Download presentation
Presentation is loading. Please wait.
1
Pulmonary Artery Pressure Monitoring
On-Line Review January 2009 Reference: Special Care Manual SP-P-6.2 – 6.7 Copyright 2008, SIUH, Staff Development on-line
2
Objectives At the end of the review the participant will be able to:
Describe pulmonary artery waveforms. Discuss pulmonary artery pressures. Identify nursing measures to increase accuracy of measurements. Describe cardiac output determination by intermittant thermodilution technique.
3
Equipment Informed consent, “side-site” verification
7.5 FR pulmonary artery catheter (PAC) 8.5 FR introducer set with “side-port” Central line insertion tray and central line dressing kit with biopatch Flush solution, pressure bag, disposable pressure tubing with transducer Transducer holder
4
Edward’s Life Science VIP Thermodilution Catheter
Thermistor port Proximal Injectate PA Distal Balloon inflation port RA infusion port The standard catheter is 7.5 FR and 110 cm long. Maximal balloon volume 1.5cc
5
Quality Control Issues for PA Catheters
“Zero” system at setup and minimally 1X shift Do square flush wave test minimally 1X shift, after zeroing, obtaining blood samples, with tubing changes, or if waveform distorted. Prime tubing and stopcocks by gravity. Maintain continuous flush system. Use standard (universal) precautions, aseptic technique Measure all values at end-expiration: *In the “valley” for a patient on a ventilator *In the “peak” for a patient breathing spontaneously.
6
Referencing the “zeroing” stopcock to Phlebostatic Axis
The phlebostatic axis is the approximate level of the left atrium. It is located midway between the anterior-posterior chest wall at the 4th intercostal space. The patient need not be flat, but must be supine.
7
Waveforms identified by movement of the catheter through
the right side of the heart into the pulmonary artery and capillary bed: Right atrial (RA) Right ventricular (RV) Pulmonary artery pressure (PAP) Pulmonary capillary wedge pressure (PCWP) or Pulmonary artery wedge pressure (PAWP)
8
Right atrial pressure (RAP) is measured by the distal tip of the PAC
on insertion or by the proximal port post insertion. The waveform has two primary characteristics: “a” wave and “v” wave. Pulmonic valve QRS P Right Atrium a = atrial systole v = ventricular systole Tricuspid valve a wave follows “P” wave on ECG v wave follows the QRS complex on ECG RAP = mean (average) of a wave
9
Measurement of Right Atrial Pressure (RA) from Pulmonary Artery Catheter
Normal RAP = 2 – 6 mmHg The “a” wave represents atrial contraction and estimates atrial filling at end-diastole. Locate the “a” wave by alignment with the PR interval on EKG. Measure the mean (average) of the height of the “a” wave and the bottom of the “a’ wave at end-expiration. *In the “valley” for a patient on a ventilator *In the “peak” for a patient breathing spontaneously
10
RA Waveform “c” wave = closure of the tricuspid valve
“x” decent = follows closure of the tricuspid valve, “y” decent = follows closure of the pulmonic valve
11
Alterations in RA Pressure
Reflects filling volume of the right atrium Low pressure consistent with hypovolemia: Trauma, blood loss Dehydration Loss of fluid from drains Vomiting, diarrhea Burns 3rd spacing of fluid Consistent with tachycardia, ↓ urine output, dry skin & mucous membranes
12
Clinical Situations Consistent with Elevations in RA Pressures
Tricuspid stenosis, regurgitation RV ischemia or failure Pulmonary hypertension Pulmonic stenosis Pulmonary embolism Atrial ventricular dissociation with loss of synchrony Atrial arrythmias, A-V conduction blocks)
13
on insertion or from the RV port (gray) post insertion.
Right ventricular pressure (RVP) is measured from the distal tip of PAC on insertion or from the RV port (gray) post insertion. RVP has a systolic and diastolic component. QRS Systolic pressure Diastolic pressure Right ventricle Onset of systole follows QRS complex on ECG End diastole occurs at the onset of systole.
14
RV Waveform Systole Diastole
15
Reflects filling volume of right ventricle
Alterations in RVP Reflects filling volume of right ventricle Normal RV = 25-30/0-5 mmHg Low pressure consistent with low volume Usually accompanies low RAP Elevation associated with: Hypervolemia Pulmonary embolism Outflow obstruction RV infarct/failure Pericarditis/tamponade LV failure Primary and secondary pulmonary hypertension (PHTN) Pulmonic stenosis COPD
16
Pulmonary artery pressure (PAP) is measured from the distal tip
of PAC on insertion and distal tip post insertion. It has a systolic and diastolic component. QRS Pulmonic valve PA systolic PA diastolic Dicrotic notch Represents closure of the pulmonic valve Systolic pressure follows QRS on ECG. Diastole begins at the closure of the pulmonic valve and continues to next onset of systole.
17
Alterations in PAP Represents filling volume in the pulmonary artery and resistance to flow within the pulmonary circuit Normal PAP = 20-30/8-15 mmHg Low pressure consistent with hypovolemia Consistent with ↓ RAP and ↓ RVP High pressure consistent with PHTN COPD Acute respiratory distress syndrome (ARDS) Pulmonary embolism (PE) Mitral stenosis Left ventricular heart failure
18
PAP Waveform
19
Pulmonary capillary wedge pressure (PCWP) or pulmonary artery wedge pressure (PAWP) is measured from the distal port of PAC with balloon inflated. a = atrial systole v = ventricular systole The PCWP represents left atrial filling pressure reflected through the pulmonary circuit. Normal value 8-14 mmHg. PCWP = mean (average) of “a” wave
20
PCWP indirectly measures left atrial pressure
Catheter tip looks “through” the pulmonary circulation to “see” the left atrial pressure.
21
Measurement of PCWP from Balloon Port
Inflate balloon with only as much air as needed to create a change in the waveform (no more than 1.5cc air). Observe for change in waveform. Locate the “a” wave on the PCWP waveform. Use the QRS as a marker to detect the “a” waveform. Measure the mean of the “a” wave to obtain the PCWP: *In the “valley” for a patient on a ventilator *In the “peak” for a patient breathing spontaneously. PA diastolic (PAD) should correlate to the PCWP and be 0-5 mmHg greater. (PAD ≥ PCWP) High levels of positive end expiration pressure (PEEP) will raise the PCWP. Correlate with MD.
22
Alterations of PCWP Low pressure consistent with hypovolemia
Elevations consistent with: Mitral stenosis/regurgitation Aortic stenosis/regurgitation Acute LV ischemia/infarct LV failure Atrial ventricular dissociation with loss of synchrony Both RA and PCWP elevated in cardiac tamponade, constrictive pericarditis, and hypervolemia
23
Once inserted balloon should remain DEFLATED and ONLY
reinflated for periodic measurement of PCWP. PAP waveform should be displayed at all other times. Distal tip measures PAP and PCWP Proximal tip measures RA and is site for CO measurement and fluid administration RA infusion port (white) Distal tip (Yellow) Proximal tip (Blue) Right ventricular port site for fluid administration Right ventricular Port (Gray)
24
Complications of PA Catheters
Air embolism Infection Atrial & ventricular arrhythmias Hemoptysis ( caused by excessive expansion of the balloon in the wedge position) Areas of infarction Thrombus formation /PE Local endothelial damage Pulmonary rupture Thrombophlebitis Pneumothorax
25
Summary of Pulmonary Artery Pressure Waveforms
Normal 2-6 mmHg Normal 20-30/ Normal 20-30/ Normal 8-14 mmHg mmHg mmHg
26
Cardiac Output by Intermittent Thermodilution Technique
1. Verify catheter placement by X-ray and appropriate PA waveform. Verify that the correct data is programmed into the computer in the CO menu * Make of catheter (Baxter) * Size of catheter (7.5 Fr) * Volume of injectate (10cc) * Use of in-line syringe * Co-factor on "Auto" * Patient temp is displayed on screen (Tb) * Injectate temp is displayed on screen (Ti) (Generally, room temp solution is used. If values are in question, use iced solution) Verify patient position: *Patient supine, HOB no > 20º (If value is doubtful, use flat position.)
27
Causes of Error in CO Determination
Erroneous catheter placement, data entry Concomitant volume infusion through PA catheter (Venous infusion port, RV or side ports) Equipment malfunction (monitor, cable, catheter) Tricuspid or mitral valve dysfunction (stenosis or regurgitation – blood must have forward flow) Fever Shivering Positive pressure ventilation, PEEP Injection technique too long (Should be smooth and no longer than 4 seconds)
29
Congratulations! This completes part one of the review.
Please return to the class page and click on the link to take the post-test. Part two of the review will be conducted “on-unit” with the hemodynamic monitoring simulator. Contact your manager.
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.