The Determination of 226Ra in Water Samples
eichrom TECHNOLOGIES INC The Determination of 226Ra in Water Samples Anil H. Thakkar & Michael Fern Eichrom Technologies, Inc. David McCurdy Duke Engineering & Services Basically this piece of work was carried for our potential client Quanterra, formerly known as IT Corp. A protocol was proposed to Quanterra based on their needs to carry out day to day analyses in their lab.
2
eichrom TECHNOLOGIES INC Objectives To design a procedure for radium in water which would: be a simple, short and quick procedure allow to measure Ra-226 and Ra-224 via alpha spectroscopy and Ra-228 via Ac-228 measurements eliminate the tedious steps of precipitation and metathesis An improved procedure was developed to separate Am,Pu and U in variety of water matrices.
3
Current Procedure Used
eichrom TECHNOLOGIES INC Current Procedure Used Barium sulfate precipitation Conversion to BaCO3- tedious and time consuming
4
Current Procedure Used
eichrom TECHNOLOGIES INC Current Procedure Used Sample + Ba carrier + Ba Conc.H2SO4_+ Heat BaSO4 Precipitate allowed to settle for several hours or overnight Precipitate collected and converted to BaCO3
5
eichrom New Approach Water samples Cation exchange resin column
TECHNOLOGIES INC New Approach Water samples Cation exchange resin column Eichrom Ln Resin column
6
Preparation of Water Samples
eichrom TECHNOLOGIES INC Preparation of Water Samples 0.5 L / 1.0 L water samples, acidified to pH 2 Add 133Ba as a tracer
7
Eichrom Cation Exchange Column
TECHNOLOGIES INC Eichrom Cation Exchange Column Load solution: 0.5 L/ 1.0 L of sample Rinse with 50 mL of 0.1M HNO3 Strip Ra and Ba with 50 mL of 8M HNO3
8
Preparation of Load Solution for Ln Resin Column
eichrom TECHNOLOGIES INC Preparation of Load Solution for Ln Resin Column Evaporate eluent from cation exchange column to dryness Dissolve residue in 10 ml of 0.095M HNO3 acid Load on Eichrom Ln Resin column
9
Ln Resin Column Separation
eichrom TECHNOLOGIES INC Ln Resin Column Separation (1) Load: 10 mL of mL 0.095M HNO3, Collect (1) and (2) for Ba & 226Ra (2) Rinse: 15 mL of 0.095M HNO3 On Uteva spec, the actinides were loaded with a load solution of 3M HNO3/0.5 M Al(NO3)3 solution, Pu was reduce by Iron. Am/Pu pass thru, Uteva, therefore the load solution and rinse are collected, Uranium is stripped by 0.01 M HCl . (3) Strip: 10 mL of 0.35M HNO3 collect (3) for 228Ac
10
226Ra Via Alpha Spectrometry
eichrom TECHNOLOGIES INC 226Ra Via Alpha Spectrometry Evaporate 226Ra/133Ba fraction to 10 mL volume Add 0.1 ml of 75 mg/ml barium carrier Add 3 mL of 40% sodium sulfate, 4 drops of 1:1 acetic acid. Swirl Add 0.2 mL of seeding suspension. Mix Place tubes in cold water bath for 30 minutes Filter on 0.1 micron filters, count via alpha spectrometry
11
eichrom 133Ba Via Gamma Counter
TECHNOLOGIES INC 133Ba Via Gamma Counter Count 133Ba tracer yield via gamma spectrometry
12
228Ac Via Gas-Flow Proportional Counter
eichrom TECHNOLOGIES INC 228Ac Via Gas-Flow Proportional Counter Add 50 micrograms of cerium carrier Add 1.0 mL of HF Filter on 0.1 micron filters Measure for beta radiation
13
226Ra % Recoveries-0.5L DI Water
eichrom TECHNOLOGIES INC 226Ra % Recoveries-0.5L DI Water
14
eichrom TECHNOLOGIES INC 226Ra Alpha Spectrum
15
eichrom TECHNOLOGIES INC 133Ba Gamma Spectrum
16
eichrom 226Ra, 224Ra, and Daughters 224Ra 226 Ra 220Rn 218Po 222Rn
TECHNOLOGIES INC 226Ra, 224Ra, and Daughters 224Ra 226 Ra 220Rn 218Po 222Rn
17
226Ra % Recoveries in Tap water
eichrom TECHNOLOGIES INC 226Ra % Recoveries in Tap water
18
eichrom TECHNOLOGIES INC 226Ra in EPA Sample Sept.18, 1998
19
eichrom TECHNOLOGIES INC 226Ra in NJ Sample
20
226Ra in Georgia Inst. Of Tech. Sample# S8933
eichrom TECHNOLOGIES INC 226Ra in Georgia Inst. Of Tech. Sample# S8933
21
226Ra in Georgia Inst. Of Tech. Sample# WS 14776
eichrom TECHNOLOGIES INC 226Ra in Georgia Inst. Of Tech. Sample# WS 14776
22
eichrom Summary Cation exchange resin option:
TECHNOLOGIES INC Summary Cation exchange resin option: provides a good replacement for the time consuming BaSO4 precipitation steps it eliminates inconsistent metathesis steps reduces usage of chemicals/reagents provides consistency and saves time Acceptable 226Ra recoveries
23
Future Work To evaluate the effect of Ba/Ra uptake in high TDS samples especially with high Ca and Mg content To measure Ra-228 concentrations in samples
24
Acknowledgements Special thanks to
Dr. B. Parsa, New Jersey department of health, NJ Dr. B. Kahn, Georgia Institute of Technology, GA