Download presentation
Presentation is loading. Please wait.
Published byJocelyn Alexander Modified over 6 years ago
1
Longitudinal Compression Second Generation DES: What You Need to Know
Ian T. Meredith MBBS, PhD, FRACP, FCSANZ, FACC, FSCAI, FAPSIC, FAHA Professor and Director of Monash Heart Monash Medical Centre, Melbourne Australia
2
Ian T Meredith, MD, PhD Honoraria: Abbott Vascular
Boston Scientific Corporation Medtronic, Inc.
3
What is Longitudinal Stent Deformation?
LONGITUDINAL / AXIAL The axial shortening or lengthening of a stent after implantation, resulting from interaction with an ancillary device such as a guide catheter, post-dilatation balloon, IVUS catheter or possibly another stent
4
Longitudinal Stent Deformation Should Not Be Confused With..
Foreshortening Inherent change in stent’s length occurring during implantation and expansion. Recoil Change in stent’s diameter due to mechanical and radial forces. Stent Fracture Cracking or breaking of a stent strut.
5
Is Longitudinal Stent Deformation a New Problem?
No The axial shortening or lengthening of a stent after implantation, resulting from interaction with another device balloon or stent reported 18 yrs ago Chalet et. al., Cathet Cardiovasc Interv 1994; 32: 376-9 Bhargava et. al., J Invasive Cardiol 1996; 8: 447-9 Experienced clinically and anecdotally with: Wiktor Stent, GR 1 and GR 2, Cordis Tantalum Stent and others
6
Can Procedural Technique Contribute to Longitudinal Stent Deformation?
Yes shortening or lengthening probably occurs when an ancillary device bumps or catches a mal-apposed strut during re-cross or subsequent withdrawal. Likely procedural Factors When nominal inflation pressure and diameter are not attained initially leading to under-deployment and mal-apposition of stent struts When the lesion tapers, leading to malapposition of struts When the stent is undersized Where there is wire bias or significant curvature resulting in poor coaxial alignment of the second device Deep seated guide catheter. .
7
Can Stent Design Contribute to Longitudinal Stent Deformation?
Yes potentially Different stent designs do have different “longitudinal strength” and integrity in bench testing * Ormiston But stent performance is the sum or a number of interrelated and interdependent design features * Ormiston et. al., J Am Coll Cardiol Intv 2011; 4:
8
Recent Attention to Stent Deformation
PROMUS® Element™, BioMatrix™, and Resolute Integrity™ Stents EuroIntervention 2011 Endeavor®/Driver® Stent EuroIntervention 2011 The issue of stent deformation has received much attention lately. Let us put this issue into perspective, particularly as it relates to the Platinum Chromium Stent Platform. First, it is important to recognize that the frequency of stent deformation is a very low.* That said, there are lessons that have developed with the recent focus on this issue. *sources: Boston Scientific Element platform clinical programs and complaints data on file; PLATNUM Trial: Stone G et.al., JACC 2011, 57; PERSEUS WH Trial: Kereiakes D et.al. JACC 2010, 56.
9
Are There Clinical Consequences of Longitudinal Distortion?
Yes may be, potentially Luminal obstruction resulting in ST or enhancing risk or restenosis Inadequate lesion coverage Mal-apposed struts may predispose to stent thrombosis & restenosis Ormiston TCT 2011
10
Are There Clinical Consequences of Longitudinal Distortion?
Yes may be, potentially Luminal obstruction resulting in ST or enhancing risk or restenosis Inadequate lesion coverage Mal-apposed struts may predispose to stent thrombosis & restenosis But This assumes ìt is unrecognized and untreated Post-dilatation to correct luminal obstruction and appose struts or additional stenting of uncovered lesion may suffice Importantly the evidence thus far is anecdotal Ormiston TCT 2011
11
Stent Design in Perspective (Delivery & Acute performance)
are a series of “sinusoidal hoops” and “connectors” Hoops may be “in phase” or “out of phase” influence radial strength and scaffolding and to a lesser extent flexibility and conformability Connectors the number, orientation, shape, thickness and composition influence flexibility, deliverability and conformability and tissue coverage and influence longitudinal strength Ormiston TCT 2011
12
Stent Design in Perspective (Outcome)
effects the vascular responses stent deployment Strut orientation and cell design influence magnitude of stent induced platelet activation and inflammatory neo-intimal response Strut thickness and alloy composition also influence extent of stent-induced vessel injury and inflammation Strut thickness RCT and registry evidence of a relationship between strut thickness and ISR Ormiston TCT 2011
13
Stent Design Progression: Platform
Material SSTL CoCr / CoNi PtCr Architecture Strength Flexibility Coverage Optimization Closed Cell Open Cell Hybrid Strut Thickness
14
Design of the Stents Tested
Figure 1. Design of the Stents Tested Shown are photographs of the stent designs, their names, the metal they are constructed from, strut thickness in microns, and the number of connectors between hoops for each design. The Cypher Select, cut from a stainless steel tube, has a strut thickness of 140 m. Its design is out-of-phase sinusoidal hoops linked by 6 sinusoidal bridges that are orientated about 30° from the stent long axis. The Liberte, cut from a stainless steel tube, has struts100 m thick. Its apparently complex design is fundamentally out-of-phase hoops (yellow line) that are joined directly by 3 links (red arrows). The Vision and MultiLink 8 and their drug-eluting counterparts, Xience V and Xience Prime, are cut from a cobalt chromium tube and have struts 81 m thick. The design is in-phase sinusoidal hoops linked by 3 bridges that are aligned with the stent long axis. Each connector has a U-shaped loop to improve flexibility. The 3-mm Driver (and the drugeluting version, Endeavor or Resolute) has sinusoidal, largely out-of-phase hoops linked by 2 welds (red arrows). The Integrity and its Resolute drug-eluting counterpart has a single sinusoidal cobalt chromium component that winds helically from 1 end of the stent to the other with 2 welds between adjacent “hoops” (red arrows). The Omega (bare-metal version) and Promus Element (everolimus-eluting version) and the paclitaxel-eluting version (Taxus Element, called “ION” in the United States) have sinusoidal hoops made from platinum chromium. These are linked by 2 straight bridges per hoop that are aligned at an angle of about 45° from the stent long axis (red arrows). Ormiston et al J. Am. Coll. Cardiol. Intv. 2011;4;
15
Comparative Stent Longitudinal Shortening and Distortion With 0
Comparative Stent Longitudinal Shortening and Distortion With 0.5 N Compressing Force Figure 4. Comparative Stent Longitudinal Shortening and Distortion With 0.5 N Compressing Force The 0.5 N force did not compress the Cypher Select stent, and the Liberte was compressed 2 mm. The Vision/Xience and MultiLink 8/Xience Prime were compressed 1 mm with minimal distortion. The Integrity shortened by 2 mm, with a small amount of strut overlap. The most shortening was with the Driver/Endeavor (4 mm) and Omega/Element (5 mm), and these stents experienced the most longitudinal distortion, with strut overlap. Ormiston et al J. Am. Coll. Cardiol. Intv. 2011;4;
16
Compressive Force and Stent Shortening
Figure 3. Compressive Force and Stent Shortening Plotted is stent shortening (millimeters) against force (Newtons) for 3 examples of the 7 stent designs. Compression graphs are not smooth but are complicated because of strut collision and over-ride. Ormiston et al J. Am. Coll. Cardiol. Intv. 2011;4;
17
Flexibility and Axial Strength A Balancing Act
SSTL PtCr Closed Cell Open Cell Axial Strength Flexibility
18
Conformability and Axial Strength Stent flexibility is inversely related to axial strength
Bending Moment (Newtons/mm) Here you can see the association between flexibility and axial stiffness. The Cypher stent is the stiffest stent and the Platinum Chromium platform provides the most flexibility. More Flexible Cypher® Stent n=6 Xience V ® (PROMUS®) Stent n=10 Integrity® Stent n=3 Driver® Stent n=7 PtCr Element Stents n=15 Bench test results may not necessarily be indicative of clinical performance. Data on file at BSC. PROMUS Stent is a private-labeled Xience V Everolimus Eluting Coronary Stent System manufactured by Abbott and distributed by Boston Scientific Corporation.
19
Measured Stent Length in PERSEUS
Cumulative frequency distribution of the ratio of QCA stent length to nominal stent length – single lesion PERSEUS RCT workhorse pts with 1 implanted study stent (N=1152) Quantiles TAXUS Express (N=289) ION/TAXUS Element (N=863) 100% Max 1.280 1.225 99% 1.173 1.059 95% 1.018 1.009 90% 0.998 75% Q3 0.989 0.988 50% Median 0.965 0.960 25% Q1 0.919 0.915 10% 0.850 0.860 5% 0.830 0.819 1% 0.730 0.745 0% Min 0.612 0.594 100 80 ION™/TAXUS Element™ TAXUS® Express® 60 Percentile 40 20 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 Ratio No cases of longitudinal stent compression or deformation were seen in either group by the core laboratory Presented by Gregg W. Stone, MD, TCT 2011.
20
Measured Stent Length in PLATINUM
Cumulative frequency distribution of the ratio of QCA stent length to nominal stent length – single lesion Platinum RCT pts with 1 implanted study stent and no non-target lesions in the target vessel (N=1251) Quantiles PROMUS (N=612) PROMUS Element (N=639) 100% Max 1.146 1.310 99% 1.081 1.077 95% 1.006 1.020 90% 0.998 1.000 75% Q3 0.986 0.984 50% Median 0.943 0.950 25% Q1 0.877 0.897 10% 0.819 0.847 5% 0.748 0.802 1% 0.643 0.664 0% Min 0.426 0.559 100 80 60 Percentile 40 Promus Element™ Promus 20 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 Ratio No cases of longitudinal stent compression or deformation were seen in either group by the core laboratory. Popma JP. 20
21
PROMUS™ (Xience V™) Stent (n=762)
PLATINUM Workhorse (12 months) TLF rates with PROMUS Element™ and PROMUS™ Stents P superiority =0.72 P noninferiority =0.0009 P=0.96 P=0.14 P=0. 51 P>0.99 Incidence Rate (%) PROMUS™ (Xience V™) Stent (n=762) PROMUS Element™ Stent (n=768) * Intent-to-Treat Population *TLF=Ischemia-driven (ID) target lesion revascularization (TLR), or cardiac death/myocardial infarction (MI) related to the target vessel (TV). GW Stone, et. al., JACC 57;16, PROMUS Stent is a private-labeled Xience V Everolimus Eluting Coronary Stent System manufactured by Abbott and distributed by Boston Scientific Corporation. Xience V is a registered trademark of Abbott Cardiovascular Systems, Inc.
22
PROMUS™ (Xience V™) Stent (n=762)
PLATINUM Workhorse Less bail-out stenting with PtCr stents due to improved lesion coverage Overall Rate Cited Reason for Bail-out Stent P=0.004 P=0.36 P=0.01 P=0.06 Incidence Rate (%) PROMUS™ (Xience V™) Stent (n=762) PROMUS Element™ Stent (n=768) Presented by Gregg W. Stone, MD, ACC PROMUS Stent is a private-labeled Xience V Everolimus Eluting Coronary Stent System manufactured by Abbott and distributed by Boston Scientific Corporation. Xience V is a registered trademark of Abbott Cardiovascular Systems, Inc.
23
Stenting Technique to Reduce Incidence
Take care when crossing a newly deployed stent Visualize stent when crossing any ancillary device Ensure proper apposition of the stent to the arterial wall Avoid deep seating the guide catheter Minimize wire bias where possible Longer inflation and deflation times with long stents Illustration for information purposes, not indicative of actual size or clinical outcome.
24
Longitudinal Stent Distortion
The potential ALC clearly exists with modern stents Not all stents are equal wrt to this potential Caution when focusing on one design feature in isolation Stent performance and outcome are the sum of many features All designs have trade-offs, strengths and weaknesses Importantly clinical trials have not demonstrated an effect Finesse rather than force is important when re-crossing stents. Elegant tools require delicate hands Illustration for information purposes, not indicative of actual size or clinical outcome.
25
Thank You
26
SCAAR Registry Crude Restenosis Rates - Unadjusted
0% 2% 4% 6% 6 12 Stents Implanted November 2009 – March 2011 Cypher™ Stent n=782 Endeavor™ Stent n=747 TAXUS™ Liberté™ Stent n=1,393 PROMUS™ (Xience V™) Stent n=1,533 Endeavor™ Resolute™ Stent n=1,566 Xience Prime™ Stent n=4,832 Cumulative Rate of Restenosis PROMUS Element™ Stent n=2,724 Total number of Stents n=13,577 Time in Months Presented by Stefan James, MD, ACC PROMUS Stent is a private-labeled Xience V Everolimus Eluting Coronary Stent System manufactured by Abbott and distributed by Boston Scientific Corporation.
27
Stent Appearances After 5-mm Shortening
Figure 5. Stent Appearances After 5-mm Shortening Shown are micro-computed tomographic images cut electronically longitudinally after the stents had been compressed by 5 mm. Of course, different magnitudes of force were required to compress different designs by 5 mm. The lower portions of the stents below the broken line were clamped, hence were not subjected to compressive forces, and serve as normal stent comparitors for insights into mechanisms of shortening. For the Cypher Select the hoops were able to bunch as the wavelength of the sinusoidal connectors shortened and the connectors angulated and became less aligned with the long axis. For the Vision and MultiLink 8 designs, the hoops were pushed together where there are no connectors. In addition, the connectors became angulated and less aligned with the long axis of the stent. Furthermore, especially with the MultiLink 8, the U-shaped loop in the connector closed up. The Liberte, Driver, and Integrity designs have only 2 connectors between hoops, and these are direct connections so there was little to prevent struts bunching and overlapping in response to compressive forces. With only 2 connectors between hoops, there is again little to prevent Omega/Element struts being pushed together. Additionally, connector orientation changed, becoming more aligned with the stent long axis. Ormiston et al J. Am. Coll. Cardiol. Intv. 2011;4;
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.