ABG Instrumentation Analytical Errors
2
Levy-Jennings Chart In Control
+2SD X -2SD All values vary about the mean in a regular manner
3
Random Error +2SD X -2SD Only one point fell out of the + range.
Causes: Statistical probability: rerun control Sample contamination: repeat analysis on different machine Sample mishandeling:
4
Random Errors Contributing Factors Corrective Actions
Statistical probability Contamination of the sample Improper handling of the sample Corrective Actions Reanalyze control sample after performing a 2-point calibration If this works no further action is needed If this does not work, follow manufacturer recommendations
5
Systematic or Bias Error Trend
+2SD X -2SD Causes: Contaminated buffers: Perform function check of suspected problem area Incorrect gas concentrations Incorrect procedures Component failure: replace, repair failed component
6
Systematic or Bias Error Shift
+2SD X -2SD The plotted points all fall with in the + 2SD range, but the mean value has shifted from where it was at the onset of data accumulation
7
Bias errors Contributing Factors Corrective Actions
Contaminated buffer solutions Incorrect gas concentrations Component degradation/ failure Incorrect procedure Corrective Actions Troubleshoot suspected problems Repair or replace problematic components Re-run 2 point calibration