TDR Measurement of MCP Anode
Camden Ertley
2
Index TDR Setup Method Results Conclusion 16 pixels tied together
3
Fast Edge from Scope 100mV and 500ps per div
TDR Setup The fast edge, ~280ps rise time, produced by the scope was used. The signal was split with a SMA T. One T output was connected to the Ch 1 input the other went to the MCP. The cables used were 2.5ns rg58 coax with SMA ends. Scope Fast Edge -0.14 to +3V 1kHz Channel 1 Input MCP Output Fast Edge from Scope 100mV and 500ps per div
4
Method TDR: Connected in manner shown on setup slide
Measured pulse height and fall time for each number of pixels Calculated C using 67% of fall time (one time constant). Assumed to be single pole.
5
16 Pixels Tied Together TDR MCP Pulse 50mV and 500ps per div
Assuming infinitely fast rise time pulse used. Fall time: ps C = pF Actual rise time of pulse used subtracted from measured. Fall time: ps C = pF LCR meter. Measured = pF Cable = 50.3 pF C = 5.32 pF MCP Pulse Rise Time 480 ps Amplitude 144.8 mV 50mV and 500ps per div 20mV and 500ps per div
6
9 Pixels Tied Together TDR MCP Pulse 50mV and 500ps per div
Assuming infinitely fast rise time pulse used. Fall time: ps C = pF Actual rise time of pulse used subtracted from measured.. Fall time: ps C = pF LCR meter. Measured = 53.8 pF Cable = 50.3 pF C = 3.5 pF MCP Pulse Rise Time 478.0 ps Amplitude 117.8 mV 50mV and 500ps per div 20mV and 500ps per div
7
4 Pixels Tied Together TDR MCP Pulse 50mV and 500ps per div
Assuming infinitely fast rise time pulse used. Fall time: ps C = pF Actual rise time of pulse used subtracted from measured. Fall time: 72.8 ps C = pF LCR meter. Measured = 51.8 pF Cable = 50.3 pF C = 1.5 pF MCP Pulse Rise Time 477.4 ps Amplitude 79.88 mV 50mV and 500ps per div 20mV and 500ps per div
8
1 Pixel TDR MCP Pulse 50mV and 500ps per div 20mV and 500ps per div
Assuming infinitely fast rise time pulse used. Fall time: ps C = pF Actual rise time of pulse used subtracted from measured. Fall time: 61.2 ps C = pF LCR meter. Measured = 53 pF Cable = 50.3 pF C = 2.7 pF MCP Pulse Rise Time 420.4 ps Amplitude 59.46 mV 50mV and 500ps per div 20mV and 500ps per div
9
Conclusion LCR meter TDR
Different cables were used on each measurement (we didn’t want to remove and reattach them) TDR Seems to show a contribution of ~0.1 pF for each pixel added.