1 GERDA meeting Dubna 27-29/06/2005
Natural Ge crystal contamination measurement and Isotopic ratio determination by ICP-MS LNGS Chemistry Service S. Nisi

2 Analysis method Ge crystal dissolution HNO3+HF+H2O 1:1:1
Dilution of samples and spiked samples preparation to check the procedure Measurement of contamination and isotopic ratio by ICP-MS Agilent Technologies model 7500a (quadrupole mass analyzer)

3 ICP-MS in the clean room at LNGS

4 Interference on contamination measurement
Element Interference 11B 10B1H, 22Ne++ N Air 27Al 31P 15N16O, 30Si1H 69Ga 71Ga 70Ge1H 75As 74Ge1H 113In 115In 73Ge40Ar 121Sb 205Tl 209Bi 232Th 238U

5 Ge93 13196 contamination results
Ge Spike Element Ge ppm Spike conc Meas. Conc 11B <0,01 27Al 0,27 0,2 0,32 31P <2 16,5 39K <5 42 34 69Ga 0,034 0,032 115In 0,014 0,1 121Sb 0,001 0,01 0,025 205Tl <0,0021 0,0089 209Bi <0,0019 <0,0018 232Th <0,0016 238U <0,0015 0,0092

6 Ge124 11888 contamination results
Ge Spike Element Ge ppm Spike conc ppm Meas. Conc ppm 11B <0,01 27Al 0,22 0,2 0,35 31P <2 16,5 39K <5 42 33 69Ga 0,031 0,033 115In 0,0038 0,13 121Sb 0,0013 0,01 0,009 205Tl <0,0019 0,0093 209Bi <0,0016 <0,0018 232Th <0,0014 238U <0,0013 0,0096

7 Ge crystal contamination comparison
Element nat Ge X8 ppm Ge Ge 11B <0,01 27Al 0,32 0,27 0,22 31P <2 39K <5 69Ga 0,039 0,034 0,031 115In 0,006 0,014 0,0038 121Sb 0,0036 0,001 0,0013 205Tl <0,0021 <0,0019 209Bi <0,0018 <0,0016 232Th <0,0015 <0,0014 238U <0,0013

8 Ge isotopic measurement for GNO
% Ge72 Ge73 Ge74 Ge76 Carrier 70Ge 96,47 96,784 1,14 1,090 0,35 0,320 1,46 1,316 0,57 0,490 Carrier 72Ge 0,36 0,355 98,54 98,609 0,23 0,223 0,73 0,683 0,14 0,129 Carrier 74Ge 0,42 0,444 0,64 0,639 0,230 98,5 98,486 0,21 0,201 BLUE VALUE=LNGS measure carried out by ICP-MS (quadrupole mass analyzer) RED VALUE=MC-ICP-MS (double focusing-multicollector) carried out from ITU of JRC Karlsrhue

9 Ge crystal isotopic ratio measurement
GeX8 % Ge Ge 70Ge 19,08+/-0,10 19,14 +/-0,07 19,18 +/-0,09 72Ge 27,00 +/-0,13 26,95 +/-0,14 26,80 +/-0,09 73Ge 7,81 +/-0,04 7,76 +/-0,04 7,82 +/-0,07 74Ge 37,58 +/-0,21 37,63 +/-0,13 37,61 +/-0,24 76Ge 8,53 +/-0,10 8,52 +/-0,05 8,59 +/-0,05

10 Conclusion Concerning the contamination measurement of Ge crystal could be possible a little improvement (2or3 factor) of the sensitivity for some elements but we are close to the DL About the isotopic composition will be possible measure the enriched crystal with the same accuracy and precision