Download presentation
Presentation is loading. Please wait.
Published byBarbara Williams Modified over 6 years ago
1
A novel gaseous X-ray polarimeter: data analysis and simulation
Astronomical telescopes and instrumentation, Waikoloa 2002 A novel gaseous X-ray polarimeter: data analysis and simulation A research by: R. Bellazzini, E. Costa, F. Angelini, L. Baldini, A. Brez, G. di Persio, L. Latronico, M. M. Massai, N. Omodei, L. Pacciani, P. Soffitta, G. Spandre. Waikoloa, 28 August 2002 Luca Baldini
2
The Micro Pattern Gas Detector
The MPGD exploit the dependence of photoemission direction on photon polarization. The photoelectron track is reconstructed by the means of a finely segmented readout plane. Waikoloa, 28 August 2002 Luca Baldini
3
Some events… The initial part of the track, with a low ionization density, evolves into a clear Bragg peak, while the photoelectron direction is randomized by Coulomb scattering (real events, 5.9 keV unpolarized radiation from 55Fe source). Large-angle scattering Auger electron Bragg peak Waikoloa, 28 August 2002 Luca Baldini
4
Basic reconstruction algorithm I
The basic algorithm for the reconstruction of the photoemission direction is based on the determination of the principal axes of charge distribution. Major axis Minor axis Waikoloa, 28 August 2002 Luca Baldini
5
Basic reconstruction algorithm II
The large average number of fired pixel per event allows a good track reconstruction. Most clusters are sensibly far from a spherical shape (M2max/M2min ~ 1), which is crucial for angular reconstruction. (Real data, 5.9 keV unpolarized radiation from 55Fe source). Waikoloa, 28 August 2002 Luca Baldini
6
Basic reconstruction algorithm III
5.9 keV unpolarized radiation 5.4 keV 100% linearly polarized radiation Cuts on the shape can be applied in order to optimize the sensitivity of the instrument: Waikoloa, 28 August 2002 Luca Baldini
7
Absorption point reconstruction I
The initial part of the track is characterized by a lower ionization density and this asymmetry can be exploited to reconstruct the conversion point. dE/dx x Principal axis xb Reconstructed conversion point ~ (M2max)1/2 Barycentre Waikoloa, 28 August 2002 Luca Baldini
8
Absorption point reconstruction II
5.4 keV unpolarized radiation entering the detector through a 50 mm collimator. 1D projection Standard deviation of gaussian fit: Barycentres Reconstructed absorption points Waikoloa, 28 August 2002 Luca Baldini
9
Absorption point reconstruction III
500 mm 1 mm Waikoloa, 28 August 2002 Luca Baldini
10
Absorption point reconstruction IV
Position of barycentres with respect to the reconstructed conversion point. 5.9 keV unpolarized radiation 5.4 keV 100% linearly polarized radiation Waikoloa, 28 August 2002 Luca Baldini
11
Angular reconstruction I
The reconstruction of the absorption point can be exploited to improve angular accuracy, rejecting the final part of the track, which is blurred by Coulomb scattering (real events, 8.0 keV polarized radiation). Basic algorithm Improved algorithm Waikoloa, 28 August 2002 Luca Baldini
12
Angular reconstruction II
8.0 keV 100% polarized radiation, basic reconstruction algorithm. Same events, analyzed exploiting reconstructed conversion point. Modulation factor rises up from 24% to 30%. Waikoloa, 28 August 2002 Luca Baldini
13
Monte Carlo simulation I
Transverse diffusion toward the GEM (5 mm in Ne). Simulation of primary ionization distribution. Auger electron Bragg peak Sampling onto readout plane (100 mm pitch). Waikoloa, 28 August 2002 Luca Baldini
14
Monte Carlo simulation II
5.0 keV photoelectrons tracks in Ne (100% linearly polarized, collimated photons beam). Waikoloa, 28 August 2002 Luca Baldini
15
Monte Carlo simulation III
5.0 keV photoelectrons tracks in Ne (100% linearly polarized, collimated photons beam). Modulation factor, as evaluated from charge released within a certain distance from conversion point. Waikoloa, 28 August 2002 Luca Baldini
16
Tested prototype simulation
Modulation factor as a function of photon energy (exploiting reconstructed conversion point). Modulation factor as a function of photon energy (basic algorithm). Experimental data Experimental data Monte Carlo simulation Monte Carlo simulation Waikoloa, 28 August 2002 Luca Baldini
17
100 mm pitch detector simulation
Modulation factor as a function of photon energy for several absorption gap thickness (100 mm readout pitch). Complete energy scan for 1 cm absorption gap. Waikoloa, 28 August 2002 Luca Baldini
18
Conclusions Some data collected during tests presented.
Algorithms used throughout analysis described in details. Monte Carlo simulation in good agreement with experimental data. Still room for great improvements. Waikoloa, 28 August 2002 Luca Baldini
Similar presentations
© 2024 SlidePlayer.com. Inc.