Weighbridge Training Course Beijing - September 2017
Eccentricity Test
2
Eccentricity - Determining the load to apply
Non- Substitution method Instruments with a load receptor with four points of support or less Establish if the instrument has four or less support points. If the instrument complies with this condition then the test load is equal to one third of the maximum capacity, plus the corresponding maximum value of any additive tare . This load must be distributed in an area approximately equal to one-quarter of the surface area of the load receptor.
3
Instruments with ≤ 4 points of support (cont’d)
Large weights should be used in preference to several small weights. Smaller weights shall be placed on top of larger weights, but unnecessary stacking should be avoided within the segment to be tested. The diagram below represents the load receptor. Test loads must be spread evenly over an area corresponding to ¼ of the area of the load receptor in each position. If a single weight is used it must be applied centrally in the ¼ under test.
4
Instruments with ≤ 4 points of support (cont’d)
Zero the instrument under test. Place a test load equal to one third of the maximum at position 1 Note: If the instrument has additive tare , use Max plus maximum additive tare instead of max Determine the error ( E ) using the formula , E = I + 0.5e - ∆L – L Where : E= error , I = Indication , e = the verification scale interval , ∆ L = delta loads and L = load Taking into account the value of the test load determine if the error is within the maximum permissible error Remove the test load Repeat steps 1 to 3 in each of the four positions, each time allowing the instrument indication to become stationary (at rest) before determining the error
5
Instruments with a load receptor having more than four points of support
If the instrument has more than four points of support, establish exactly how many support points there are, as this factor determines the value of the test load and the area that the test load will be placed on the load receptor. Now that you know the number of support points, establish what the maximum capacity is Note: If the instrument has additive tare , use Max plus maximum additive tare instead of max The formulae for calculating the value of the test load, and the area in which it will be placed are:
6
Instruments with a load receptor having more than four points of support ( cont’d)
7
Instruments with a load receptor having more than four points of support ( cont’d)
Test each support point with a test load of 6 tonne spread over 1/6th of the surface area of the load receptor. The numbers represent positions of the test load on the receptor for the weighbridge with six support points. Note: Test loads must be spread evenly over an area equal to the surface area of the load receptor divided by the number of support points. In the case of our example this would be an area equal to 1/6th of the surface area. If a single weight is used it must be applied centrally in the 1/6th under test.
9
Instruments with a load receptor having more than four points of support ( cont’d)
Zero the instrument under test Place the test load at position 1 Determine the error ( E ) using the formula , E = I + 0.5e - ∆L – L Where : E= error , I = Indication , e = the verification scale interval , ∆ L = delta loads and L = load Remove the test load. Repeat steps 1 to 3 in each of the positions, each time allowing the instrument indication to become stationary (at rest) before determining the error. Taking into account the value of the test load determine if the errors are within the maximum permissible error ( refer to table 1) Note: On any weighing instrument where the load cannot be distributed as required because the load supports are too close together the test load is doubled and placed on both sides of the axis connecting the two points of support.
10
Worked Example A New weighbridge of 50000kg Maximum capacity and e = 20kg is being tested for eccentricity. The weighbridge has 6 points of support and is a Class III instrument The test load is calculated to be 10000kg Because the weighbridge has 6 points of support the load receptor is divided into 6 equal parts and the test load placed evenly in each area The results are in the table below and record the delta loads required to cause the indication to change. The formula E = I + 0.5e - ∆L – L is used to calculate the errors Position Indication + 0.5e - ∆L - load = Error 1 10000kg 10kg 8kg 2kg 2 9980kg 18kg -28kg 3 16kg -26kg 4 4kg 6kg 5 9960kg -48kg 6 1000kg 0kg
11
Worked example continued
Looking at our table of results we can see we have a range of errors depending on the position of the load on the load receptor. If we convert the load to no. of verification scale intervals we can look at the MPE table for the correct load and see what the allowable MPE is for that load. The load of 10000kg / 20kg (e ) = 500e In table 1 below the MPE allowed for an “Initial Verification” test is ± 0.5e or ± 10kg Table 1 Class of Weighing Instrument Load (expressed in the number of verification scale intervals (e)) Error (±) Initial Verification In Service Class I 0 to 0.5 1 to 2 Over 1.5 3 Class II 0 to 5 000 5 001 to Over Class III 0 to 500 501 to 2 000 Over 2 000 Class IIII 0 to 50 51 to 200 Over 200
12
Worked example Looking at our table we can see which positions failed and which positions passed MPE = + or – 0.5e or 10kg Green = pass Red = Fail Position Indication + 0.5e - ∆L - load = Error 1 10000kg 10kg 8kg 2kg 2 9980kg 18kg -28kg 3 16kg -26kg 4 4kg 6kg 5 9960kg -48kg 6 1000kg 0kg
13
Worked example In the worked example there are 3 positions that exceeded the MPE therefore the instrument fails the eccentricity test for Initial verification Would the instrument pass if it was an In Service test ?
14
Substitution method Use a suitable vehicle to move the loads. Ensure that its: Wheel track does not exceed 0.5 the width of the load receptor Wheel base length does not exceed 1/n the length of the load receptor gross weight is greater than 0.5 times and less than the nominated weight required in the non- substitution method Determine the no. of support points Divide the load receptor into n approximately equal segments where n is the number of points of support. Note each point of support and assign numbers to the segments with position 1 to the left closest to the viewing position and then label the other segments in a clockwise direction
15
Substitution method Note: For rail weighbridges . If two support points are too close together for the load to be distributed as indicated , double the load and distribute over twice the area on both sides of the axis connecting the two points of support. Determine 1/(n-1) of the Max Note: if the instrument has additive tare , use max plus maximum additive tare , instead of max. If this value is : greater than 5 t go to step 4 Less than or equal to 5 t use the non-substitution method Determine the weight required for testing. The substitution shall be: At least 0.5 of the weight determined in step 3: and No more than the weight determined in step 3. Zero the instrument Place standard weights onto the load receptor in the required position , equal to or greater than the weight of the vehicle , provided it is within 0.3 t of the vehicle weight. Record this load (L) Note: ensure that the placement of the weights does not exceed the wheel track or base dimensions of the vehicle
16
Substitution method
17
Substitution method Apply additional standard weights of 0.1 e to the load until the indicator changes up and stabilises Record this additional load (∆L) and the indication (I) Calculate the error in the weighbridge (E) for the load applied (L) using E = I + 0.5e- ∆L- L 10. Remove the standard weights and ∆L. For digital instruments , ensure that a suitable load (eg: 10e) is left on the load receptor to avoid zero tracking 11. Drive the vehicle as close as possible to the footprint of the weights in step 6. Remove the 10e placed on the load receptor in step 10. Record the indication for the substitution load (Isub) Add additional standard weights of 0.1e until the indication changes up and stabilises. Leave these additional weights (∆L) with the substitution load. Calculate the actual load (Lsub) of the substitution load using Lsub = Isub e - E Round the true value of Lsub to a whole scale interval Lsub ( rounded) by applying or removing additional standard weights , keeping the weights with the vehicle.
18
Substitution method
19
Substitution method 16. Reposition the vehicle to each point of support and record the indication 17. Determine if the instrument passes or fails . To pass , each indication for all points of support must be within 0.5e of the applied load Lsub ( rounded) Note: If the instrument fails use the non substitution method described earlier
20
Rolling load test Weighbridges used for weighing rolling loads , e.g. vehicles, are tested by using a rolling load The rolling load is applied at different positions on the load receptor. These positions will be at the beginning, the middle and at the end of the load receptor in the normal direction of travel. The test is then repeated in the reverse direction. There is no requirement restricting the area of the load receptor to be tested. The test load selected should be of the heaviest and most concentrated load that is normally weighed on the instrument under test and should be between ≥ 20% and ≤ 80% of the maximum capacity. The actual value of the rolling load must be determined by using the procedure detailed in Eccentricity Substitution method
21
Rolling load test Zero the instrument under test.
Place the test load at position 1 on the load receptor. Taking into account the value of the test load determine if the error is within the maximum permissible error Repeat the above steps in positions 2 and 3 and then in the reverse direction, positions 3, 2 and 1 in turn, each time allowing the instrument indication to become stationary (at rest) before determining the error.
22
Group Exercise Work out the loads to apply for the following Instruments; 30000kg , 4 points of support 50000kg , 8 points of support 60000kg , 8 points of support 100000kg , 12 points of support
23
Eccentricity Test Acknowledgement
To the National Measurement Institute of Australia for the section on Substitution testing