Download presentation
Presentation is loading. Please wait.
Published byOttavio Gianni Modified over 6 years ago
1
Successful transdradial removal of a dislodged coronary stent
Italo Porto Interventional Cardiology Unit Università Cattolica del Sacro Cuore Rome, Italy Fondi, 28 Giugno 2008
2
Italo Porto, MD, PhD DISCLOSURES
I have no real or apparent conflicts of interest to report.
3
Case Male, 66 years High blood pressure, hypercholesterolemia, family history of cardiovascular disease History of coronary artery disease from 2007 with previous PCI on RCA (BMS) and LAD (DES) Relapse of angina in August 2008 with positive exercise stress test Fondi, 28 Giugno 2008 3
4
Coronary angiography Fondi, 28 Giugno 2008
RCA: ostial 85% lesion proximally to previously implanted stent Fondi, 28 Giugno 2008 4
5
Coronary angiography Fondi, 28 Giugno 2008
LAD: mid-distal 55% lesion no instent restenosis; LCX: distal collateralized occlusion Fondi, 28 Giugno 2008 5
6
Revascularization strategy
Percutaneous coronary intervention to proximal right coronary artery Drug pre-treatment: ASA 100 mg daily, clopidogrel loading dose 300 mg Anticoagulation: heparin 100 UI/kg ACT >250 sec Vascular Access: Right radial artery Guiding catheter: Judkins Right 4 6F Fondi, 28 Giugno 2008 6
7
Wires in Guide Wire: 2xBalance Middleweight Universal 0.014” Fondi, 28 Giugno 2008 7
8
Stent: Chrono Carbostent 3.5 x 16 mm
Direct stenting Stent: Chrono Carbostent 3.5 x 16 mm Fondi, 28 Giugno 2008 8
9
Incomplete ostial coverage
Fondi, 28 Giugno 2008 9
10
Stent: Chrono Carbostent 3.5 x 8 mm
Overlap stenting Stent: Chrono Carbostent 3.5 x 8 mm Fondi, 28 Giugno 2008 10
11
Ooopss…..
12
A simple solution to a complex problem ?
Wrong!!!
13
Capture of dislodged stent
Fondi, 28 Giugno 2008 13
14
Stent withdrawal Fondi, 28 Giugno 2008 14
15
Stent withdrawal Fondi, 28 Giugno 2008 15
16
What next? A 3.5 mm deployed stent is engaging the right radial artery The 6 French radial sheath is exchanged for a 7 French one over the two wires An Amplatz GOOSE NECK Microsnare 4 system is implemented to achieve stent removal Size Matters Fondi, 28 Giugno 2008 16
17
Korean group, CCI 2001
18
Stent snaring Fondi, 28 Giugno 2008 18
19
Stent snaring Fondi, 28 Giugno 2008 19
20
Stent removal Fondi, 28 Giugno 2008 20
23
Stent: Chrono Carbostent 4.0 x 12 mm
Ostium stenting Stent: Chrono Carbostent 4.0 x 12 mm Fondi, 28 Giugno 2008 23
24
Final result Fondi, 28 Giugno 2008 24
25
Final radial artery angiography
Fondi, 28 Giugno 2008 25
26
Discussion Stent dislodgement is an uncommon PCI complication which can be secondary to severe coronary angulation, coronary calcification, and direct stenting Two previous report described transradial stent retrieval using small-balloon technique and forceps We describe a safe and successful retrieve of a coronary stent dislodged in ascending aorta using a Microsnare system Fondi, 28 Giugno 2008 26