Presentation is loading. Please wait.

Presentation is loading. Please wait.

Protection of Patients

Similar presentations


Presentation on theme: "Protection of Patients"— Presentation transcript:

1 Protection of Patients
Stacy Kopso, M.Ed. RT (R)(M)

2 Protection of patients
Immobilization Beam Restriction Beam-limiting devices Filtration Gonadal shielding Exposure and technique factors Film Screen Electronic Imaging Patient positioning Grids Pregnant patient Repeat Radiographs Fluoroscopic Procedures

3 Protection of patients
Immobilization Communication Good directions Discomfort Voluntary motion Immobilization devices Involuntary motion heart movement/bowel peristalsis short exposure times

4 Beam Restriction Beam limiting devices
Restrict the primary beam to the specfici area of interest Reduce the amount of tissue exposed to the radiation beam Small field size Mindful of kVp Beam limiting devices Aperture diaphragms Flat lead containing hole in center Attached to tube head angio Cones/Cylinders Circular metal structures Attach to tube housing Predetermined circular field size Straight or flared Longest cone, smallest diameter Sinuses Designed to operate most effectively at a designated SID

5 Beam limiting devices Variable aperture collimators
Manually adjustable beam restricting device Upper and lower pair of shutters Upper absorbs off-focus radiation inside tube Lower located below lamp and mirror Restrict the beam to area of interest Right angles to each other Each pair move independently Square or rectangular field size Lamp,mirror and crosshair (illuminate and deflect light) Positive beam limitation (PBL) Electronically interlocked with bucky tray. Automatically opens the radiation field size to the size of the image receptor Accuracy within 2% of SID 4 sided border

6 Beam filtration Beam Filtration Primary purpose
Reduce patient exposure/reduces entrance skin exposure Aluminum filter inserted in the path of the xray beam Absorbs low energy photons(non diagnostic) Improves beam quality Increases beam energy As filtration increases, exposure rate decreases

7

8 Beam filtration Inherent filtration Added filtration Total filtration
Glass window of tube and cooling oil surrounding tube housing Increases as tube ages Between mm Al Added filtration Aluminum Located outside the glass window Placed in the path of the primary beam to bring the total filtration amount up to the minimum standards Total filtration Combination of inherent and added

9 Beam filtration Operating kVp Minimum total filtration Below 50
.5 mm Al 50-70 1.5 mm Al Above 70 2.5 mm Al Mobile/fluoro unit 2.5mm Al

10 Beam Filtration Half value layer (HVL)
Used to determine whether a diagnostic x-ray beam is adequately filtered The thickness of material that will reduce the x-ray intensity to half its original value Measured by QC personnel on an annual basis or after repairs have been performed on tube housing or collimator An increase in kVp requires an increase in HVL Measures beam quality Example; Intensity of 100 mR. If one HVL is added the new intensity would be 50 mR If inherent filtration is 1.0 how much added filtration is needed if tube is operating at above 70 kVp: 1.5mm Al

11 Gonadal shielding Used when the gonads are in or within 5cm of primary beam Used on anyone who is within reproductive age Under the patient for fluoro procedures, over the patient for overheads Types Flat contact Recumbent studies Shadow Suspended from tube housing Cast a shadow over area Sterile fields Shaped Male patients, held in place by disposable supporter Oblique lateral and erect studies Fluor studies rare Lead apron

12 Exposure factors kVp factors to reduce patient dose mAs
Increase kVp/lower entrance skin exposure Less absorption/photoelectric effect Decrease mAs/pt dose Disadvantage Poor contrast mAs “Rate of exposure” Direct effect on quantity of radiation Decrease mAs “Patient dose” 3 phase and high frequency generators Lower entrance skin exposure

13 Film screen considerations
High speed film and intensifying screen Rare earth screens More efficient than calcium tungstate Absorb 3x more Screens accelerate the action of xrays by converting energy into visible light

14 Electronic imaging Computed radiography (CR) & Digital Radiography (DR) Eliminates need for repeat due to technical factors because of the large dynamic range

15 PATIENT POSITIONING PA vs AP T-spine (PA) Scoliosis series (PA)
breast Scoliosis series (PA) Rib (PA) Skull/sinuses (PA) Lens of eye Cervical spine (RAO and LAO) thyroid

16 grids An increase in grid ratio, increase patient dose
Non grid to grid 70 kvp or 10cm Air gap technique long SID/increased OID

17 Pregnant patient Elective abdomen exams should be postponed until 10 days following the onset of menstruation Physicist should perform calcuations to estimate the actual fetal dose

18 Repeat radiographs Positioning Pt motion
Incorrect radiographic techniques Dirty screens Incorrectly loaded cassettes Light leaks Chemical fog Processor artifacts Incorrect projections Positioning errors Grid errors Double exposure

19 Fluoroscopic procedures
Greatest patient and personnel radiation exposure Intermittent/pulsed fluro Digital form of fluoroscopy Decreases pt exposure Stores image from last time fluoro switch was depressed Limitation field size (collimate) Decrease total time Decrease image intensifier to patient distance will decrease entrance exposure Automatically adjusts mA (exposure) 2.5mm Al Source-tabletop/Fixed 15 inches Mobile 12 inches Cumulative timing device Five minutes

20 Exposure rate no greater than 10R/min for intensified fluoro units
Cannot exceed 2.1R/minute for each mA of operation at 80 kVp Non digital fluoro units operate within a range of 2 to 5 mA If 4 mA is utilized at 80 kVp, what is the exposure if the fluoroscopy time is 2.5 minutes? 21R 2.1 R x 4mA=8.4 8.4 x 2.5 minutes=21R


Download ppt "Protection of Patients"

Similar presentations


Ads by Google