Presentation is loading. Please wait.

Presentation is loading. Please wait.

The Heidelberg Dark Matter Search Experiment

Similar presentations


Presentation on theme: "The Heidelberg Dark Matter Search Experiment"— Presentation transcript:

1 The Heidelberg Dark Matter Search Experiment
Laura Baudis Stanford University

2 The HDMS collaboration
MPI fuer Kernphysik, Heidelberg Group of Prof. H.V. Klapdor-Kleingrothaus

3 The HDMS detectors WIMP 2 HPGe detectors in one cryostat system
inner: WIMP target outer: active veto against multiple Comptons passive Ge shield gamma

4 Some detector properties
Property Inner Detector Outer Detector Crystal type p-type n-type Mass [g] 202 2111 Active vol [cc] 37 383 Diameter [mm] 35.2 84.4 Length [mm] 40.3 86.2 Op. bias [V] + 2500 -1500 300 keV (1332 keV) [keV] 1.2 (1.87) 3.2 (4.45) Threshold [keV] 2.0 7.5

5 Cross talk between detectors
linear correlation: correction offline kio = /- 0.43 x 10-4 koi = /- 0.27 x 10-3

6 Cross talk induced error
compare zero energy resolutions: inner: no effect outer: from 0.3% to 0.4% at 1332 keV

7 The HDMS detector

8 HDMS in Gran Sasso prototype: March 98-July 99

9 Performance @ GS stable running over ~ 15 months
individual runs ~ 0.9 d checked daily: leakage current nitrogen flow overall trigger individual count rate, etc weekly: calibration with 133Ba, 152Eu-232Th source

10 Outer detector: 363 raw live-d
68Ge, 57Co, 58Co, 54Mn, 60Co, 65Zn, 137Cs, 238U, 232Th, 40K

11 Inner detector: 363 raw live-d
68Ge, 54Mn, 57Co, 60Co, 210Pb, 212Pb, 214Pb, 40K

12 Background model MC simulation of dominant backgrd.
sources with extended GEANT3.21

13 Inner: low energy spectrum
68Ga X-ray 49Ti, 55Mn cross talk?

14 Structure at 32 keV: cross talk?

15 Anticoincidence between inner and outer
Eth for anticoincidence: 3 sigma away from zero threshold [keV] FWHM/2.35 inner 2.0 0.40 outer 7.5 1.42 3s 1.2 keV 4.3 keV keV: factor 4.3 11-40 keV: 0.2 events/kg d keV keV: 0.07 events/kg d keV

16 HDMS prototype limits only last 49 d of data taking: 9.9 kg d
2-30 keV: 0.5 events/kg y keV L. Baudis et al., PRD63,2000

17 73Ge-HDMS: full scale inner crystal replaced by 73Ge crystal
sensitive to spin-dependent interactions 70Ge de-enriched by factor >50 new copper holder system no soldering of contacts installed in GS: August 2000 working well first analysis: fall 2001? expectation: test DAMA with independent technique and raw data

18 HDMS goal Based on: 0.07 events/kg d keV in keV region

19 Conclusions HDMS prototype worked well
limits competitive to other Ge experiments 73Ge-HDMS installed in August 2000 first 73Ge detector! first data to be analyzed soon... independent test of DAMA region sensitive to spin-dependent interaction however, increase of target mass improbable no sensitivity to a WIMP signature


Download ppt "The Heidelberg Dark Matter Search Experiment"

Similar presentations


Ads by Google