Download presentation
Presentation is loading. Please wait.
1
NAC flat fielding and intensity calibration
Flatfields Method Examples Open points Intensity calibration Definition Results
2
Flatfields Data: Data taken with integrating sphere and calibrated lamps Halogen in visible and IR Xenon in UV Data from July 2003 All raw images full frame (2096x2048) with amplifier B All images taken with cold CCD (~190 K)
3
Flatfields Method: Select 4-5 images (must pass check for bad data and shutter problems) Remove coherent noise (if necessary), subtract bias, cut off preclocked pixels Create flat field pixelwise At each pixel, discard values which deviate > 5 sigma from median at that pixel Create flat field averaging the “good” pixels Normalize so that the central 200x200 pixels average to 1
4
NAC FFP-UV/Far UV Obvious UV-pattern caused by the detector
No filter pinholes for the NAC Some vignetting
5
NAC Neutral/Blue “Humbug” pattern nearly disappeared Vignetting
Straylight (?) at the edges
6
NAC FFP-vis/Orange Extremely homogeneous (consider the scale!)
No vignetting Dust obvious
7
NAC IR/FFP-IR Not as homogeneous as in the visible
Little or no IR-fringing!
8
Open points Should we improve the statistics of the flat fields by using more input files (as done at LAS) Can improve S/N by factor (pixel-to-pixel variation calibrated down to below 0.1 % for some filters) Needs understanding of systematic effects Difference between flat fields in different lamp positions Constant flat field in space? Care is needed to avoid bad images (however, there were not many during the NAC calibration in 2003)
9
Intensity calibration
Definition of calibration factor A = C/ I A: Calibration factor [DN s-1 / W m-2 nm-1 sterad-1] C: Count rate [DN s-1] I: Irradiance of the lamp [W m-2 nm-1 sterad-1] at the central wavelength of the filter combination
10
Intensity calibration (2)
Calibration factor depends on input spectrum This would also be true if integrated irradiance would be used in the definition Knowledge about relative spectral response can in principle be used to predict calibration factor for different input spectrum Nevertheless, calibration with solar analogs is essential for accurate science measurements with OSIRIS
11
Intensity Calibration (3)
Exposure time corrected for difference between commanded and real exposure time (assumed to be 2.7 ms) Can be done more accurately (maybe) Temporal change of irradiance of the lamps was monitored with photo diodes Reliability of monitoring hard to access Occasionally diodes were not used during calibration Diode correction sometimes applied to calibration factors
12
Results (NAC) Solid lines: Prediction Symbols: Measurements
The large deviations are for the neutral density filter!! Laboratory Standard Star (ε Aqr)
13
Results Prediction and 2 measurements generally agree more or less
Complete agreement is not expected due to different input spectra However, transmissivities calculated from measurements sometimes appear high Prediction completely wrong for neutral density filter Inaccurate filter curve ? Wrong filter curve in data base?
14
Open points Error of intensity calibration hard to evaluate
Lab: Dominated by temporal changes in the output from the integrating sphere Star: Calibration may be more accurate, evaluation needs more stars Current best estimate ~5-10 % Some differences between predicted and measured transmissivity are unexplained
15
Future work Compare LAS and MPS flat fields and calibration factors
Reassess calibration procedure to understand discrepancies Analyse additional data from in flight calibration
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.