Download presentation
Presentation is loading. Please wait.
1
Micros 60 & Quality Control
MLAB 1415: Hematology Micros 60 & Quality Control
2
CBC Quality Control Commercial Controls: 3 levels (low, normal, high)
Values stored in instrument computer Levey-Jennings graph generated and stored for each parameter Mode to Mode QC: Most automated hematology instruments have a primary and secondary mode of sample aspiration. Controls must be run on BOTH and correlate. Primary=Automated or Closed Secondary=Manual or Open Delta Checks When the Laboratory Information System (LIS) and the instrument are interfaced (connected) delta checks are conducted by the LIS on select parameters. Current values compared to most previous result Differences greater than the limits set within the LIS are flagged
3
Micros 60 3-part differential Sample size= 10µL Environment
Indoor only Level work station Avoid direct exposure to sunlight Proper ventilation 60-90oF Linearity (AMR) WBC (103/mm3): 0-100 RBC (106/mm3): 0-8.0 Hgb (g/dL): Hct (%): Plt (103/mm3):
4
Linearity (Reportable Range/ AMR)
Instruments are calibrated for each analyte with a range that is clinically relevant. With concentrations above or below the reportable range or Linearity range, the result does not correspond with the calibration curve in linear fashion. Results outside of linearity are NOT acceptable. Linearity ranges vary by instrument. Example: Parameter Coulter STKS Advia WBC X 103 /µL X 103 /µL
5
Parameters Parameters Measurements/Calculations WBC/RBC/PLT
Hgb Hct MCV MCH MCHC RDW Platelet MPV Lym # Lym% Mono # Mono% Gra# Gra% WBC/RBC/PLT Impedance change HGB Spectrophotometry LYM%/MON%/GRA% Calculations MCV, MCH,MCHC,RDW,MPV,LYM#, MON#, GRA#
6
Automation Specs Principle of operation: The counting principle used for the red blood cells, white blood cells and platelets is based on the variations in impedance generated by the passage of cells through a calibrated aperture. Number of samples which can be analyzed per hour: 60 Number of tests performed on this analyzer: 1 Quality control: Three levels run each day of patient testing, or following instrument repair, reagent lot change, or when indicated. Calibration: Shift in control values, replacement of instrument parts, reagent number change, or when indicated Preventative maintenance: Daily: start up and shut down, general cleaning Acceptable samples: whole blood from an EDTA tube ran within 4 hours of collection Limitations: Lipemia, cold agglutinins, high WBC’s
7
Equipment 8 Main parts Electrical supply Electronic board
Dilution pneumatics Control panel with keyboard and LCD screen Reagent compartment Printer Smart card reader Bar code reader (optional)
8
Reagents Waste Diluent Cleaner LYSE: potassium cyanide
ABX minidil: sodium hydroxide/azide Determines and differenciates blood cells Cleaner ABX miniclean: enzymatic detergent Cleans system LYSE: potassium cyanide ABX Alphalyse: Lyses RBCs, releasing hemoglobin and WBCs. WBCs are counted by impedance. Waste
9
Distribution curves WBC RBC Plt
10
Daily Start Up Turn on instrument, once on, the start-up will begin automatically Instrument is performing a cleaning and rinsing cycle: detergent in the chamber is rinsed with diluent Cycle lasts about 130 seconds At cycle completion, background values will be displayed. These values must be less than the below: WBC < 0.3 x 103/mm3 RBC < 0.02 x 106/mm3 Hgb < 0.0 g/dL Plt < 10x 103/ mm3 If the values are not within this range, another cycle will begin. If still not in range, the Micros will do another start-up. After 3 failed cycles, it will give a “start-up failed” message. Troubleshooting will now start.
11
Daily start-up con’t Run a primer sample Perform QC and evaluate
Complete maintenance sheet If all QC in, can continue with patient samples.
12
Running a Patient Press the ID button on the key pad
Manually enter the ID. Use the up and down arrows to enter alphabetical characters. Please Enter after each entry.
13
Shutdown Hit the Stand by key, wait 1 minute
Switch off instrument and printer. When the shutdown hit is pressed, the Micros 60 performs a complete cleaning with the Miniclean and puts the system into a Standby mode. Once this process is complete, the instrument can be switched off.
Similar presentations
© 2024 SlidePlayer.com. Inc.
All rights reserved.