Download presentation
Presentation is loading. Please wait.
1
Introducing Ansur ver. 2.9.1 and INDEX2 Plug-in ver. 1.0.3
Subscribe to twibes.com/ansur for instant notification of new versions! ©2010 Fluke Biomedical.
2
Presentation overview
Introducing Index 2 Plug-In Custom R-curve utility How these participate in productivity improvement and quality assurance through reduction in human error Ordering information FAQs Summary How to get answers to questions
3
Ansur is now compatible with Index 2!
Performance Analyzers Ventilator Tester QAVTM Defibrillators Impulse 7000 Impulse 4000 QED6 QA-40/45 Electro-surgery QA-ES Infusion Device LAGU Electrical Safety Analyzers ESA620 ESA612 ESA601 QA-90 QA-ST Non-Invasive Blood Pressure Simulators BP Pump 2 QA-1290 X-Ray Test Device TNT 12000 Pulse Oximeter Simulator Index 2
4
Increase productivity with Ansur Index 2
January 29-February 1, 2003 Reduce total test time on pulse oximeters Automate FBC test instruments and extract data electronically Create SpO2 sensor specific R-curves, which can then be quickly and easily stored in six download slots in the Index 2. Automatically assesses Pass/Fail against specified test limits Generate & archive professional test reports Minimize training dependence, and allow users to leave bulky service manuals behind Guided step-wise testing process to ensure standardization Create standard electronic templates that mirror recommended test procedures Estimated savings of $1700 per year in testing and documentation
5
Estimated savings of $1700 per year
©2010 Fluke Biomedical. Company confidential: not for public use.
6
Manage risk and ensure compliance
Ensure entire teams test and document same way every time Reduce human error through electronic data collection and transfer from test devices to minimize human error Trace data easily with electronic archival of detailed preventative maintenance and repair records for medical devices ©2010 Fluke Biomedical. Company confidential: not for public use.
7
What’s new in Ansur ver. 2.9.1 PDF creator now built in!
New Test report layout selections icon Easier to add customer’s logo to Test Report Print Header
8
Introducing Index 2 Plug-in ver. 1.0.3
medTester-equivalent templates in Ansur Library Optical Electrical (using electrical simulation cables) Custom R-curve utility Restores ability to create custom R-curves missing since Index 2M-series Auto-configures Index 2 for test to be performed Automatically evaluates keyboard entered test data against pre-determined test limits for objective pass/fail
9
Custom R-Curve Utility
Designed to follow custom R-curve creation process from Index 2 Operator’s Manual, chapter 12; replaces Index R-curve data sheet Includes settings for TLC (adjusts for lower absorption due to finger thickness), RDC, IRDC, IRAC, light level (medium or low) Save in computer for later use (set up your own R-curve library) or click Send to install into appropriate Index 2 R-curve position (5 locations available: 3 medium light level, 2 low light level)
10
Installing a Custom R-Curve into Index 2
With Index 2 connected to computer: Open Ansur (using Rune) Select Tools > Options Scroll down to Index 2 thumbnail and select Index 2 Use Oximeter Make drop-down menu to select R-curve to be installed Click to select “Send custom R-curve to Index 2XL before performing the test” or view R-curve
11
Test Automation: Productivity and QA
Use with any other ECG simulation-capable test tool (e.g., ESA612, ESA620, Impulse 6000/7000, etc.), or any patient simulator (MPS450, PS420, etc.) in an overall patient monitor test procedure or multi-parameter monitor test procedure Use with BP Pump 2 for static and/or dynamic NIBP in an enhanced patient monitor or multi-parameter monitor test procedure Overall test procedure covers all required inspection steps and tests (regulatory compliance) Reduces sources of human error (quality assurance improvement) Reduces combined testing and documentation time (productivity improvement)
12
Ordering information Item Number Model Description 2817203
ANSUR INDEX2 ANSUR TEST SOFTWARE, INDEX2 PLUG-IN LICENSE
13
FAQs Question: Why can’t Ansur/Index 2 automatically report data?
Answer: Index 2 is a simulator. There is no connection between Ansur (computer) and the oximeter. Ansur auto-configures Index 2 input to the oximeter. Keyboard entry of displayed information is required, which Ansur evaluates against pre-determined test limits for pass/fail result Question: Will using the Ansur Index 2 Plug-in allow it to work with the newest Nellcor oximeters? How about Masimo Rainbow? Answer: No. Ansur cannot make any test instrument do more than its design allows. There remain hardware incompatibilities with the newest Nellcor technology and Masimo Rainbow uses multiple light wavelengths (multiple LEDs) which Index 2 cannot simulate. Question: How fast is the operation of the Ansur Index 2 Plug-in? Answer: The configuration of Index 2 by Ansur for any particular test is very fast. However, the signal processing time of oximeters varies widely; and the older the oximeter technology, the slower its processing and settling time. Newer oximeter technologies process quicker and settle quicker.
14
Summary Ansur ver makes paperless possible with or without CMMS interface with new .pdf test reports Ansur Index 2 plug-in ver reduces sources of human error via auto-configuration of Index 2, and auto-evaluation of keyboard-entered test data against pre-determined test limits Ansur Index 2 plug-in ver improves regulatory compliance when used in an overall patient monitor or multiple-parameter monitor inspection procedure (standard work) to document test results Ansur Index 2 plug-in license pricing is included in the latest price lists
15
Questions? For more information on Ansur Test Automation, contact Fluke Biomedical today. Fluke Biomedical Toll free: (800) option 0, ask for Tech Support Direct: (440) Fax: (440) If you find you have other questions about products from Fluke Biomedical, please contact us at the phone numbers or address shown. Thank you for using Fluke Biomedical products.
Similar presentations
© 2025 SlidePlayer.com. Inc.