Presentation is loading. Please wait.

Presentation is loading. Please wait.

Momentum and Momentum Spread Measurements

Similar presentations


Presentation on theme: "Momentum and Momentum Spread Measurements"— Presentation transcript:

1 Momentum and Momentum Spread Measurements

2 Momentum Measurement For Dipole 1 New addition to procedure
 (in MeV/c) = I (in Amps) as implemented in the magnet table New addition to procedure measure charge on BPM-02 and BPM-03 to make sure all beam is measured Is BPM-03 a better measure of 45 than YAG-04? Need shot-by-shot measurement of beam centroid on YAG-04 to use!

3 Momentum v. Phase New data shows no change from ‘run 3’ and ASTRA simulations (parameters not identical)

4 Momentum v. Power New data shows same behaviour as previous experiments Fit parameter indicates effective shunt impedance of 23.3 M/m Missing momentum is still missing!

5 Momentum Spread Measurement
Beam size given by If x is minimised using QUADS 1 - 4 Can calculate momentum spread from dispersion D and the beam size at YAG-04

6 Dispersion Calculate QUADS 5-6 must be off – may be a problem if beam image is large Does not account for any residual fields Calculated = pixels Measured = pixels Measure Take data for three or more magnet settings/image positions on the screen Measure on-the-fly Measure dispersion for each data point

7 Measurement on-the-fly
Bottom 5 points taken at low charge – not representative! Even then there is considerable statistical variation on dispersion for nominally identical measurements Using an average value for dispersion produces smoother parameter response (all data re-analysed this way) Measuring dispersion up-front and using a single value may lead to a small systematic error but removes large random error allowing trends to be seen more easily

8 Data extraction Summation Doesn’t assume particular line shape
Quick – no fitting! Susceptible to problems with background Gaussian Fitting Seems to get width of signal even for non-Gaussian profiles Still has issues with background at low charge By-eye (pencil and paper) Allows intelligent background subtraction Cannot be used for large data sets

9 Momentum Spread v. Solenoid
Previous data showed higher momentum spread at optimum solenoid settings for transverse properties Not confirmed by first re-measurement Single shot data! Subsequent data Low charge data not good Some hint in higher charge data More detailed scans required ASTRA simulations would also be good

10 Momentum Spread v. Phase
Previous data looked convincing Minimum momentum spread not identical to simulation but parameters differ Single shot data, phase for minimum momentum spread depends on analysis technique! x not minimised x not minimised Data collected so far does not allow the phase for minimum momentum spread to be accurately assessed!

11 Momentum Spread v. Phase (TDC)
Two data sets, one with the beam set up for TDC work (y minimised?) and one with both x and y ‘minimised’ and TDC actually on. x minimisation not good enough on either set to be useful Getting slice momentum spread measurements could be extremely challenging!

12 Momentum Spread v. Charge
Previous work Low charge data limited by clipping New data Low charge data influenced by background subtraction issues Clipping in the magnet vessel

13 Charge Dependence Along the Beamline
BA2 – Measure on YAG-09, careful setting of QUAD-12 required BA1 – Measure on BA1-YAG-02, beam transport could be an issue, no QUADs after dipole so no control of dispersion

14 Charge Dependence Along the Beamline
Some interpolation and limited extrapolation Is behaviour as expected? Charge transport to BA1 an issue? ICT in BA1 BPMs for charge measurement AP shifts for characterising beam in BA1?

15 Summary Momentum measurements Momentum spread measurements
New data confirms previous results – momentum still missing! Momentum spread measurements Lots of data but much of it useless - need to improve procedures!  minimisation Eliminating background issues (turning lights off, subtracting dark images, optimising data extraction) Shot-to shot YAG image analysis, automated data collection routines? Important beam parameters not accurately measured Phase for minimum energy spread Evolution of charge dependence along the beamline Simulations with matching parameters to data would help!


Download ppt "Momentum and Momentum Spread Measurements"

Similar presentations


Ads by Google