Download presentation
Presentation is loading. Please wait.
1
Dynamic methods to construct phase diagrams
Lection 9 Dynamic methods to construct phase diagrams Dynamic methods of phase equilibrium studies – DTA, HF-DSC a. Unary system; b. Binary and ternary systems c. TGA; d. Heat capacity measurements using DSC
2
Thermal Analysis: Dynamic Methods
Measurement of physical property of a substance as a function of temperature using controlled temperature program Method Measured property Application Differential Thermal Analysis (DTA) Temperature difference Phase reactions, phase transformations Differential Scanning Calorimetry (DSC) Heat flow Specific heat, Heat of transition Thermal Gravimetric Analysis (TGA) Mass Change Reactions with the gas phase, decomposition reactions Dilatometry Size Change Phase transformations, Thermal expansion Thermomechanical Analysis (TMA) Mechanical properties Materials testing
3
DTA vs. DSC DTA DSC Heat-flux Power compensation
DT between sample and reference DT between sample and reference Power compensation to keep the same temperature in both furnaces T and DH of transformation More robust, measurements can be done in wider T range, in more aggressive environment (oxidation atmosphere), possible combination with TGA to measure mass change T and DH of transformation, Cp measurements T and DH of transformation, Cp measurements HF-DSC is more sensitive than DTA, possible to measure heat capacity More effective, since response time is shorter than in HF-DSC DTA/HF-DSC signal is difference between sample and reference thermocouples. It is usually given in mV. For some devices it is given as temperature difference; this means that reference table or equation was used for recalculation. The temperature is not measured in the sample, but at the bottom of crucible. Temperature calibration is necessary.
4
The DTA Signal Heat is transferred between furnace, crucible, sample (reference) and thermocouple Steady –state condition ∆Φ 𝑆𝑅 =− 𝐴𝜆 Δ𝑙 Δ 𝑇 𝑆𝑅 =−𝐾Δ𝑇 DFSR – difference in heat flow rate, l – thermal conductivity TW is furnace temperature, TC is temperature of crucible TT is temperature of thermocouple, TS is sample temperature Idealized curve Real curve
5
DTA signal for unary system
DTA responses to melting and freezing of pure substance: a- onset temperature, b- peak at temperature c. Due to dynamic character of experiment, the temperature distribution is never completely homogeneous. The temperature is not measured in the sample, but at the bottom of crucible. That is why temperature correction is necessary. Temperature calibration: establishment of relation between measured temperature Tmeas indicated by the instrument and the true temperature Ttr. At least three substances (usually pure metals) with melting temperature covering temperature range of interest should be selected. Mass should be corresponded to recommended mass for measurement in this instrument. Measurement should be done with different heating rate b and extrapolated to b=0.
6
Temperature calibration
Measured temperature of thermal event depends on mass and cooling rate b. Extrapolated T melting of Sn to b=0 for different sample mass Substance T, °C Sn Al Ag 961.78 Au Pd 1554.8 Temperature correction for Ga, In and Sn
7
Enthalpy calibration KH is instrument sensitivity for Ga, In, Sn
𝑚 𝑠𝑎𝑚𝑝𝑙𝑒 ∆ 𝐻 𝑠𝑎𝑚𝑝𝑙𝑒 = 𝐾 𝐻 𝑡1 𝑡2 ∆𝑇 𝑡 𝑑𝑡
8
Recommended values of temperatures and enthalpies of melting of metals
Enthalpy calibration is establishment of relationship between the enthalpy change DHmeas (peak area A) measured by the instrument and true enthalpy change DtrsH absorbed or released by the sample as a result of transition at transition temperature Ttrs: DtrsH=KH(Ttrs)DHmeas. Enthalpy calibration factors for each calibration substance are represented as a function of transition temperature. Provided the dependence on heating rate b and sample mass are negligible (within scatter of individual experiments) the enthalpy calibration factor KH(Ttr, m, b) give the enthalpy calibration function KH(T). Element Tmelt (°C) DHm(J/g) Ga 29.764 80.07 In 28.62 Sn 60.38 Zn 108.09 Al 399.87 Ag 961.78 104.61 Au 64.58
9
Heat capacity measurement using DSC: classical three-step method
Empty crucible measurement F0(T) Calibration substance measurement (e.g. synthetic sapphire, Pt) 𝐶 𝑃,𝑐 𝑚 𝑐 𝛽= 𝐾 Φ (𝑇)( Φ 𝑐 − Φ 0 ) KF(T) – calibration function Sample measurements 𝐶 𝑃,𝑆 𝑚 𝑆 𝛽= 𝐾 Φ (𝑇)( Φ 𝑆 − Φ 0 ) All three measurements have to be performed under exactly the same conditions Good thermal contact between sample and crucible is required 𝐶 𝑃,𝑆 = 𝑚 𝑐 𝑚 𝑠 ∆Φ 𝑆 ∆Φ 𝑐 𝐶 𝑃,𝑐 DFS=FS-F0, DFc=Fc-F0, - heat flow rate difference between substance S / calibration substance c and zero-line F0(T), mc-mass of calibration substance, ms – mass of sample, CP,c – heat capacity of calibration substance
10
Binary systems Equilibrium: example Sn-Bi system
11
Binary system Equilibrium: Example Sn-Bi system
12
Scheil Solidification
Fast diffusion in liquid Slow diffusion in solid Local equilibrium
13
Latent heat: Equilibrium vs. Scheil solidification
Equilibrium solidification Scheil solidification
14
Example: Ag-Cu system. Comparison of calculated dHS/dTS with DTA results
Phase diagram and calculated dHS/dTS for 1 and 5 mass.% Cu. dHS/dTS (J/kgK) DTA results at different heating rates: black -15 K/min, red – 10 K/min, blue – 5 K/min.
15
Example: binary system Ag-Cu
Comparison of experimental DTA with calculated dHS/dTS for alloys with 9, 23 and 28 mass.% Cu: Black line – heating rate 15 K/min, red – 10 K/min, blue 5 K/min. dHS/dTS (J/kgK)
16
General DTA curve analysis for binary system
Alloy Ag-15%Cu: dHS/dTS vs. TS using equilibrium enthalpy. Delta function is eutectic, vertical jump is liquidus. DTA scan for melting and freezing at 5 K/min for Ag-15%Cu alloy: Important points are labeled by i, not important by n.
17
Eutectic reactions (L a+b) vs. Peritectic reactions (L+b a)
The kinetics of peritectic reaction is different from eutectic because the diffusion rate is very different in liquid and substitutional solids. If only interstitial diffusion is required the peritectic reaction occurs more easily. If one assume that no diffusion occurs in the solid upon cooling, solidification merely switch from freezing of high temperature phase L b to freezing of low temperature phase L a. Then b phase usually surrounds a phase resulting in coarser two phase microstructure than eutectic one.
18
Phase diagram with eutectic and peritectic reactions
Example Au-Sn system Sn-rich part of Au-Sn phase diagram (a). Hs and dHs/dTs curves for Sn-25%Au calculated for equilibrium (black) and Scheil (red) conditions (b). DTA scan for Sn-25%Au (c) (c)
19
Types of reactions in ternary system
Sn E2: LAg3Sn+(Sn)+Cu6Sn5 U5:L+Cu3Sn Ag3Sn+Cu6Sn5 P2: L+(Ag)+Cu10Sn3 Cu3Sn
20
Example: system Ag-Cu-Sn
Part of liquidus surface U5 L+Cu3Sn=Ag3Sn+Cu6Sn5, E2 L=Ag3Sn+(Sn)+Cu6Sn5 Phase fraction diagram Sn-25%Ag-10%Cu Vertical section at 25 mass% Ag Latent heat release in equilibrium Latent heat release in Scheil conditions dH/dT in equilibrium
21
Example Ag-Cu-Sn Phase fraction Sn-25%Ag-27%Cu
Latent heat release in equilibrium dHS/dT in equilibrium Latent heat release in Scheil conditions
22
Vertical section construction using DTA/DSC
23
TGA: Thermo Gravimetric Analysis
Determination of the mass as function of temperature. It is applicable for all reaction involving gas phase Decomposition reactions Vaporization and sublimation Desorption Oxidation and reduction Corrosion Small sample Gas flow essential Different temperature modes Linear heating, isothermal, stepwise isothermal
24
Solid lines- TG, dashed lines - DTG
25
TGA: thermogravimetric analysis
TGA measure mass loss or gain with temperature and/or oxygen partial pressure. TGA can be used to investigate red/ox reactions in M/Oxide system and decomposition reactions accompanied by gas formation (e.g. CaCO3CaO+CO2 (gas)) 778°C 3Mn2O3 2aMn3O4+0.5O2 1167°C aMn3O4 bMn3O4 1458°C bMn3O4 3MnO+0.5O2 Calculated phase diagram of the Mn-O system pO2=10-2 bar
26
Good praxis for DTA experiments
Problems Influence of mass and heating rate: Larger mass and larger heating rate produce larger peak, but make detection of closely spaced thermal events more difficult. Sample shape is typically not conform to the shape of the sample cup. The thermal contact area between the sample and crucible will change during melting process. Undercooling: Many metals and alloys are prone to undercooling before the nucleation of solid phase start from melt. Nucleation temperature can differ from liquidus up to 100 or more degree depending on nature of alloy system and other factors. Determination of melting on heating is more reliable. Good praxis for DTA experiments Calibration: Based on the melting point of pure substances. Crucible, standard material, heating rate, sample mass, atmosphere are kept constant Characterisation: The composition of samples and crystal structure have to be investigated before and after the measurement Combination: DTA experiments tell us that something is happening at a specific temperature. They do not tell us, what is happening. Combination with other methods like X-ray diffraction, spectroscopy, microscopic investigation and composition analysis (e.g. Electron probe microanalysis) are required to interpret the results
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.