Effect of Open fittings
May, 2004 Johnny H. Hansen Director, Global Technical Operations GN ReSound as.
2
What’s the problem? M&M Hearing Uppsala Sweden – Experienced problems 6 months after fitting – Based on 1361 responses
3
It’s all about occlusion…
4
Occlusion 1k 8k 2k 4k 500 250 Frequency (Hz) dB SPL in ear canal CIC with pressure vent BTE with 1.5 mm vent Open mold Sound pressure measured in ear for own vocalization of vowel /ee/
5
…..and feedback
6
The dilemma.. and the solution
Effect of Stabilizer DFS
7
Why ResoundAIR and Canta7/Open are so revolutionary?
Because we have combined: state-of-art technology in earmould design/venting signal processing in ComforTech .
8
ComforTech: WarpOpen
9
Stabilizer™ DFS Technology
Feedback path S WarpOpen Compression system Static filter bank Adaptation Adaptive filter - The feedback path is measured and stored in the static filter bank which is a “mirror” of the actual acoustic feedback path
10
The proof….
11
Hearing threshold level (dB HL)
The occlusion research project in Giessen Germany… 20 40 60 80 100 120 125 1k Frequency (Hz) Hearing threshold level (dB HL) 250 500 2k 4k 8k AC HTL BC HTL NH group (n = 9) HI group (n = 10)
12
FlexVent and ResoundAIR Ear Tip
(front) FlexVent (rear) ResoundAIR ear tip
13
Occlusion measure (dB)
Measured occlusion: SPL increase in ear canal at low frequencies when reading a standard text -5 5 10 15 20 Occlusion measure (dB) Test Retest Flex- Vent 1.6 mm Earmold vent 2.4 mm ReSound AIR ear tip Un- occluded all subjects (n = 19)
14
Experienced occlusion:
NH group (n=9) -2 2 4 6 8 10 Occlusion rating (arb. units) Flex- Vent 1.6 mm Earmold vent 2.4 mm ReSound AIR ear tip Un- occluded Test Retest HI group (n = 10) -2 2 4 6 8 10 Occlusion rating (arb. units) Flex- Vent 1.6 mm Earmold vent 2.4 mm ReSound AIR ear tip Un- occluded Test Retest
15
The challenge: Open fitting in custom (CIC) products
16
Stepped vent in CIC device
Here you see an example of how the stepped vent looks like in a CIC device.
17
Feedback path response
10 2 3 4 -80 -70 -60 -50 -40 -30 -20 -10 Frequency, Hz Feedback path response, dB Stepped Not stepped Stepping makes the vent shorter and shifts the peak of the high frequency resonance, reducing the likelihood of feedback and optimizes Stabilizer function. A vent produces a low frequency peak at the cut off frequency which is evident on a coupler measurement and is largely damped in a real ear. This resonance produces gain (insertion gain) and reduces the amount of attenuation for the frequencies in the region of the peak. That is, the attenuation slope is not a straight line. The stepped vent have a significant effect in reducing the risk of feedback in the high frequencies by shifting the high frequency resonant vent peak up in frequency - beyond the useful bandwidth of the aid. The effect of the Stabilizer is thereby increased
18
The Challenge..
19
The Challenge…. Stepping optimize the work conditions for Stabilizer-DFS but don’t create significant occlusion reduction To obtain full Open comfort benefits in CIC’s we need very large vent diameter combining new skills and mindset challenge with dispenser, the sales force and the Lab More in the next sessions……….
20
Vents….
21
How do Vents work ? Sound bore Dampers Vents
Vents affect the low frequencies Vents Dampers Sound bore (Dillon, 2001)
22
What determines the vent effectiveness
Acoustic Mass determines the vent effect The lower the Acoustic Mass the higher the vent effect (occlusion relief) Formula Vent Length x Constant Vent diameter2
23
Effects of Ventilation
Low frequency sounds can enter and leave the ear canal through the vent. Transmission of low frequency sounds is easier the smaller the acoustic mass of the vent. The acoustic mass of the vent is calculated from the length & diameter of the vent. Increasing the diameter reduces the acoustic mass Decreasing the length reduces the acoustic mass
24
Openness of vents Non-occluding Non-occluding Open Closed Closed
25
How open are our instruments
ReSoundAir rating 1 FlexV (L) rating 3 FlexV (M) rating 4 Tmold (2.4) rating 6 Tmold (1.6) rating 7 S.CIC (2.3) rating 5 T.CIC (2.3) rating 5 FV(M):4 FV(L):3 Air:1 IROS SCIC:5 Here you see the subjective occlusion rating (HI subjects) from the Giessen trial. Explain why the stepped vent doesn’t alleviate occlusion. Venten kortes af, men kun ud til et hulrum som ikke er åbent, idet resten af apparatet fylder en del af øregangen ud. Vent length in custom products: 19 x 2 (right and left) CIC devices vent length (stepped vent): average of 1,48 cm 19 x 2 (right and left) CIC devices vent lengh (traditionel vent): average of 1,99 cm. Vent length of tradional earmold vents: 6 Earmolds 2,4 mm:average of 1,9 cm (vent length) 6 Earmolds 1,6 mm: average of 1,8 cm (vent length). TCIC:5 Tmold:7 Tmold:6
26
Sound Demo´s Speak Crunching
27
Custom Products
28
Problems in the Lab Custom products with open fittings, especially CIC’s often cause trouble for the Lab. To get room for the vent canal you occasionally have to compromise. You can focus on either Cosmetics or Comfort. You can compromise on either vent diameter or instrument size.
29
Full cosmetic benefits
Custom Open Products Vent diameter Small Large Full cosmetic benefits Small Select cases* * Cases of larger ear canals Instrument size Full comfort benefits No perceived benefits Large Ø 1.6 mm (~2.0 mm2) Ø 2.6 mm Ø 3.0 mm (~7.0 mm2) <-Closed Open ->
30
Vents on Custom Products
Non-occluding Non-occluding Open Custom Closed Closed
31
The conscious choice Compromise on instrument size.
Risk of Return due to protrusion Comfort Limited room for vent Cosmetics Compromise on vent diameter. Risk of Return due to occlusion. Keep in mind that 40% of users experience Occlusion as the primary problem
32
Venting Process 1
33
Large vent issues Faceplate Module Battery Hybrid Receiver
34
Large vent issues Faceplate Module Battery Hybrid Receiver
35
Large vent issues Faceplate Module Battery Hybrid Receiver
36
Lab Training America Asia Europe North Europe South
4 Lab training sessions to be co-ordinated with Product/Mkt/Sales training America (Bloomington) March 3rd US – CND Asia (DanaJapan) March 9th JPN – CHI – AUS - NZ Europe North (Taastrup) March 9th GER – HOL – FRA DEN - SWE – NOR - BRA Europe South (Madrid) March 12th SPA – ITA – U.K. BEL - SWI - A
37
Thank you. Questions ?