
26
Filter Applied filter is the product of: Ramp
User selected/characterized filter Shepp-Logan Hahn Butterworth Weiner Hamming Hanning (MCG: Philips, generally turned off)
27
Filter Applied filter is the product of: Procedure for applying filter
Ramp User selected/characterized filter Shepp-Logan (cut-off freq) Hann (cut-off freq) Butterworth (order (slope), critical freq (0.707 response) Procedure for applying filter 1D projection of each view is converted to spatial frequency using a Fourier transform Ramp filter with roll-off is applied in spatial frequency space (k-space) Filtered projection is recovered with inverse Fourier transform Back projections performed to reconstruct image
28
Selection of Filters for SPECT
Filters trade noise for resolution No standard way to optimize filter choice Patient to patient variation Physician preferences Vendor recommendation Embellish from Cherry.
29
Iterative Reconstruction (IR)
Filtered Back Projection has some limits Various corrections needed Attenuation Compton Scatter Ordered Subsets Expectation Maximization (OSEM) is a common iterative reconstruction algorithm Improve comments on limitations of filtered back projection.
30
Calculation Includes: Attenuation Scatter Blur with depth
Assume Some Image (I) Calculate Projections (P’) Calculation Includes: Attenuation Scatter Blur with depth Compare to Measured Projection (P) Use P’ & P to form corrections Image estimate may be a uniform image Assumed image undergoes forward projection to produce sinogram Form New Image (I’) Is I-I’< * Done
31
Iterative Reconstruction
Slow compared to filtered back projection Commonly used for PET Being used increasingly in SPECT IR used for ~all Philips SPECT at MCG Siemens C-Cam?????? Has become the standard for SPECT?????
32
Image recon - Iterative
Common IR recon is the OSEM For OSEM, # iterations (I) and # subsets (S) affect image quality # (I/S) noise, but sharper images Nuclear Medicine Physics for Radiology Residents Sameer Tipnis, PhD, DABR
33
Iteration 1 Nuclear Medicine Physics for Radiology Residents Sameer Tipnis, PhD, DABR
34
Iteration 5 Nuclear Medicine Physics for Radiology Residents Sameer Tipnis, PhD, DABR
35
Iteration 10 Nuclear Medicine Physics for Radiology Residents Sameer Tipnis, PhD, DABR
36
Brain Phantom IR OSEM FBP
Nuclear Medicine Physics for Radiology Residents Sameer Tipnis, PhD, DABR
37
Non-filter Noise Factors
Collimator Matrix 64 x 64 128 x128 Slice thickness Time per stop/ Number of Stops Administered Dose Deleted: Number of matrix element rows
38
Data Collection Whole Image is collected for each view
64 x 64 or 128 x 128 Each row makes a slice Multiple slices can be added to reduce noise Anything higher than 128 x 128?
39
Data Collection: Counts
Determination of Number of Image Counts Activity in patient Time per stop Number of Stops
40
Attenuation Correction
Like all radionuclide imaging there is a problem due to attenuation. Correction can be important for judging the activity of lesions
41
Attenuation correction
s traveling smaller paths through pt (nearer to camera) have less attenuation compared to those from deeper in pt for AC can be assumed or measured Chang (assumed), Measured - Gd rods (older) or CT (new) CT can be non-diagnostic (low power, cone-beam) or fully diagnostic depending on the scanner model Nuclear Medicine Physics for Radiology Residents Sameer Tipnis, PhD, DABR
42
Attenuation in SPECT I0 = I1e+a (attenuation corrected intensity)
Probability of detection / correct intensity I0, dependent on the depth at which originates need to know “a” Nuclear Medicine Physics for Radiology Residents Sameer Tipnis, PhD, DABR
43
Chang’s AC method Image first reconstructed without AC
Contours of image used to estimate t for each projection, assumed to be constant Average ACF determined for each pixel (x,y) from all projections Reconstructed image corrected pixel-by-pixel Works well for area with approximately constant attenuation like head, abdomen but not for areas like chest / thorax Nuclear Medicine Physics for Radiology Residents Sameer Tipnis, PhD, DABR
44
I(x) = I0e-mx (x is thickness of tissue between pixel & detector)
Chang’s Method Assume uniform attenuation m linear attenuation coefficient m is ~0.15/cm for Tc-99m for soft tissue I(x) = I0e-mx (x is thickness of tissue between pixel & detector) 0.15/cm is the narrowbeam attenuation coefficient for 140 keV photons in soft tissue Use of the broadbeam coefficient, 0.12/cm, helps to account for attenuation and scatter in uniform radioactivity distributions In Chang’s method, m is often set to ~0.12/cm to better account for Compton scatter
45
Uniform phantom with evenly distributed 99mTc
Chang method Proper AC Low counts in center Nuclear Medicine Physics for Radiology Residents Sameer Tipnis, PhD, DABR
46
SPECT/CT
47
AC in SPECT/CT Accurate / realistic -map obtained for each projection using CT values used in Chang’s algorithm to correct pixel-by-pixel AC here is more realistic (since is not assumed to be constant) Current SPECT/CT systems use this method Nuclear Medicine Physics for Radiology Residents Sameer Tipnis, PhD, DABR
48
Philips Astonish NM Recon Software
The Astonish: Ordered Subsets Expectation Maximization (OSEM) Compensation for the blurring effects of the collimator built into the reconstruction Resolution Recovery allows recovery of some of the original resolution Astonish uses the distance from the detector to the object of interest recorded as a function of angle by the camera during acquisition and geometric properties of the specific collimator Has become the standard for SPECT?????
49
Philips Astonish NM Recon Software
Used on essentially all Philips NM SPECT images at MCG Iterations: 4 Subsets: 16 CT data used for attenuation correction (except brains due to EEG electrode artifacts) Chang’s AC invoked for brains Has become the standard for SPECT?????
50
Common SPECT Problems Patient motion
System Alignment (Center of rotation issues) Collimator issues Distance issues Loss of resolution with distance
51
Center of Rotation SPECT assumes heads always look at a constant central rotation point
52
COR (Spatial Alignment)
Image will have blurring and circular artifacts COR must be tested periodically for all heads SPECT of three point sources Generally done with system QC software SPECT of 3 point sources ?????
54
COR correction SPECT image of a uniform phantom on a camera with poor COR correction Incorrect COR correction introducing a ring artifact, degrading spatial resolution Nuclear Medicine Physics for Radiology Residents Sameer Tipnis, PhD, DABR
55
COR correction COR correction is used when reconstructing tomo-graphic data to correct for minor misalignment between the center of the image and the axis of rotation. COR corrections are stored in a correction table and are applied automatically after a data set has been acquired. Nuclear Medicine Physics for Radiology Residents Sameer Tipnis, PhD, DABR
56
Collimator Issues Collimators are not completely uniform
A high count flood must be stored to correct for collimator non-uniformities 20 M for 5% for 128 x128 Check or delete “20 M for 5% for 128 x128”
57
Patient Studies Advantages Disadvantages No overlapping structures
3 dimensional lesion locations Fusion with high resolution images (CT, MRI) Disadvantages Time consuming (motion) Images are noisy
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.