AMANDA L. STEBER, MARIYAM FATIMA, CRISTÓBAL PÉREZ, and MELANIE SCHNELL
AN GHz SEGMENTED CHIRPED PULSE FOURIER TRANSFORM MICROWAVE SPECTROMETER FOR ASTROCHEMICAL APPLICATIONS AMANDA L. STEBER, MARIYAM FATIMA, CRISTÓBAL PÉREZ, and MELANIE SCHNELL Styling: MPSD color:
2
Introduction Chirped pulse Fourier transform microwave spectroscopy
Increased data throughput Used in both a large and small bandwidth configuration Segmented Chirped pulse Introduced in 2013 Used at mmw frequencies Utilizes lower cost components Introduces real time averaging Goals: Decrease the cost Decrease acquisition times G.G. Brown et al., J. Mol. Spectro, 238, 200 (2006) J.L. Neill et al., Opt. Express, 21, (2013)
3
Instrumentation N.A. Seifert et al., J. Mol. Spectro., 312, 13 (2015)
4
Instrumentation
5
Segmented, Fast Frame Chirped Pulse
Divide 8 GHz bandwidth into 10 – 800 MHz segments Pulse train Depending upon the pumping speed you can fit multiple segments into one gas pulse
6
Phase Stability We are able to collect the pulses (that have been heavily attenuated by the switch)
7
Phase stability
8
OCS / Sensitivity 0.2 % OCS in Neon 3 bar backing pressure
200,000 averages 18O13CS SNR ~3:1 % Not intensity calibrated
9
1-Hexanal 0.2% 1-Hexanal in Neon 1 bar backing pressure 1 nozzle
10
1-Hexanal N.A. Seifert et al., J. Mol. Spectro., 312, 13 (2015)
11
Spurious Signals Generated from the local oscillator Occur ~200 MHz
2M average spectrum
12
1-Hexanal Spurious SIgnals
N.A. Seifert et al., J. Mol. Spectro., 312, 13 (2015)
13
Comparison between two approaches
Broadband Segmented Effective Rep Rate 30 Hz 15 Hz Power/GHz 5 W/GHz 5-7 W/GHz Bandwidth 8 GHz /segment 800 MHz/segment Nozzles 3 1 Pulse Duration 1 ms 2 ms FID duration 10 ms LNA 48 dBm 45 dBm ~ Cost ~350,000 € ~150,000 € SNR of conformer 1 of 1-Hexanal 3500:1 ~1300:1
14
Astrochemistry Applications
Broadband reaction screening Based on the work done at Uva Discharge nozzle Paired with the W-band instrument Astrochemically relevant molecules Alaninol Isoleucinol M.C. McCarthy et al., Astrophys. J. Suppl. Ser., 129, 129 (2000) D.P. Zaleski et al., Astrophys. J. Lett., 765, L10 (2013) R. A. Loomis et al., Astrophys. J. Lett., 765, L9 (2013)
15
Conclusions and Future Directions
A new segmented, fast frame chirped pulse spectrometer has been built Potential to reduce time Reduces cost Future Directions Intensity calibration needs to be improved Spurious signals better cateloged Pumping speed improved
16
Acknowledgement ERC Louise Johnson Fellowship
17
Power Curve Power loss across electronics Frequency Dependent
18
1-Hexanal N.A. Seifert et al., J. Mol. Spectro., 312, 13 (2015)