Presentation is loading. Please wait.

Presentation is loading. Please wait.

Single GEM Measurement

Similar presentations


Presentation on theme: "Single GEM Measurement"— Presentation transcript:

1 Single GEM Measurement
Matteo Alfonsi,Gabriele Croci and Bat-El Pinchasik June 11th 2008 GDD Meeting

2 Parameters Standard GEM 3 cm x 3 cm surface area
Pitch 140 µm Hole Diameter 70 µm Thickness 50 µm Standard Gaps Geometry 2 mm Induction gap 3 mm drift gap Standard Gas Ar/CO2 70%/30% 8.9 keV X-Rays

3 Picture of the GEM used and of a Standard Stock GEM
The focus used for the two pictures was different. From optical microscope measurement (by eyes) we found that the hole diameter is the same for both. We still miss the measurement with Miranda’s camera microscope (out of order).

4 Pulse Height Spectrum The rate used in the following current measurements is higher, around a value of 250 kHz/mm2 The Gain is evaluated using (Rate Method)

5 Drift Field = 1.9 kV/cm; Induction Field = 3 kV/cm
Voltage Scan Drift Field = 1.9 kV/cm; Induction Field = 3 kV/cm

6 GEM Voltage = 500 V; Induction Field = 3 kV/cm
Drift Scan GEM Voltage = 500 V; Induction Field = 3 kV/cm

7 Drift Field = 1.9 kV/cm; GEM Voltage = 500 V
Induction Scan Drift Field = 1.9 kV/cm; GEM Voltage = 500 V

8 Literature comparison
J. Benlloch et al. /Nucl. Instr. and Meth. in Phys. Res. A 419 (1998) 410) S Bachmann et al. Nucl. Instrum. Methods Phys. Res., A, 438(CERN-EP ):376– p, Apr 1999.

9 Simulation Comparison
Simulations ????? ?????

10 An other method to measure Effective Gain
We tried to measure the Gain without measuring the rate (as in the usual method) but measuring the ionization current We used this setup to measure the ionization current We fixed the rate to a value where we can read the ionization current 8.9 keV X-Rays Vdrift = 0 Edrift VTopGEM < 0 EGEM VBottomGEM = 0 Vanode= 0 We read the ionization electron current on the drift electrode using Keithley 6517A electrometer

11 Ionization Current Measurement
Top GEM Current (when measurable) No X-Rays Ionization Current on Drift Electrode We found a mean value for the ionization current around 150 pA

12 GAIN Measurement: Two Methods Comparison
Rate Method Ed = 1.9 kV/cm ; Ei = 3 kV/cm Ionization Method Ed = 1.7 kV/cm ; Ei = 3 kV/cm It seems that the two results are compatible

13 New detectors Coming.. Standard Bulk Micromegas THGEM “Closed”
Geometry: Amplification gap = 128 µm THGEM “Closed”

14 THGEM Closed THGEM Copper FiberGlass BackPlane
Conductive Glue Copper FiberGlass BackPlane THGEM (geometry pitch 0.7 mm, hole diameter 0.3 mm, thickness 0.6) closed on one side. Glue a kapton foil on the bottom of the THGEM with conductive glue. Glue a copper readout electrode to this kapton foil (anode) and then glue a fiberglass support to this anode.

15 Testing Standard Bulk Micromegas in air (with Paul Colas)
Vmesh (V) I (nA) 200 2 300 400 2.1 450 2.2 500 2.3 600 2.7 650 2.75 700 2.95 800 3.1 830 First Spark, than stable 870 Last Stable Value before having many sparks Mesh Photos The gas that will be used to test this detector (under Paul’s suggestion) is Ar/CF4/Iso 95%/3%/2% and the maximum voltage that will be applied to the mesh (being Vanode =0) is 360V Vanode = 0


Download ppt "Single GEM Measurement"

Similar presentations


Ads by Google