Download presentation
Presentation is loading. Please wait.
1
ISOCAM Parallel Survey
The 6.7m Point-Source Catalogue Duilio Farina Supervisor: Stephan Ott September 8, 2008
2
ISO ISO Instruments: ISOPHOT, SWS, LWS and ISOCAM
ISOCAM: a camera consisted in a SW (2.5-5m) detector and a LW (4-18m) detector of 32x32 pixel Prime Survey was performed when ISOCAM was the prime instrument Parallel Survey was performed in the so-called Paralled Mode when one of the other three instruments observed a target
3
ISOCAM Parallel Survey
Configuration: LW2: 6.7m PFOV: 6’’ Time: 3971 hours 24036 pointings
4
Difference between Prime and Parallel
Reduced telemetry rate, the 50% of the Parallel pointings have less than 10 images. For the majority of Parallel-mode data there is no spatial redudancy. Sources are distributed all over the detector PRIME In Prime there are typically more than 12 images per pointing Prime surveys were performed in raster observations The target was generally centered at the middle of the detector Prime observations were quality checked by ISOCAM Instrument Dedicated Team
5
1. Type Classification A list of 207 sources observed in Prime Mode and in Parallel Mode Type Classification: Point Many Extended Cloud
6
2. Match Example of Bad Match Example of Good Match
7
3. Photometry Manual Photometry was performed for all 207 sources in Prime and in Parallel Total: 414 sources with manual photometry
8
1. Results: Auto Vs Manual
Relative Error: Only Point Sources Only sources with good match Only sources far from the edge and the dark column at least 1 pixel Remaining sources: 95
9
PRIME PARALLEL Total: 95 Rel Err Mean: 0.030 Rel Err RMS: 0.264
10
2. Results: Prime and Parallel with same configuration
Sources viewed with the same setting in Prime and in Parallel (LW2 and PFOV 6’’) have to give the same flux We selected: Point sources Good match Far from the edge and dark column at least 1 pixel Num Mean RMS XPHOT 16 0.011 0.140 AUTO -0.032 0.170
11
Conclusion Good agreement between Prime flux and Parallel flux with the same settings Good agreement between Auto flux and Manual flux in both modes Parallel Mode is consistent with the Prime Mode
12
Comparison between IPSC and other catalogues
The ISOCAM Point Source Catalogue consist of individual sources with a flux range between 0.48 mJy and 1960 mJy with a median of 3.84 mJy Comparison with bands: J, H, K (2MASS), B and V (other catalogues) using the Kurucz’s model We selected 2041 sources: No indication of source confusion Number of frame > 10 No other ISO sources in a radius of 25’’ No other 2MASS sources in a radius of 25’’ The five fluxes (B, V, J, H, K) and the filter type are available for 224 of the 2041 sources
13
An example: J H K V Cam Par B
14
Results Relative Error: Tot: 224 Mean: -0.07 RMS: 0.16
On average the ISOCAM fluxes are lower of about the 7% compared to the Kurucz interpolation
15
Comparison between IPSC and Spitzer
In Spitzer catalogue there are 2618 matches with ISOCAM Point Source Catalogue We considered only sources with a match in a radius < 2’’ and with good quality flags in Spitzer Remaining Sources : 1395 2’’
16
An example IR1 K IR2 IR3 IR4 H Cam Par J
17
Results Tot: 1395 Mean: -0.020 RMS: 0.28
Positive outliers due to the better Spitzer’s spatial resolution
18
With cut: Mean: -0.5 RMS: On average the ISOCAM fluxes are lower of about the 5% compared to the Spitzer interpolation
19
Future Complete the comparison with Spitzer:
Interpolation with the Kurucz’s model Delete the multiple matches of Spitzer Investigate the possibles reasons of the low fluxes of ISOCAM Check if there is this negative shift also in the Prime observations Continue to perform the manual photometry: For the moment we have about 2000 sources with manual photometry
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.