New Pressure Wires and Catheters Systems for Measuring FFR:

Similar presentations


Presentation on theme: "New Pressure Wires and Catheters Systems for Measuring FFR:"— Presentation transcript:

1 New Pressure Wires and Catheters Systems for Measuring FFR:
Opsens FFR Technology and Clinical Data Update Morton J. Kern , MD MSCAI, FACC, FAHA Chief of Medicine VALB Healthcare System Long Beach, California Professor of Medicine, UCI Orange, California 28 Oct

2 Disclosure: Morton J. Kern, MD Within the past 12 months, the presenter or their spouse/partner have had a financial interest/arrangement or affiliation with the organization listed below. Company Name Relationship Abbott Medical Inc. Speakers’ Bureau Philips Volcano Speakers’ Bureau Merit Medical Inc. Consultant Acist Medical Inc. Consultant Opsens Consultant Heartflow Consultant

3 Some Historical Challenges in FFR
Accuracy – Dealing with drift Need for a workhorse wire – Dealing with pressure wire handling characteristics Connections – Dealing with the procedural challenges of connecting and re-connecting during a case without impacting the pressure readings

4 Pressure Sensor wire and catheter Technologies
Piezo-Resistive Sensors Micro-Catheter Optical Fiber Technology Specially constructed ” wires Sensor at 3cm from tip Piezo-electric sensor Performance thought below dedicated coronary wires St. Jude Medical Volcano Therapeutics Specially constructed ” nitinol wire Sensor i3cm from tip Fiber-optic sensor Performance thought below dedicated coronary wire Opsens Boston Scientific Rapid Exchange Micro catheter over wire with sensor incorporated into distal end Fiber-optic sensor Use over personal best wire Acist Medical

5 Pressure Wire/Catheter Construction
Piezo-Resistive Sensors Micro-Catheter Optical Fiber Technology Optical fiber

6 Available Sensor Construction
Piezo-Resistive Sensors Micro-Catheter Optical Fiber Technology Piezo Resistive Fiber Optic Fiber Optic

7 Opsens OP2 Pressure Wire-Unique Characteristics
Deliverability: Improved wire handling with spiral nitinol construction Reliability: Improved signal stability with optical sensors Connectivity

8 Nitinol based construction enhances guide wire performance coupled with optical pressure sensor - Opsens Inc. Optical fiber Handle FOIC optical connector to OptoMonitor 2 meter cable Laser cut spiral stainless steel Larger diameter Nitinol core – enhanced torque characteristic Same form factor of market available guide wire. Respecting the FFR cut off value of landmark studies.

9 FFR Guidewire torque performance
OptoWire Torque transmission compared with two existing pressure guide wires and BMW.

10 What is the cause of signal drift?
AO/Procedure Changing height of AO transducer Capillary forces in catheter Wedging guide in Ostium Needle in Y-connector Catheter with side-holes Piezo-Resistive sensor Temperature shift from room to patient Calibration error Blood, saline remnants on connector Microscopic-air bubbles trapped inside sensor capsule Optical sensor Cavity width using a Fabry–Pérot white-light interferometer Sensor not vulnerable to electrical interference.

11 Optical Pressure Sensors – Environment Effects
Conventional Optical: Temperature-induced pressure shift - - CTE silicon ≠ CTE Pyrex. Moisture-induced pr drift - Adhesive swelling induces mechanical stress Opsens Conventional Optical Opsens: No temperature or moisture-induced shift - Corrugated diaphragm with central dot compensates moves diaphragm outward. Small amount of adhesive.

12 Zero drift test (mmHg / hr) <Not specified in IFU
FFR with Opsens - Zero drift1 Opsens - Accurate FFR measurement showing minimal drift1. Published drift characteristics OpSens St. Jude2 Volcano Boston Scientific3 Acist4 Zero drift test (mmHg / hr) < 1 < 7 <Not specified in IFU < 3 Per IEC ed3 St-Jude documentation # BSCI Dr. Chambers presentation on TIP conference 2016 Acist 510K K filing.

13 Electrical pressure wire
144 lesions (11%) showed PD>3mmHg! Assessed after 1 measurement Courtesy of Hitoshi Matsuo M.D.,Gifu Heart Center

14 Optic fiber pressure wire Pressure drift>3.0mmHg (4/102=4%)
Pressure wire use as the workhorse wire (2016/1/ /10/12) 102 lesions from 95 patients Total procedure time (min) 98±29 min Total contrast volume (ml) 110±43 Device exchange (times) 6±2 Max pressure (atm) 21±4 Rotablator/Jailed wire (case) 10/13 Drift in FFR value 0.013±0.013 [-0.07 – 0.02] Drift in mmHg 1.2±1.1 [-5.0 – 2.0] Optic fiber pressure wire Pressure drift>3.0mmHg (4/102=4%) FFR drift>0.03 (6/102=6%) Assessed after whole procedure ミリマーキュリー Courtesy of Hitoshi Matsuo M.D.,Gifu Heart Center

15 Procedural FREEDOM and ease of use
Opsens - Optical Pressure Guidewire Optical contact, not electrical contact and not affected by saline and blood. Reliable and simple disconnect and reconnect. Piezo resistive Piezo resistive pressure guidewire with electrical contact, more susceptible to moisture - saline and blood. Electrical contacts alignment is critical in order to get accurate readings. Physicians fear disconnecting current FFR wires due to constant loss of signal. You can disconnect with faith OptoWire, perform PCI and reconnect once you need accurate measurement. This is possible because of Opsens’ optical sensing technology with user-friendly connector. You can also perform reconnection for multi vessels or to obtain post PCI FFR measurement. Conclusion statement: FFR procedure is easier and simplify by Opsens’ connectivity to be almost a routine PCI procedure. Electrical contacts

16 Optical Sensor connection not effected by blood – Courtesy of Dr
Optical Sensor connection not effected by blood – Courtesy of Dr. Nico Pijls

17 FFR with Opsens Optowire 2
Most Accurate FFR Device Zero Drift* – Opsens’ latest generation fiber optic sensor Fast, stable and accurate pressure signal FFR Workhorse Wire Excellent torquability, steerability, and deliverability Using the FFR guidewire as standard PCI guidewire Confidently disconnect for better vessel manipulation, over the wire procedures and post-stent FFR. Continuous display of pressure throughout procedure