Download presentation
Presentation is loading. Please wait.
1
Low intensity versus high intensity CXL
Arthur Cummings FRCSEd Wellington Eye Clinic, Dublin, Ireland
2
Financial disclosure:
I have no financial interest in the materials being presented in this presentation
3
Purpose To review the clinical data of corneal cross-linking with the IROC UV-X 2000 To determine the performance and safety of corneal cross-linking with the UV-X 2000 compared to the IROC UV-X 1000
4
Important factors in CXL
Light sources Wavelength Time Energy dose Riboflavin Concentration Spatial distribution Oxygen Reactive environment Type of collagen Radical absorbers Additives
5
Main principles of today‘s photooxidative CXL
The photooxidative cross-links are induced by a photosensitizer (Riboflavin) and UV-light UV-light 3O2 1O2 + Tissue molecule + Cross-linked tissue molecule Riboflavin 3RF* triplet Riboflavin singlet 1RF* Riboflavin RF
6
Induction of Cross-links
Activation of photosensitizer induce 1O2 3O2 decay Molecule products interaction RF Additives Generation of singlet oxygen The induction of cross-links depends on: Amount of singlet oxygen molecules Probability of inducing cross-links Induction of cross-links
7
Rate equation for generation of cross-links
Maximize Minimize Is it possible to shorten the CXL-treatment? Is it possible to make the CXL-treatment more effective?
8
Rate equation for generation of cross-links
Maximize Minimize Factors which can be influenced Concentration Intensity Energy dose (time) Additives Glucose D2O Sodium acids 11
9
Intensity CXL-Congress 2008, R. Krueger & E. Spörl Schumacher et al., IOVS, 2011 10 mW/cm2 9 min 3 mW/cm2 30 min Control 2 mW/cm2 45 min 3 mW/cm2 30 min 10 mW/cm2 9 min 15 mW/cm2 6 min Further increase possible ? CXL induces an 1.3/1.5 fold increase in corneal stiffness. No statistical differences between doses. 12
10
Oxygen depletion Higher intensities lead to a faster depletion of oxygen Diffusion of oxygen into the cornea takes several minutes Missing reactive partners in short time might lead to a reduced effect Are we inducing hypoxia with CXL?
11
Summary CXL is a complex physical process Consider concentration of
Photons Riboflavin molecules Oxygen Reactive partners (collagen, …)
12
Biomechanical changes in standard and rapid CXL
Jeremy Wernli1 Silvia Schumacher1 Eberhard Spörl2 Michael Mrochen1 Biomechanical changes in standard and rapid CXL
13
Purpose What is the problem? Progressive keratectasia
UV-light corneal-crosslinking (CXL) Standard treatment parameters 3 mW/cm2 for 30 minutes Disadvantages Patient’s comfort Surgeon’s patient throughput Goal Shorter treatment time! Question How far can we go? VISION - OPTICS - LIGHT we focus your ideas!
14
Bunsen reciprocity law: Energy dose of standard protocol:
Purpose Bunsen-Roscoe reciprocity law: Bunsen reciprocity law: Photochemical processes depend on the absorbed energy dose Energy dose: Energy dose = Intensity x Time Energy dose of standard protocol: 3 mW/cm2 x 30 min = 5.4 J/cm2 3mW/cm2 30 min 10 min 9mW/cm2 30mW/cm2 3 min 90mW/cm2 1 min Same CXL effect in theory…really? VISION - OPTICS - LIGHT we focus your ideas!
15
Purpose …really? ok ok ? Established standard protocol
3mW/cm2 30 min 10 min 9mW/cm2 30mW/cm2 3 min 90mW/cm2 1 min Same CXL effect in theory… really? ok ok ? Established standard protocol Scientific proof of biomechanical strengthening effect is insufficient in the literature Schumacher et al., IOVS, 2011 Let’s prove! (ex vivo first) VISION - OPTICS - LIGHT we focus your ideas!
16
Methods Experimental study design
Cross-linking of ex vivo porcine corneal tissue In compliance with the ARVO Statement for the Use of Animals in Ophthalmic and Vision Research 180 porcine eyes: Randomly assigned to 10 different treatment groups with different CXL illumination intensitites ranging from 3 to 90 mW/cm2 and correpsonding illumination times from 30 minutes to 1 minute constant energy dose of 5.4 J/cm2 To each of theses groups control eyes of the same date were tested, which were not irradiated with UV light, but otherwise underwent the same procedure Two main steps: Cross-linking procedure Stress-strain measurement Stiffness (Young‘s modulus) as a derivative of the stress-strain curve at 10% strain Nonparametric Kruskal-Wallis test for the comparison between groups (not normally distributed population) VISION - OPTICS - LIGHT we focus your ideas!
17
Comparison of the different groups to the control group
Results Comparison of the different groups to the control group Table 2 - Post hoc analysis of Kruskal-Wallis test. All treatment groups (‘Group B’) were compared against the control group (‘Group A’). Q represents the calculated values which are compared to Q(0.01) from the Kruskal-Wallis test. VISION - OPTICS - LIGHT we focus your ideas!
18
Results Relative stiffness increase
Average increase in stiffness of all groups follows a typical threshold function (Boltzmann function) Function dips at about 40 mW/cm2 50% limit is associated at approximately 47mW/cm2 ± 1.5mw/cm2 Figure 4 – Stiffness increase of all treatment groups compared to control group. The second x-axis at the top indicates the corresponding irradiation times to maintain a constant energy dose of 5.4 J/cm2 VISION - OPTICS - LIGHT we focus your ideas!
19
Discussion Why does the Bunsen-Roscoe law fail?
We can only speculate at this time Hypothesis: Suppression of initiating processes at high intensities Oxygen availability for reaction Riboflavin bleaching Activation of termination processes at high intensities Reaction of activated Riboflavin Oxygen radical Other substances pre-existing in the cornea (e.g. Vitamin C as quencher) VISION - OPTICS - LIGHT we focus your ideas!
20
Optimized beam profile
Peripheral intensity 30% higher than ENERGY DOSE (not intensity) = Increased biomechanical strengthening of the cornea Improved corneal flattening
21
Corneal cross-linking procedure
Abrasion of corneal epithelium Application of Riboflavin solution to cornea Illumination of cornea with UV-light
22
Standard and high intensity procedure
Standard protocol UV-X 1000 3 mW/cm2 for 30 minutes 5 min min min 30 min Diffusion of Riboflavin Illumination with UV light Abrasion Check diffusion High intensity protocol UV-X 2000 9 mW/cm2 for 10 minutes Abrasion Check diffusion Diffusion of Riboflavin Illumination with UV light 5 min min min min
23
UV-X 2000: Optimized beam profile
Hypothesis: more flattening of the cornea due to optimized beam profile
24
UV-X 2000: Optimized Beam Profile
Crosslinking of more corneal tissue Post OP comparison of demarcation line shows more tissue crosslinked in the periphery with the UV-X 2000
25
Center treatment on thinnest point
Treatment should be centered on the thinnest point with UV-X 2000 to avoid direct illumination of limbus Protect limbus When doing a treatment it is important to center the the treatment on the thinnest point and not on the center of the pupil. In keratoconic eyes the thinnest point is mostly not the center of the pupil Dislocation of the cone and the thinnest point Seiler et al. (2010)
26
Respect corneal thickness profile
Corneal thickness profiles are different for normal and keratoconus cornea. Ambrosio et al. (2006)
27
Results for endothelium safety (human donor eyes)
10 mW / cm2; 9 min Pair Untreated Treated 24 hours before 24 hours after M1529B 2800 2900 M15309 2600 2400 M15457 3150 3000 No statistical difference in endothelium cell count or apoptosis Apoptotic cell appear green Only the anterior region of the stroma showed apoptosis No apoptosis was found in the endothelium Untreated controls had no UV but otherwise were treated in the same way as the treated corneas. Hilarby at al. 2011, University of Manchester
28
Comparing different UV-A Intensities
IOP = 8mm Hg Control 3mW 10mW 30mW 100mW Backward SH Imaging
29
Comparing different UV-A Intensities
IOP = 11mm Hg Control 3mW 10mW 30mW 100mW Backward SH Imaging
30
Comparing different UV-A Intensities
IOP = 16mm Hg Control 3mW 10mW 30mW 100mW Backward SH Imaging
31
Comparing different UV-A Intensities
No IOP measurement Control 3mW 10mW 100mW Backward SH Imaging
32
Comparing different UV-A Intensities
No IOP measurement Control 3mW 10mW Forward SH Imaging
33
Clinical Outcomes Ethics Committee Approved Study
Collecting UV-X 2000 data Comparing to database of UV-X 1000 cases and new UV-X 1000 cases performed during the clinical trial Recruitment has stopped Analysis ongoing
34
Outcome measures Performance: Safety: Adverse events Change in K-Max
Change in BSCVA Endothelial cell loss Adverse events
35
Data source Clinical data collected by:
A. Cummings, Wellington Eye Clinic, Ireland T.S. Seiler / T. Koller, IROC Clinic, Switzerland F. Raiskup, TU Dresden, Germany June 2012: 92 eyes treated 33 eyes available for interim data analysis Up to 12 months follow-up
36
Demographic data Number of eyes (OD/OS): 33 (18/15) (55%/45%)
Gender: % male 18 % female Age (years): mean: 26.2 ± min: 15 max: 40 Kmax (D) (Patients with pre-op Kmax > 65 D were excluded) mean: 52.9 ± 5.8 min: 41.8 max: 64.0 Sphere (D) (11 eyes): mean : 0.6 ± 1.7 min: max: 3.75 Cylinder (D) (11 eyes): mean: -3.4 ± 2.4 min: max: -0.5 Corneal thickness (micron) (21 eyes): mean: 477 ± 39 min: 392 max: 533
37
Comparison UV-X 1000 and UV-X 2000
Interim data analysis for UV-X 2000 at 6 months Status Kmax (D) Number of eyes (%) Flattening < -1 D 14 (44.8) Same from -1 D to 1 D 15 (46.9) Steepening > 1 D (9.4) UV-X 2000 (32 eyes): After 6 months, progression is halted in 90.6 %. Status Kmax (D) Number of eyes (%) Flattening < -1 D 39 (37.1) Same from -1 D to 1 D 58 (55.2) Steepening > 1 D (7.6) UV-X 1000 (117 eyes): After 12 months, progression is halted in 92.4 %. Comparable results Higher percentages of eyes are flattened with UV-X2000 Equal rate to stop progression after 6 months UV-X2000 compared to 12 months UV-X 1000 We expect further improvement of flattening over time for UV-X 2000
38
UV-X1000 improvements over time
-0.75D -1.5D -2.0D -2.5D We expect further improvement for UV-X 2000 from 6 to 12 months as known from UV-X1000 results Raiskup et al. 2008
39
Performance – Change in Kmax
Decreased Unchanged Increased 7 eyes show a strong flattening effect at 6 months
40
Performance – Change in Kmax
More flattening for higher Kmax values
41
Performance – Change in Kmax
Progression of corneal steepening 6 month follow-up in 32 eyes Mean Kmax = -1.0 D Status Kmax (D) Number of eyes (%) Flattening < -1 D 14 (44.8) Same from -1 D to 1 D 15 (46.9) Steepening > 1 D (9.4) Approx. 45% of the eyes are significantly flattened In more than 90 % the progression of ectasia has stopped
42
Safety – Change in BSCVA
Mean increase in BSCVA by 1 or more lines: 36% No eye lost more than 2 lines Safety is comparable to published data. Koller / Raiskup 80% of eyes are equal or improved in BSCVA compared to pre-OP 3 eyes lost two lines of BSCVA at 6 months None of the eyes lost more than 2 lines
43
Safety – Change in BSCVA
UV-X 2000: 0 eyes lost more than 2 lines (this data summary) min: 1-2% lost more than 2 lines (Raiskup et al. 2008) UV-X 1000: 1-2 % lost more than 2 lines (Hersh et al. 2011) Comparable safety for UV-X 2000
44
ECC: Comparison UV-X 1000 vs UV-X 2000
Number of eyes Follow-up in months Cell loss? Statistitical significance of endothelial cell loss Percentage of endothelial cell loss 8 6 N.S. No UV-X 2000: After 6 months: No statistically significant cell loss reported Number of eyes Follow-up in months Endothelial cell loss in % or reported damages? Statistitical significance of endothelial cell loss Percentage of endothelial cell loss 88 52 Yes 2% 51 12 No cell loss No N.S. 24 66 36 30 18 14 Standard: After 12 to 52 months: Percentage of endothelial cell loss is less than naturally occurring 1% per year
45
AE: Comparison UV-X 1000 vs UV-X 2000
UV-X 2000 (92 eyes): 3.3% of cases Corneal edema: 1 Corneal edema with Descemet folds: 2 Standard (more than 2800 eyes): 3.5% of cases Corneal scarring Sterile infiltrates Delayed epithelium healing Keratitis Anterior chamber reaction Corneal edema Comparatively low incidence of adverse events
46
Risk of UV damage UV-X 2000 has equal total energy dose compared to UV-X 1000 Thus, lens and retina get the same theoretical low UV light dose
47
Add…….. Graphs from WEC comparing AXL to CXL off IBRA
48
WEC Clinical Data AXL CXL
49
Typical clinical observation UV-X 2000
Corneal haze ring can be found in the mid-periphery of the cornea. Similar to UV-X 1000, the haze is reduced within the first 6 months
50
Summary Performance: Strong flattening effect
More flattening for steeper corneas Approx. 45% of the eyes are significant flattened In more than 90 % the progression of ectasia has stopped Further improvement expected in corneal flattening from 6 to 12 months follow-up Safety: Comparable safety regarding vision & endothelium Comparable theoretical risk for UV damage of lens & retina Adverse events: Comparable low rate of adverse events
51
Conclusion The UV-X 2000 provides: better performance,
equal safety and a shorter treatment time compared to standard corneal cross-linking procedures.
52
Ongoing Studies at the WEC
Epi-ON CXL (CXLO) Conductive Keratoplasty (CK) + CXL
53
4th Keratoconus Experts Meeting
Pre-ESCRS One item on agenda Define parameters to measure keratoconus progression and regression John Kanellopoulos ISV IHD
54
9th CXL Meeting 6-7 December 2013 Dublin, Ireland
55
Thank you for your attention
Similar presentations
© 2024 SlidePlayer.com. Inc.
All rights reserved.