Calibration of 6000 Thermometers

Similar presentations


Presentation on theme: "Calibration of 6000 Thermometers"— Presentation transcript:

1 Calibration of 6000 Thermometers
J.P. Thermeau1, T. Junquera1, F. Chatelet1, C. Joly1, F. Mellac1 J. Casas-Cubillos2, C. Balle2, N. Vauthier2 1 Institut de Physique Nucléaire, CNRS-IN2P3, Orsay, France 2 CERN, ACR Division, CH-1211 Geneva 23, Switzerland LHC Inauguration : 31 May - 1 June 2007

2 LHC Inauguration : 31 May - 1 June 2007
Summary LHC thermometry requirements Calibration facility Calibration sequence description Production rate of the LHC calibrated thermometers Conclusions LHC Inauguration : 31 May - 1 June 2007

3 Engineering specifications
For the temperature range 1.6 K – 300 K : - each thermometer needs an individual calibration, - several sensor models could be used → program tests. Program tests to select the sensors : - CERN : thermal cycling → long term stability, - CNRS/IN2P3 : neutron irradiation test in liquid helium (1015 neutron/cm²) → radiation hardness.  CERN have chosen CERNOX sensors produced by Lake Shore LHC Inauguration : 31 May - 1 June 2007

4 LHC Inauguration : 31 May - 1 June 2007
LHC requirements 6000 thermometers units are required, including 500 spare units : thermometers for vacuum condition are mounted on long thermometric blocks,  -  2200 thermometers for liquid helium environment are mounted on short thermometric blocks,  -   500 thermometers are used on cold fingers,    Thermometers overall accuracy requirements : range 1.6 K K 2.2 K - 4 K 4 K - 6 K accuracy ± 10 mK ± 20 mK ± 30 mK range 6 K - 25 K 25 K - 50 K 50 K K accuracy ± 1K ± 5 K LHC Inauguration : 31 May - 1 June 2007

5 View of the calibration facility
Calibration Insert Helium vapor heater Cryostat Dewar LN2, LHe LHC Inauguration : 31 May - 1 June 2007

6 Calibration under vacuum condition
75 thermometers to calibrate Calibration and thermalization blocks are cooled down with the liquid helium loop. T > 30 K : no temperature regulation, T < 30 K : temperature regulates by Proportional Integral regulator using a heater. 7 days to realize a calibration run in vacuum condition LHC Inauguration : 31 May - 1 June 2007

7 Mounting of thermometer with connector
Measuring wires sold on RF connector Thermometric block slides under RF connector to realize the electrical connections LHC Inauguration : 31 May - 1 June 2007

8 Insert for liquid helium calibration
- Copper plate on which 100 thermometers can be fixed. A pressure regulator is used to regulate the temperature of liquid helium, Measurement sequence is the same that in vacuum condition. 2 days to realize a calibration run in liquid helium LHC Inauguration : 31 May - 1 June 2007

9 Calibration procedures
PROCESSUS DIREC DIRECT Réception Cyclage Préparation Étalonnage Sous VIDE Démontage Après Analyse Conditionn t Expédition LIQUIDE 1 2 3 4 5 6 7 8 9 10 Maintenance Des Équipements D'étalonnage 12 Exploitation de la Base de Données 13 PROCESSUS SUPPORTS ET CONTRÔ PROCESSUS SUPPORTS ET CONTRÔLE Contrôle Électrique Unitaire 11 CALIBRATION PROCESS Reception Cycling Vacuum Calibration Setting-up Dismounting after Analysis Tests Shipping Liquid Equipments Data Base CONTROL AND MAINTENANCE Electrical Test and Quality Assurance Plan Calibration sequence for thermometers used in liquid Calibration sequence for thermometers used in vacuum LHC Inauguration : 31 May - 1 June 2007

10 IPNO Calibration Cryostat
Network of the calibration facilities IPNO Calibration Cryostat IPN Network Local Data Base CERN Network Internet CERN PC Calibration Analysis Control & Command PLC Back-up LHC Inauguration : 31 May - 1 June 2007

11 Number of LHC calibrated thermometers since january 2001
1 vacuum insert 1 liquid insert 2 vacuum inserts 2003 2004 2005 2002 2001 LHC Inauguration : 31 May - 1 June 2007

12 Production rate of LHC calibrated thermometers
Pre-serie LHC Inauguration : 31 May - 1 June 2007

13 LHC Inauguration : 31 May - 1 June 2007
Conclusions In five years (2001 – 2005), more than six thousands thermometers have been successfully calibrated for the CERN. 129 calibrations runs in vacuum condition, 44 runs in liquid helium. Two thousands five hundred thermometers can be calibrated per year with three calibration inserts, High calibration accuracy :   7 mK for T < 2.2 K,   10 mK for 2.2 K < T < 30 K, 0.25 % for T > 30 K. Now, the calibration facility is part of SUPRATECH a technological platform for superconducting cavities at Orsay, and can be used for new projects such Spiral2, FAIR, ITER, … LHC Inauguration : 31 May - 1 June 2007