Download presentation
Presentation is loading. Please wait.
1
Accuracy Significance in Measurement.
Arumugham.V Calibration Cell BMTWing, SCTIMST
2
Index Accuracy – Concepts. Definition. Terms related to Accuracy.
Significance of Accuracy. How Accuracy specified ? Expressing Accuracy.
3
Accuracy- Concepts Accuracy of a measuring instrument: is a qualitative indication of the ability of a measuring instrument to give responses close to the true value of the measurand (parameter being measured.) [VIM, 5.18] This accuracy is a design specification and it is what is verified during calibration.
4
Accuracy- Concepts Accuracy of a measurement: is a qualitative indication of how closely the result of a measurement agrees with the true value of the measurand. [VIM, 3.5] Because the true value is always unknown, accuracy of a measurement is always an estimate. An accuracy statement by itself has no meaning other than as an indicator of quality. It has quantitative value only when accompanied by information about the uncertainty of the measuring system.
5
Accuracy- Concepts Measured Value
6
Definition of Accuracy
ISO[5725-1] “closeness of agreement between a test result and the accepted reference value” ISO[3534-2] “closeness of agreement between a test result or measurement result and the true value” [VIM 3] “closeness of agreement between a measured quantity value and a true quantity value of a measurand”
7
Terms related to Accuracy
Precision Resolution Error Bias Tolerance Uncertainty
8
Precision The degree to which an instrument will repeat the same measurement over a period of time under the same conditions.
9
Resolution Resolution is 0.01 Volt Resolution is 1 Volt
The smallest change in a measured value that the instrument can detect. Resolution is also known as sensitivity. Resolution is 0.01 Volt Resolution is 1 Volt
10
Error Difference between measured value and accepted reference value in an instance is called as 'ERROR'. E = 5.0 – 4.6 = 0.4 Volts
11
Bias The difference between the mean of repeat measurement results and the accepted reference value or true value is called the bias of the measurement process.
12
Tolerance The unwanted but acceptable deviation from a desired dimension. It is the probability of largest error in a measurement. It can be estimated as half of the largest measurement and smallest measurement. T = 1/2 ( largest measurement - smallest measurement)
13
Uncertainty Uncertainty of measurement [Sec3.9, VIM]: Parameter, associated with the result of a measurement, that characterizes the dispersion of the values that could reasonably be attributed to the measurand Calibration determines the measurement value’s deviation from a standard with a known value in order to assess and apply a systematic correction to those measurements. The associated uncertainty of every calibration result is also calculated. The calibration certificate compiles measurement results with their associated expanded uncertainties regardless of the instrument specifications.
14
Uncertainty Sources Environment Reference Equipment
Metrologist / Operator Measurement Procedure
15
Accuracy in measurements
Measurements with adequate accuracy improves credibility of reporting numerical results and their application in research / testing. But how to ensure that they are adequate ? By using equipments with stated accuracy Reproducing results with other standard methods
16
Significance of Accuracy
By knowing the accuracy in a measurement, one will be always aware of the deviation in the measurement results. Hence improves the confidence in measurement results or performance of equipments. Note: In short accuracy is one of the contributing element in uncertainty estimations.
17
Applicable to Equipments Methods
Ideally a measuring device should be both accurate and precise, with all measurements close to and tightly clustered around the reference value.
18
How accuracy can be specified
According to equipment literature / manuals According to the requirement As per reference/standard literature As per norms specified by standards (TAR/TUR)
19
Test Accuracy Ratio The comparison between the accuracy of the Unit Under Test/Calibration (UUT/UUC) and the accuracy of the standard is known as a Test Accuracy Ratio (TAR). However, this ratio does not consider other potential sources of error in the calibration process.
20
Test Uncertainty Ratio
The comparison between the accuracy of the UUT and the estimated calibration uncertainty is known as a Test Uncertainty Ratio (TUR). Based on ANSI/NCSL Z , “Collective uncertainty of the measurement standards shall not exceed 25% of the acceptable tolerance (e.g. Manufacturer specifications)”. This 25% represents a TUR of 4:1.
21
Equipment Literature Scenario
We want to extract plasma from blood using a centrifuge with following conditions. Speed : 3500 rpm ± 10 rpm Temperature : 22 to 25 degrees Time : 20 to 60 minutes A centrifuge with it’s specification is shown in next slide. Can it be used for the specific purpouse ?
22
Equipment Literature Cont.d
Specifications from Manual Capacity L (4 x 400mL) Control Microprocessor Display Digital Max. Speed 15,200rpm Speed Accuracy ± 1% of Reading Temperature 25C ± 5C Timer 9 hr., 99 min. Noise Level <61dBA
23
Equipment Literature Cont.d
Specifications from Manual of Given Centrifuge Capacity L (4 x 400mL) Control Microprocessor Display Digital Max. Speed 15,200rpm Speed Accuracy ± 1% of Reading Temperature 25C ± 5C Timer 9 hr., 99 min. Noise Level <61dBA ± 10 rpm 22 to 25 degrees 20 to 60 minutes Conclusion : The specific centrifuge does meet the requirements
24
Reference / Standard Literature
From Work Procedures. From National / International Standards Scenario We want to measure the temperature of a water bath. The work procedure describes the the incubation should be done in the water in 37C ± 1C. Maximum set temperature for Bath is 100C Two thermometers are shown in next slide, which one to use ?
25
Reference / Standard Literature Cont.d
Thermometer 1 Temperature Range : 0C to 100C Resolution : 0.1 C Accuracy : ± 0.1 C Thermometer 2 Temperature Range : 0C to 100C Resolution : 0.02 C Accuracy : ± 0.05 C
26
Reference / Standard Literature Cont.d
Thermometer 1 Temperature Range : 0C to 100C Resolution : 0.1 C Accuracy : ± 0.1 C Both Thermometer 1 and Thermometer 2 will satisfy the requirement, 37C ± 1C. Thermometer 2 Temperature Range : 0C to 100C Resolution : 0.02 C Accuracy : ± 0.05 C Conclusion : Thermometer 1 is cost effective, so it can be used.
27
Expressing Accuracy as a tolerance limit as a % of Nominal Capacity
as a % of measurement reading as a % of operating range as a combination of any of the above.
28
Tolerance Limit
29
% of Nominal Capacity
30
% of Measurement Reading
Voltage DC Specification Maximum Voltage: V Accuracy: Fluke 83 V: ± (0.1% of Reading) Fluke 87 V: ± (0.05% of Reading) Resolution: Fluke 83 V: 100 µV Fluke 87 V: 10 µV
31
% of Operating Range Temperature Probe Operating Range -20°C to +135°C
Accuracy <0.1% upto 100.0°C Response time 0.8 sec. Electrode Material Stainless Steel Electrode Dimension 1.2 mm
32
NABL 141 Guidelines for estimation of uncertainty in measurement
FOR MORE INFO... NABL 141 Guidelines for estimation of uncertainty in measurement A Guide for ISO – 9000 : CALIBRATION ISO Standard Fluke : NATA : NIST Hand Book :
33
Thank You.....
Similar presentations
© 2024 SlidePlayer.com. Inc.
All rights reserved.