MSE Radial Resolution with ‘Dirty’ Lens
2
MSE Radial Resolution with ‘Clean’ Lens
FWHM about a factor 2-3 smaller. ‘Tails’ are reduced. Tried various fitting functions … settled on a spline
3
Actual magnetic Pitch-angle Degrees geometry Angle that MSE sees radius
4
The Poor Spatial Resolution of the Dirty Lens Changes the
Measured Field-Line Pitch Angle by up to ~0.5 degrees.
5
There is Little Difference in Inferred Pitch-Angle between a Local
And a Radially-averaged MSE Measurement with a Clean Lens
6
Measured MSE Light is the Convolution of the Diagnostic Response (spatial resolution) and MSE Emission Intensity Profile At edge: emission intensity weighted to inside of viewing region. At center: emission intensity weighted to outside of viewing region. Beam-into-gas: emission intensity weighted to outside of viewing region. MSE Edge Channel Density profile MSE Intensity Profile radius
7
Edge Density Gradients in L- and H-modes
L-mode H-mode