Presentation is loading. Please wait.

Presentation is loading. Please wait.

Dialysis System SW Training module 10 – Single Needle Double Pump

Similar presentations


Presentation on theme: "Dialysis System SW Training module 10 – Single Needle Double Pump"— Presentation transcript:

1 Dialysis System SW Training module 10 – Single Needle Double Pump Ref: Operator’s Manual SW - Chapter 6: HD-SN Treatment

2 HD-SN Double Pump Important Notice: Important:
HD-SN Double Pump Single Needle Double Pump Ref: Operator’s Manual SW 8.08 Service code - Chapter 6: HD-SN Treatment Important Notice: This presentation does not relieve any user of the Artis Dialysis System from his or her duty to carefully read the full text of the Operator’s Manual before using the system. Important: Please note that the SW change to of this presentation does not change the training of the Artis Dialysis System for the user.

3 SNDP Principle SN cassette Arterial phase; Arterial blood pump (Art_BP) is running until the automatically calculated post dialyzer pressure limit is met (SN pressure max) - venous line is clamped (VC) Venous phase; Venous blood pump (Ven_BP) is running until the fixed low system pressure limit is met. (SN pressure min) - arterial line is clamped (AC) SNDP Principle SN maximum and minimum pressures are measured in the post dialyzer chamber The extracorporeal flow: Arterial clamp  pre pump chamber with arterial pressure measurement  arterial blood pump  pre dialyzer chamber  dialyzer  post dialyzer chamber with SN pressure measurement  venous pump  venous chamber with venous pressure measurement  Venous clamp

4 Programming HD-SN DP on Artis
Programming HD-SN DP on Artis 1. Select treatment type 2. Confirm parameters Single Needle Double Pump – SNDP Operator’s Manual, section and 3.4.2 SN must be preset in service level 2 to be available in the Treatment type list Treatment Type must be confirmed before loading the SNDP ArtiSet The Stroke Volume Is a mandatory parameter – must be Confirmed before treatment start Can be set within the range of ml - default value is 30 ml Changeable before and during the treatment (via the Prescription screen Parameter list or on the Blood screen) Highlight Always set the HD-SN Treatment before starting the HD-SN DP cassette loading procedure If not, ADS will expect a standard ArtiSet and guide the Operator to open only the Arterial pump cover … Always set the HD-SN Treatment before starting the HD-SN cassette loading procedure

5 Priming performed in SN
HD-SN DP HD SN HC Priming performed in SN Patient Connection SN or DN Special Procedures SN  DN DN  SN HD SN (Double Pump) Ref: Operator’s Manual: Section Machine dressing and priming and Special Procedures 8:20 and 8:21 Install ArtiSet HD SN HC Preparation and priming is performed in SN mode Patient Connection may be performed in SN or DN mode From Patient Connection until Stop Treatment it is possible to switch from DN to SN (or SN to DN) as many times as needed, keeping the SN cassette installed. New SKU’s and IFU’s because of the new Intended Use (DNDP) and performances in DNDP – regulatory purposes Needle mode switch is available, from Connect Patient until Stop Treatment keeping the SN cassette installed

6 Switching from SN to DN (DP) …
Switching from SN to DN (DP) … Switching from SN to DN (DP) … Ref: Operator’s Manual: Special Procedures 8:20 Note: It is not possible to switch from SN to DN when Isolated UF is still activated. Performing this switch means the treatment continues in DN DP mode. The Arterial and Venous pumps are then running with the same rate, which is controlled by the Operator using the Arterial Blood Pump hard keys. The Venous Blood Pump is further controlled by a fixed system pressure of 100 mmHg (displayed in the Blood screen).. Diascan and Hemoscan functions will remain available if they were activated from start. Control loop on Venous pump with system pressure fixed at 100mmHg when in Double needle Double pump mode

7 Needle mode switch is possible as many times as needed
… and from DN to SN (DP) ... and from DN to SN (DP) Ref: Operator’s Manual: Special Procedures 8:21 Note: It is not possible to switch from DN to SN when Isolated UF is still activated. To Confirm the switch to SN DP mode, the Stroke Volume first has to be set/Confirmed (mandatory parameter). The Mean Blood Flow and Stroke Volume are displayed in the treatment overview screen. The Blood Setting Sub-screen becomes available. Diascan and Hemoscan functions will remain available if they were activated from start. If activated, the Hemocontrol function deactivation Confirm Window appears; Check the UF and dialysis fluid Sodium parameters before Confirmation. (The UF volume/rate and dialysis fluid sodium parameters may be adjusted via the prescription screen select pad if needed.) Needle mode switch is possible as many times as needed

8 SNDP ArtiSet Dialyzer Art. line Post dialyzer Expansion Chamber Pre dialyzer Dialyzer Ven. line SN Service lines Integrated blood circuit including 2 expansion chambers and a venous pump segment Automatic loading of the complete blood circuit Automatic priming and chamber level adjustment SNDP ArtiSet Operator’s Manual, section and section 16.8 concerning alarm 594 description Demonstrate the SN ArtiSet and specifically point out the additional components Double cassette; standard blood cassette + 2 expansion chambers (pre-dialyzer and post-dialyzer) and a pump segment Autopriming - standard preparation All chamber fluid levels - blood cassette and expansion chambers - are automatically adjusted Blood circuit and pump management are designed to provide continuous and almost constant flow in the dialyzer - expansion chamber before and after the dialyzer is patented by Gambro Highlight: The fluid level in the post dialyzer chamber is normally quite low It really looks too low but once the treatment is running it will be OK SNDP alarm “594 SN Service Line(s) on SN cassette open” during priming. NOTE: If the clamps are closed, open the venous blood pump cover and check that the pump segment is correctly loaded. If not, it is possible to load it manually and then reset the alarm (once or twice) The fluid level in the post dialyzer chamber is normally quite low

9 Blood Levels in the Chambers
Blood Levels in the Chambers Blood level adjustment during SNDP treatment 1. Arterial chamber During the arterial phase when the arterial automatic clamp is open 2. Venous chamber During the venous phase when the venous automatic clamp is open 3. Expansion chambers (if blood levels in the two expansion chambers become greatly different; e.g. after an alarm situation): Decrease the “Stroke Volume” to 20 ml and the “Mean Blood Flow” to 50 ml/min and wait at least one cycle (arterial + venous phase) before stopping the blood pumps Attach a sterile luer-lock syringe (at least 20cc) to the Service Line on the expansion chamber with the higher blood level and gently adjust by pushing or pulling the blood level in accordance with the other expansion chamber Restart the blood pumps and adjust the Mean Blood Flow and “Stroke Volume” Blood Levels in the Chambers Operator’s Manual, section 8.28

10 SNDP Parameters Mean blood flow rate
SNDP Parameters Mean blood flow rate The effective blood flow rate during one cycle (20 to 270 mL/min, 10 mL/min step) Stroke volume The volume of blood passing through the dialyzer during a cycle (20 to 60 mL, 5 mL step) SN pressure max Automatically calculated in accordance with the set stroke volume SN pressure Actual pressure in the post dialyzer chamber SN pressure min Fixed at 40 mmHg SNDP Parameters Operator’s Manual, section 6.3.2

11 Treatment Overview Screen
Treatment Overview Screen Mean Blood Flow directly adjustable with BP hard keys Stroke Volume value box 200 Treatment Overview Screen Operator’s Manual, section 6.2 and 6.3 2 differences vs. DN treatment Overview screen Mean blood flow – set and actual Stroke volume – actual The Mean Blood Flow can be adjusted either by: Using the Blood Flow Increase/Decrease keys on the Hard Key Panel - In this case, the ratio Arterial Flow / Venous Flow is kept constant. Using the “Arterial Flow” and “Venous Flow” buttons on the Blood Settings sub-screen during treatment. Each time the set Mean Blood Flow is changed in this way, the ratio Arterial Flow / Venous Flow is changed (see slide 13). Highlight: Adjustable mean blood flow I.e. no need to set one flow for the arterial pump and one flow for the venous pump Adjustable mean blood flow

12 Blood Settings Screen Adjustable Stroke Volume
Blood Settings Screen Stroke volume set button (also available on Prescription screen) Stroke Volume value box Arterial and Venous pressure values and trend graph Blood Settings button 200 Blood Screen Operator’s Manual, section 6.3 The Blood screen contains additional information: The set and actual stroke volume An arterial and venous pressure graph that looks a bit different due to bigger changes in the arterial and venous pressures along with the blood pump shifts A blood Settings button Highlight: Adjustable Stroke Volume That is as most parameters reachable via the prescription screen or the blood screen Adjustable Stroke Volume

13 Stroke Volume + = 40ml Arterial Phase Venous Phase Stroke Volume
Stroke Volume + Stroke Volume The stroke volume or the volume of blood passing through the dialyzer during one cycle can be visualized according to this slide During the arterial phase, the levels are increasing in the two chamber pre and post dialyzer During the venous phase, the levels are decreasing  the amount now pumped back to the patient is the stroke volume Since one chamber is pre and the other is post, the flow through the dialyzer is more or less continuous = 40ml Arterial Phase Venous Phase

14 Blood Settings sub-screen
Blood Settings sub-screen Arterial flow set button Venous flow set button 200 The Mean Blood Flow is automatically recalculated when Arterial or Venous Flow values are changed Blood Setting sub-screen Operator’s Manual, section 6.3 The blood flow for the arterial and venous pumps can be set individually If another flow ratio than the preset 30% higher venous flow is needed, the arterial and venous flows can be adjusted using the two flow set buttons This subs-screen also contains some additional information Maximum SN pressure Actual SN pressure Minimum SN pressure A graph displaying the actual SN pressure Highlight: Usually the venous pump runs about 30% faster than the arterial pump Normally, it is possible to return the blood with a higher flow than it is removed from the patient i.e. decreasing the cycle time – this is the reason for the preset 30% higher venous flow (default) SN Pressure Max, dependant of the stroke volume SN Pressure, post dialyzer chamber pressure SN Pressure Min, fixed at 40mmHg Usually the venous pump runs about 30% faster than the arterial pump as it is easier to return blood faster - it also decreases cycle time

15 During Rinseback the A/V pumps run synchronously
SNDP Supervision Operator control: Mean blood flow and Stroke Volume Arterial and Venous flow adjustment (blood pump speed) for fine-tuning System control: Post dialyzer pressure management; auto adjustment of pump speed and pressure threshold Safety system: Pressure monitoring and Time-cycle SNDP Supervision Operator’s Manual, section System Control; SN Max pressure is automatically calculated by the machine according to the Stroke Volume set by the operator. When the “SN pressure” reaches this value, the phase is switched from Arterial to Venous. Post-dialyzer Pressure: -110 mmHg to +800 mmHg Safety system; There is a built in time limit that prevents the pumps from running for too long time Highlight: During Rinseback the A/V pumps run synchronously This saves time since the blood is being return with a 100% continuous flow – normally in SN/DP the arterial pump is only running part of the time During Rinseback the A/V pumps run synchronously

16 Nice to Know ! Functions available with SNDP – Diascan*, Hemoscan, Isolated UF (*=accuracy not guaranteed for SN: blood pump does not run as fast as in double needle mode resulting in a decreased volume of blood cleared/processed) The fluid level in the SN Venous Chamber is normally quite low The Blood setting button becomes available at blood detection When the Blood Setting sub-screen is open, the BP hard keys ▼ and ▲ are disabled When a SN parameter is changed, 3-4 cycles are required for the change to become valid Nice to know Last bullet – be patient, it takes a while for the machine to adjust

17 Nice to Know – Switching Needle Mode
Nice to Know – Switching Needle Mode ! Isolated UF, Hemoscan and Diascan* are available when the treatment is started in HD DN (SP/DP) or SN DP mode. (*=accuracy not guaranteed for SN) The Isolated UF function must be deactivated before switching Needle Mode (not available in the Select pad) and cannot be reactivated after the switch. HemoControl is available when the treatment is started in DN. When switching to SN (DP or SP), HemoControl is deactivated upon the Operator’s confirmation of UF rate and dialysis fluid sodium concentration. Nice to Know

18 Hands on Module 10 SNDP Switch between the needle modes
Hands on Module 10 SNDP Switch between the needle modes Hands on Module 10 The Trainer’s Check List should be used and ticked off during the hands-on training By following the different hands-on prescriptions all function will have been practiced at the end of the course

19 Abbreviations for products
HD SN DP: Hemodialysis Single Needle Double Pump HD DN HC: Hemodialysis Double Needle Hemocontrol HD DN DP: Hemodialysis Double Needle Double Pump

20 For the safe and proper use of the devices referenced within, refer to Warnings, Precautions and the complete Instructions for Use or the appropriate manual avaialble by contacting customer support ArtiSet, Diascan, Hemoscan and Hemocontrol are Trademarks of Gambro AB. Gambro is an indirect subsidiary of Baxter International Inc.


Download ppt "Dialysis System SW Training module 10 – Single Needle Double Pump"

Similar presentations


Ads by Google