Download presentation
Presentation is loading. Please wait.
1
The physics of Radiation Therapy, pp. 315 - 345
Ch13 Treatment Planning III: Field Shaping, Skin Dose, and Field Separation
2
13.1 Field Blocks A. Block Thickness B. Block Divergence
13.2 Field Shaping A. Custom Blocks B. Independent Jaws C. Multileaf Collimators 13.3 Skin Dose A. Electron Contamination of Photon Beams B. Measurement of Dose Distribution in the Build-up region C. Skin Sparing as a Function of Photon Energy D. Effect of Absorber – Skin Distance E. Effect of Field Size F. Electron Filters G. Skin Sparing at Oblique Incidence 13.4 Separation of Adjacent Fields A. Methods of Fields Separation B. Orthogonal Field Junctions C. Guildlines for Field Matching
3
A. block Thickness Material of block Required thickness of block
lead (commonly) Required thickness of block Beam quality The allowed transmission (5% transmission is acceptable)
4
A. block Thickness Calculation of the of the number (n) of HVL to achieve 5% transmission Recommended of block thickness = 4.5 – 5 HVL Further increase of block thickness is not significant The predominance of scattered radiation from the adjoining open areas of the field
5
Recommended Minimum Thickness of Lead for Shielding Beam Quality
Required Lead Thickness 1.0 mm Al HVL 0.2 mm 2.0 mm Al HVL 0.3 mm 3.0 mm Al HVL 0.4 mm 1.0 mm Cu HVL 1.0 mm 2.0 mm Cu HVL 2.0 mm 3.0 mm Cu HVL 2.5 mm 137 Cs 3.0 cm 60Co 5.0 cm 4 MV 6.0 cm 6 MV 6.5 cm 10 MV 7.0 cm 25 MV * Approximate values to give ≦5% primary transmission
6
B. Block Divergence The blocks is shaped with the geometric divergence of the beam Minimized the block transmission penumbra Little advantage for beams with large geometric penumbra (e.g. 60Co) Most suited from beams having small focal spots
7
C. Multileaf Collimators
13.1 Field Blocks A. Block Thickness B. Block Divergence 13.2 Field Shaping A. Custom Blocks B. Independent Jaws C. Multileaf Collimators 13.3 Skin Dose A. Electron Contamination of Photon Beams B. Measurement of Dose Distribution in the Build-up region C. Skin Sparing as a Function of Photon Energy D. Effect of Absorber – Skin Distance E. Effect of Field Size F. Electron Filters G. Skin Sparing at Oblique Incidence 13.4 Separation of Adjacent Fields A. Methods of Fields Separation B. Orthogonal Field Junctions C. Guildlines for Field Matching
8
A. Custom Blocking Lipowitz metal (Cerrobend®) Main advantage
A low melting point alloy Density: 9.4g / cm3 at 20℃ (83% of lead) Consisted of : bismuth (50%), lead (26.7%), tin (13.3%), cadmium (10%) Main advantage Melting point: 70℃ (c.f. 327℃of lead) Harder than lead in room temperature Common thickness of block (Cerrobend®) using in MV range of photon 7.5 cm Density ratio relative to lead (~lead thickness × 1.21) Equivalent to about 6 cm of pure lead
9
A. Custom Blocking Procedure for constructing Cerrobend block
Outline of the treatment on simulator radiography (port film) Construct divergent cavities in a Styrofoam block by cutting device Cutted stylofoam was placed on Lucite plate and carefully aligned relative to the central axis Pouring Cerrobend into the cavity Mounting the block on the Lucite plate
14
B. Independent Jaws Blocking off a part of the field without changing the position of the isocenter In the past: using beam splitters In present, simply moving the collimators (or jaws) independently Application: matching field (blocked off at the central axis to remove divergence) Effects of the asymmetric collimation Change in physical penumbra The tilt of the isodose curves toward the blocked edge Eliminating photon and electron scatter from the blocked portion of the field Reducing the dose near the edge
15
Fig.13.3 Comparison of isodose distribution with half the beam blocked by an independent jaw versus a block on a tray Notice close agreement as well as the tile of the isodose curves toward the blocked edge
16
C. Multileaf Collimators
Large number of collimating blocks (or leaves) that can be driven automatically and independently, to generate a field of any shape Thickness of the leaf (along the beam direction): ≦1% transmission Width of the leaf: 1 cm as projected at the isocenter (usually)
17
C. Multileaf Collimators
Double focused MLC Form a cone of irregular cross-section diverging from the source position Move on a spherical shell centered at the source Provide a sharp beam cutoff at the edge Difficult to manufacture In high-energy beams, the dose falloff at the edge is largely determined by scattered photons and electrons Rounded leaf edges MLC Directions of travel perpendicular to the central ray Provide constant transmission through a leaf edge, regardless of its position in the field
18
Conformity between the planned field boundary and MLC boundary
Continuous vs. jagged Depending on: projected leaf width, shape of the target volume, angle of the rotation of the collimator Optimization of the MLC rotation Direction of motion of the leaves is parallel with the direction in which the target volume has the smallest cross-section Potential applications of the MLC systems Replacement of custom Cerrobend blocking Automatic beam shaping for multiple fields Dynamic conformal RT (beam shaping during rotation of the gantry) Modifying dose distributions within the field (computer-controlled dwell time of the individual leaves)
19
A. Electron Contamination of Photon Beams
13.1 Field Blocks A. Block Thickness B. Block Divergence 13.2 Field Shaping A. Custom Blocks B. Independent Jaws C. Multileaf Collimators 13.3 Skin Dose A. Electron Contamination of Photon Beams B. Measurement of Dose Distribution in the Build-up region C. Skin Sparing as a Function of Photon Energy D. Effect of Absorber – Skin Distance E. Effect of Field Size F. Electron Filters G. Skin Sparing at Oblique Incidence 13.4 Separation of Adjacent Fields A. Methods of Fields Separation B. Orthogonal Field Junctions C. Guildlines for Field Matching
20
13.3 Skin Dose Skin sparing is one of the most desirable features of high-energy photon beams This effect may be reduced or even lost if the beam is excessively contaminated with secondary electrons
21
A. Electron Contamination of Photons Beams
Contribution to the skin dose Electron contamination Backscattered radiation from the medium Source of the contaminating electrons Photons interactions in the air, in the collimator, other scattering material in the path of the beam (e.g. shadow tray, thick enough to absorb most of the electrons)
22
A. Electron Contamination of Photons Beams
Controversy as to the relative contribution of secondary electrons vs. low energy scattered photons to the dose of the build-up region Field size , PDD in the build-up region (dmax shift to a shallower depths) Current evidence favors that the secondary electrons is the predominant cause of the effect
23
B. Measurement of Dose Distribution in the build-up region
The size of the dosimeter should be as small as possible Steep dose gradient Type of the dosimeter used Extrapolation chamber (Fig.6.15, p.113) Fixed-separation plane-parallel ionization chambers (most common used) Variable electrode spacing (section 6.7B) Thin layers (<0.5cm) of thermoluminescent dosimeter (TLD)
24
C. Skin Sparing as the Function of Photon Energy
Factors affecting the build-up region (shown by the studies) beam energy, SSD, field size, configuration of secondary blocking tray The effect becomes more pronounced as photon energy increases
25
Build-Up Dose Distribution in Polystyrene for a10 × 10 cm Field
Depth (mm) 60Co 80 cm 4 MV 10 MV 100 cm 25 MV 18.0 14.0 12.0 17.0 1 70.5 57.0 30.0 28.0 2 90.0 74.0 46.0 39.5 3 98.0 84.0 55.0 47.0 4 100.0 63.0 54.5 5 94.0 72.0 60.5 6 96.5 76.0 66.0 8 99.5 73.0 10 91.0 79.0 15 97.0 88.5 20 95.0 25 99.0 30 * Data from Velkely DE, Manson DS, Purdy JA, Oliver GD. Buildup region of megavoltage photon radiation sources. Med Phys 1975; 2: 14. Phenomenon: The dose increase rapidly within the first few mm, and gradually achieves its max value at the depth of peak dose
26
C. Skin Sparing as the Function of Photon Energy
A practical application Build-up bolus to maximized the dose on the skin e.g. covering a scar with a strip of bolus Thickness of the bolus required < the depth of maximum dose
27
D. Effect of Absorber – Skin Distance
Absorber is introduced in the beam to eliminate the “collimator electrons” (main source of electron contamination) e.g. Lucite shadow tray Thickness > the range of secondary electrons Absorber becomes the principal source of electron contamination Solution: increase the distance between the tray and the surface Reasons: Divergence, absorption and scattering of electrons in the air Preserve the skin sparing e.g. 60Co γ rays, 15 – 20 cm air gap, skin dose <50% of the Dmax
28
Figure 13.5 Effect of Lucite shadow tray on dose buildup for 10 MV x-rays
% Depth Dose Beam: 10 MV Tray thickness: 1.5 g/cm2 Field size: 15 x 15 cm SSD: 100 cm Phenomenon The relative surface dose The Dmax move closer to the surface
29
D. Effect of Absorber – Skin Distance
Beam spoiler A low atomic number absorber (e.g. Lucite shadow tray), placed at an appropriate distance from the surface, can be used to modify the build up curve e.g. H&N cancer using 10 MV x-rays, using a beam spoiler to increase the dose to the surface neck nodes
30
E. Effect of Field Size The relative skin dose depends strongly on field size Field size, the dose in build up region Reasons: increase electron emission from the collimator and air Fig.13.6 Percent surface dose as a function of field size
31
Radius of an equvalent circular field
E. Effect of Field Size Relationship of the field size and the tray-to-skin distance (studied by Salylor and Quillin) The ratio of optimum skin sparing is For large treatment field, tray-to-skin distance is not possible for isocentric treatment Maintaining the skin-sparing effect using electron filter Tray to skin distance Radius of an equvalent circular field Field size Optimum tray-to-skin distance 5 × 5 cm 12 cm 30 × 30 cm 67 cm
32
F. Electron Filter Using absorber of medium atomic number can reduce the skin dose e.g. Z in the range of 30 – 80 Reduces the secondary electron scatter in the forward direction (showed by Hine) Proved to reduce the surface dose, and improve the build-up region of large field in 60Co teletherapy (studied by Khan) Thickness of electron filter Equal to the maximum range of secondary electron e.g. 60Co, 0.5 g/cm2 or 0.9 mm of tin
33
Fig.13.7 Variation of percent surface dose with atomic number of absorber
Field size: 15 × 15 cm Phenomenon The minimum occurs at about Z = 50 (Tin)
34
Fig.13.8 demonstration of the effectiveness of tin in reducing skin dose
35
G. Skin Sparing at Oblique Incidence
Surface Phantom Phenomenon: Angle of incidence , skin dose The concept of electron range surface (suggested by Jackson) The ERS is a 3D representation of secondary electron range and distribution, produced by a pencil beam of photons interacting with the medium Electron generated insides the ERS volume will reach P and contribute dose to there ERS for 60Co γ rays is an ellipsoid with axial dimensions of 5 × 2.4 mm P
36
G. Skin Sparing at Oblique Incidence
For tangential beam incidence 1/2 of ERS is below the phantom surface upper estimate dose to the skin P
37
G. Skin Sparing at Oblique Incidence
Oblique factor (defined by Gerbi et al.) The dose at a point in phantom on central axis of a beam incident at angle θ°, with respect to the perpendicular to the surface Representing dose enhancement due to beam obliquity for the same depth along central axis
38
Fig.13.10 Oblique factor at the surface plotted as a function of beam angle for various energy beams
Phenomenon The angle of incidence , Obliquity factor First gradually and then dramatically beyond 45° 45°
39
13.4 Separation of Adjacent Fields A. Methods of Fields Separation
13.1 Field Blocks A. Block Thickness B. Block Divergence 13.2 Field Shaping A. Custom Blocks B. Independent Jaws C. Multileaf Collimators 13.3 Skin Dose A. Electron Contamination of Photon Beams B. Measurement of Dose Distribution in the Build-up region C. Skin Sparing as a Function of Photon Energy D. Effect of Absorber – Skin Distance E. Effect of Field Size F. Electron Filters G. Skin Sparing at Oblique Incidence 13.4 Separation of Adjacent Fields A. Methods of Fields Separation B. Orthogonal Field Junctions C. Guildlines for Field Matching
40
13.4 Separation of Adjacent Fields
Examples of adjacent fields “Mantle” or inverted-Y fields for Hodgkin’s disease Craniospinal fields for medulloblastoma (orthogonal) Placing lateral neck fields adjacent to the ant. supraclavicular field
41
A number of techniques have been devised to achieve dose uniformity in the field junction
A. Angling the beams aways from each other B. Field separation C. Isocetric split beam technique D. Penumbra generator Fig Schematic representation of various techniques used for field matching
42
Geometry of two adjacent beams
The dose above the junction will be lower The dose below the junction will be higher The total separation S on the surface is given by:
43
An ideal geometry No overlap between a field and its adjacent opposing neighbor
44
Creates regions of “Three-field overlap”
Bigger fields divergence into the opposing smaller fields The maximum length of three-field overlap (ΔS) ΔS = S1 – S2 Field #1 30 × 30 cm Field #1 15 × 15 cm Fig.13.14a Field separate at surface = 2.3 cm
45
Eliminated “Three-field overlap” by adjusting SSD
ΔS = 0, if A cold spot occurs at the midline Field #1 30 × 30 cm Field #1 15 × 15 cm Fig.13.14b Field separate at surface = 3 cm
46
Eliminating the “Three-field overlap” in a specific region
e.g. spinal cord Depth of the cord Midline depth Ant. surface Field #1 30 × 30 cm Field #1 15 × 15 cm 15 cm Depth of spinal cord Fig.13.14c: Field separate at surface = 2.7 cm elimination of the 3-field overlap at the cord
47
B. Orthogonal Field Junction
The central axes of the adjacent fields perpendicular to each other e.g.1. The craniospinal irradiation of medullobastoma Lateral parallel opposed brain field + posterior spine field e.g.2. Head & Neck Lateral parallel opposed head field + anterior supraclavicular field
48
Methods for not possible separation of the field
Beam splitters (abut the field along or close to their central axes) Drawing match line each time before treatment Additional block for sensitive structure (e.g. spinal cord) A geometrical method of orthogonal field separation (Werner et al.) d depth at which the orthogonal fields are allowed to join
49
B.1. Craniospinal Fields Techinque A
Bilateral cranial fields adjacent to a spinal field The inferior border of cranial field meet at a point midway on the posterior neck surface
50
Technique B (1) Rotation the couch and collimator
Calculation of the two angles θcoll and θcouch
51
Technique B (2) Rotation of the collimator only with using hemiblock of the cranial fields Calculation of the collimator angles θcoll only
52
C. Guidelines for Field Matching
The sites of field matching should be chosen over an area that dose not contain tumor or a critically sensitive organs If the tumor is superficial at the junction site, the fields should not be separated since a cold spot on the tumor will risk recurrence For deep-seated tumors, the fields may be separated on the skin surface so that the junction point lies at the midline The line of field matching must be drawn at each treatment session on the basis of the first field treated A field-matching technique must be verified by actual isodose distribution before it is adopted for general clinical use
Similar presentations
© 2024 SlidePlayer.com. Inc.
All rights reserved.