From: Retinal Prosthesis Safety: Alterations in Microglia Morphology due to Thermal Damage and Retinal Implant Contact Invest. Ophthalmol. Vis. Sci ;53(12): doi: /iovs Figure Legend: Schematic of experimental apparatus for implant contact eyes. Dissected eye cups were placed in a Petri dish containing Ames' medium bubbled with carbogen on a heated plate such that bath temperature was maintained at 36.7°C. Examples of an incubation control eye and implant contact response eye are presented. The implant contact eye was placed in a specialty holder, and an X,Y,Z micromanipulator was used to centrally position the implant in contact with the ganglion cell side of the retinal surface. The implant was constructed by attaching a 20 Ω surface-mounted resistor to a 1000 Ω platinum resistance temperature detector (PRTD), with the two 1.6 × 0.85 mm components overlaid to form a cross (heating element surface area, 1.36 mm2). Temperature at the surface of the retinal tissue was detected with the PRTD sensor in the implant, while bath temperature was detected by a sensor in the bath. The Petri dish, specialty holder, eye cup, and implant are all to scale (scale bar, 10 mm). Date of download: 10/10/2017 The Association for Research in Vision and Ophthalmology Copyright © All rights reserved.