1 UV/Visible Spectroscopy

2 UV/Vis Spectrometer Measure?
Spectroscopy What does a UV/Vis Spectrometer Measure?

3 UV/VIS Spectroscopy T = P/Po
Transmittance (T) is defined as the amount of light passing through the sample solution (P) divided by the amount of incident radiation (Po) T = P/Po In practice Po is estimated by a “blank” which accounts for reflection and scattering losses.

4 UV/VIS Spectroscopy As the radiation is absorbed in the sample,
the total intensity of radiation is reduced as it travels through the sample. This results in a non-linear relationship between transmittance and concentration.

5 UV/VIS Spectroscopy A = log10(1/T) Absorbance (A) is based
on the amount of light absorbed by the solution and is defined as the log of the inverse of the transmittance. A = log10(1/T)

6 UV/VIS Spectroscopy Within limits, the relationship between absorbance and concentration is linear.

7 A = bc UV/VIS Spectroscopy Beer’s law or the Beer-Lambert law
 is the molar absorptivity with units of (L/mol-cm) b is the path length of the sample ie. the inside cross section of the sample cuvette (cm) c is the concentration of the compound in solution in (mol/L)

8 UV/VIS Spectroscopy Increased interaction between the molecules
can affect absorbance at high concentrations. Electrolytes can also alter the molar absorptivity of the analyte by electostatic interactions. Molar absorptivity is also altered by the refractive index of the solution.

9 UV/VIS Spectroscopy Hindicator H+ + Indicator-
When an analyte associates or dissociates in solution producing products with different absorption spectra, deviations from Beer’s law can be observed if the equilibrium shifts at different concentrations. Hindicator H+ + Indicator-

10 UV/VIS Spectroscopy Other causes of deviation from Beer’s law:
Radiation is not monochromatic Stray radiation Absorbance readings less than 0.10 and higher than 1.5 can contain significant error and should be avoided.