Dialysis System SW Training module 7 – Report and Service Environment

Similar presentations


Presentation on theme: "Dialysis System SW Training module 7 – Report and Service Environment"— Presentation transcript:

1 Dialysis System SW Training module 7 – Report and Service Environment Ref: Operator’s Manual SW - Chapter 15 – Report Environment - Service Man chapters 1.2, 1.3 and 10

2 Report and Service Environment
Report and Service Environment Report and Service Environment Ref: Operator’s Manual SW 8.08 Service code Chapter 15 – Report Environment - Service Man chapters 1.2, 1.3 and 10 Important Notice: This presentation does not relieve any user of the Artis Dialysis System from his or her duty to carefully read the full text of the Operator’s Manual before using the system. Important: Please note that the SW change to of this presentation does not change the training of the Artis Dialysis System for the users.

3 Report Screen Usability
Report Screen Usability Event List Automatic treatment records at regular intervals (default 30 min) Manual snapshot New Reading button Present Time button (available in all 5 Main screens) Report Screen Usability Operator’s Manual, section In the Report screen Event list, treatment data are automatically recorded and displayed along the treatment (Start treatment to Disconnect Patient) The time interval for automatic recording is settable minutes (default 30 minutes) Also specific operations during treatment are automatically recorded and displayed in the Event list (UF off, UF on, Pause treatment, Stop Treatment, etc.) Treatment data is recorded every time the New Reading or Present Time button is pressed When doing so, a Snapshot button appears in the Event list Pressing the Snapshot button opens the Modify Snapshot Event screen (see next slide) Data collected during a treatment is available until a new prescription is selected and confirmed on the network OR the Connect Patient button is pressed and confirmed All treatment-related parameters recorded (automatically or manually) and displayed in the Event list are also available for network acquisition through the Exalis System. Highlight Press the Snapshot button to access the Modify Snapshot Event screen See next slide Press the New Reading button to access Modify Snapshot Event screen

4 Snapshot and Event reporting
Snapshot and Event reporting Snapshot and Event Reporting Operator’s Manual, section and Pressing the Event list Snapshot button OR the Report screen Event button, opens the Modify Snapshot Event screen In the Modify Snapshot Event screen, the Operator can enter incidents and actions occurred and medications administered during the treatment be selecting and Confirming the appropriate choice from the three select pads available When Confirming a choice, the Event list is updated accordingly This screen also offers the possibility to change the time interval for automatic recordings minutes (default 30 minutes) Highlight Press the Auto-chart time button to change the records time interval Needs to be emphasized as this button is kind of “hidden” (not on the most logical screen) Press the Auto-chart time button to change the records time interval

5 Hygiene sub-screen Disinfection history table always available
Hygiene sub-screen Ultrafilters history table is always available Ultrafilters change procedure button is available in the standard environment Default values for Ultrafilters history counters correspond to the maximum authorized duration Hygiene Sub-screen Ref. Operator’s Manual, Hygiene , Ultra Filter Change 8.25 and Disinfection History table Disinfection history table always available

6 Button Bar Usability LCD test
Button Bar Usability LCD test The test should be performed if the Operator observes or suspects that data are not correctly displayed on the Touch Screen Follow the instructions appearing on the screen (click = press) Call for service if not functioning correctly Screen cleaning using alcohol solution only Button Bar Usability Operator’s Manual, section and section

7 Button Bar usability – System data
Button Bar usability – System data Button Bar Usability – System Data Operator’s Manual, section Service related data can be reviewed during the treatment without entering the service environment The service data are displayed in four different pages according to the module the data belongs to

8 Artis Service mode – Two levels
Artis Service mode – Two levels Two Service Levels with different access codes Service Level 1: Access code Date and time change Auto start preset Displays all preset parameters Service Level 2: Technicians access Preset environment Transfer of presets via an USB key Calibration of sensors BlackBox downloading Artis Service Mode – Two Levels Service Manual, section 1.3 The Service button is available in the Report Screen button bar during preparation until Auto-prime is booked or when no process is running Pressing the Service button a Key pad appears Two different codes are required depending on Service Level: Service Level 1 accessibility is not restricted but an access code is required to enter the service screens (Gambro Lund local phone number) Service Level 2 is restricted by access code – is specifically intended for the service technicians

9 Artis Service mode – NavPad
Artis Service mode – NavPad + Service code Preset Screen Service Overview Screen Exit Service Mode Blood Screen Blood circuit schematics and sensors/actuators status Fluid Screen Fluid circuit schematics and sensors/actuators status Artis Service Mode – NavPad Service Manual, section 1.3 The Service NavPad follows the Treatment NavPad concept in general and give access to the Service Mode screens and menus The Service Mode Preset Screen button, give access to the Service Level 1 screens and menus via its respective button situated in the button bar Technical Log Screen

10 Artis Service mode – Preset Screen
Artis Service mode – Preset Screen All presets can be viewed by scrolling the different select pads Service level 1 Possible to set/change: Time date Auto Start Service level 2 Possible to set/change: Installed features Default settings Treatment mode Preset transfer (next slide) Artiset Service Mode – Preset Screen Preset Manual Specifically notice The screenshot shows the Service Level 2 Preset screen (USB Configuration Transfer instead of the Ultrafiletr Change button in the button bar) It is highly recommended that the list of presets and default functions (systematic activation of features such as Diascan, Hemoscan, Hemocontrol…) is defined through a collaboration with the clinical and commercial managers Highlight The Exit Service Mode button must always be pressed to save the changes and exit Service mode The Exit Service button = Service Overview button The Exit Service Mode button must always be pressed to save the changes and exit Service mode

11 Artis Service mode – Preset transfer
Artis Service mode – Preset transfer Import Export Load from … Save to … By the use of a USB key presets can be: Downloaded from one machine (Import) Uploaded to another (Export) Artis Service Mode – Preset Transfer Service Manual, section 1.3 and Preset Manual

12 Artis Service mode - Calibration
Artis Service mode - Calibration The Calibration button available in the button bar gives access to the Calibration Component Map from which individual calibrations and tests can be performed Artis Service Mode – Calibration Service Manual, section 1.3

13 Artis Service mode - Blackbox
Artis Service mode - Blackbox In the Blackbox, Artis stores the latest 1 to max 24 sessions According to the principle “first in- first out” Example of information stored: Treatment data Alarms Operator’s action Machine processes and phases Actuators information status Limiting factors: Blackbox 24 Mb memory size Sampling time, 10 sec > 0.7Mb/session Artis Service Mode - Blackbox Operator’s Manual, section and Service Man chapter 1.2, 1.3 and 10 Blackbox downloading of session data is highly valuable facilitating the investigation of problems that may have occurred during treatment Data is logged every 10 second according to the Blackbox recorder configuration - from Function check start to Disinfection end When the maximum BlackBox memory size is reached, the system overwrites data starting from the oldest ones The log list is set by the service technician

14 Blackbox files The Blackbox files can be downloaded to an USB key via the Report screen, requiring Service Level 2 access code During preparation until Auto-Prime is booked or when no process is running The downloaded Blackbox files should be named accordingly: BBoxnn_ddmmyy_hhmm.dat BBoxnn = nn progressive ID of session (recycled from1 to 24) ddmmyy = date, month and year hhmm = time of session start (hours and minutes) Blackbox files can only be read with Mediartis Tech sw application Blackbox Files Service Manual, section 1.3

15 Blackbox downloading procedure
Blackbox downloading procedure Export All Export Files Target + Service code Export Blackbox File Close Current Session Place a USB stick in the Connectivity Panel USB port Press Blackbox button and enter Service Level 2 code in the key pad appearing Press Export Blackbox File button in the screen appearing → only selected file downloaded to USB key Press Export All button  all files in left window are downloaded to the USB key Close Blackbox screens and remove USB key from the port Press Service Overview button and Exit Service Mode button Blackbox Downloading Procedure Service Manual, section 1.3

16 Hands on Module 7 Event list New reading/Snapshot Time interval
Hands on Module 7 Event list New reading/Snapshot Time interval Hygiene screen LDC test Screen Cleaning Access Service Level 1 and 2 Preset process Blackbox downloading Hands on Module 7 The Trainer’s Check List should be used and ticked off during the hands-on training By following the different hands-on prescriptions all function will have been practiced at the end of the course

17 Gambro is an indirect subsidiary of Baxter International Inc.
For the safe and proper use of the Artis Dialysis System, refer to Warnings, Precautions and complete Instructions for Use in the Operator’s Manual available by contacting customer support. Gambro is an indirect subsidiary of Baxter International Inc.