1 Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale
IMRT Commissioning Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale

2 Welcome to Arizona

3 Commissioning elements
Validating the dosimetry system Commissioning the delivery system Commissioning the planning/delivery systems for dosimetric accuracy Learning how to use IMRT

4 Step 0: the dosimetry system
Decide on the system that will be used to measure dose distributions Determine the measurement uncertainty Compare measurements of unmodulated fields to ion chamber scans Needed to establish acceptance criteria

5 Delivery system tests Need to create test patterns independent of planning system Need to check MLC positioning accuracy MLC dynamic motion accuracy Effect of rounded leaf ends MU linearity ...

6 Radiation field “offset” for rounded leaf ends
For rounded MLC leaf ends, there is an offset between the light and radiation field edges: ~0.6 mm Important in IMRT Light Radiation

7 Why important for IMRT?

8 Measuring the offset Irradiate a series of strips that abut
Use different values of offset Select offset value that gives best uniformity

9 Measuring the offset No offset 0.6 mm offset 1.0 mm offset
Here, 0.6 is best, i.e. subtract 0.6 mm from MLC settings Needed?? most uniform at junction? Integrated dose?

10 Measuring the offset

11 MLC accuracy Tenths of millimeters are important for IMRT
Use patterns of abutting strips for a sensitive test Test at different gantry and collimator angles Select subset for routine QA

12 Abutments are very sensitive to positioning
Offset 0.5 mm Offset 0.7 mm Offset 0.9 mm

13 DMLC tests For dynamic MLC motion, leaf speed and dose rate also need to be controlled Need to test for range of leaf speed and dose rates, as well as gantry and collimator angles Create patterns that move a 1 cm gap using all leaves at same or variable rates

14 Sweep all leaves at same rate
Chamber reading at center should be proportional to MU Film should show uniformity

15 Sweep leaves at different rates
e.g. Travel 5, 7, 9, 11, 13 cm in same MU Check relative dose

16 Special problem of matching accelerators
In ’99 Mayo Clinic Scottsdale had two “matched” Varian 2100C linacs, but IMRT doses differed by ~2.5% Needed to adjust MLC calibration on one machine by 0.13 mm

17 Delivery system QA Sweep a 5mm gap across a 10 cm span with Farmer chamber at center Take ratio to 10x10 open field 2.5% for 0.2 mm change

18 Routine QA checks for positioning accuracy
Abutment films 1 mm gap films Check at different gantry, collimator angles Check frequency(?)

19 …and now for a change

20 Planning system: Dosimetric accuracy
For IMRT, the MLC leaves move through the area of interest Final distribution is created by summing many beamlets New things become important Leakage through MLC leaves Penumbra defined by MLC leaves Small fields

21 MLC leakage - measure average value
Leakage through leaf (~2%) Between neighboring leaves (~5%) Measure using a pattern that fully closes all leaves - careful not to be under carriage or jaw

22 Penumbra Measure with film, diode, or microchamber, conventional scanning chamber too wide Subtle effects make a difference in IMRT Beam model based on penumbra measured with chamber Beam model based on penumbra measured with film

23 Dosimetric validation (1) Start with the basics
Define simple open fields irradiating a simple phantom check calculated output, PDD, profiles against standard measurements (just like commissioning any planning system) And shaped static fields also

24

25

26 Dosimetric validation (2)
Define simply-modulated fields with wide regions that can be measured with chamber and film (or equivalent) Check: Leakage/transmission Basic modulation Penumbra Out of field dose

27 “Leak100” – “Leak0” patterns

28 “Ridge” and “Trench” patterns
Useful for deciding minimum segment width 100% centers, 10% wings 10% centers, 100% wings “Incline” patterns 50% centers, 100% -10% wings

29

30 5mm strips

31 10mm strips

32 “Band100” – “Band0” patterns
60 0% 80 20

33 Finally, highly modulated single fields

34 What to do about differences?
May need to adjust the beam model May need to live with it That is, take known deficiencies into account when evaluating plans Very important to know about it, especially for critical structures

35 And once you are over the simple stuff …

36 Dosimetric validation (3)
Define targets and structures in a simple phantom: mock clinical situations Create inverse plan Measure doses with chamber, film, … Evaluate dose/MU in low gradient region Evaluate isodoses on many planes

37 Mock prostate

38

39 Colors: film, Gray: plan

40 Mock Head/Neck

41

42 Colors: film, Gray: plan

43 Mock spine

44 Critical structure plane

45 Target plane

46 Colors: film, Gray: plan

47 Dosimetric evaluation of mock clinical plans
What kind of agreement to expect in high dose, low gradient regions? ~2 - 4% is reasonable (chamber) What kind of agreement to expect for isodoses? Much harder to specify ~2 - 4 mm for 50% - 90% lines

48 How to be more quantitative?
Acceptance criteria should be stated statistically, such as: “within the region of interest, 95% of the calculation points should agree with the measurement to within 5%, which combines the desired degree of agreement with reality (3%) with the two-sigma uncertainty in the measurement technique (4%).”

49 Questions What is the region of interest?
What is the desired degree of agreement with reality? What is the uncertainty in the measurement technique? What percentage of the points should agree to the specified tolerance?

50 Questions For the points that do not agree to that tolerance, is there an outer limit of acceptability? Should the agreement be expressed as a percent of the local dose, prescription dose, or some other dose? Should distance to agreement be incorporated?

51 Another question What is the role of quantitative assessment of individual IMRT fields vs. composite doses?

52 Dosimetric validation (4)
Test calculations in the presence of inhomogeneities Measured/Corvus: 0.985

53 Dose/MU at 5 cm below 12 cm inhomogeneity
Measured/Corvus Bone Air Lung Water 0.978

54

55 Dosimetric validation - Summary
Know your measurement uncertainties Start with simple situations that can be more easily analyzed Design tests that focus on specific parameters, e.g. Transmission Small or narrow fields Penumbra and dose outside field

56 Finding your way around …

57 General principles (1) Don’t ask for the impossible
If you ask for NO dose to the cord, 60 Gy may appear just as bad as 40 Gy to the optimizer Look at a good 3D conformal alternative to get a starting point

58 General principles (2) Explain the problem sufficiently to the system
Define what needs to be treated Define what needs to be avoided The system is going to choose the beam shapes according to the structures defined

59 General principles (3) Define the problem sufficiently to the human who is doing the planning What is absolutely necessary What is desirable Where you are able to compromise

60 General principles (4) Understand the difference between three kinds of “prescriptions”: What you want and tell the human planner to get What the planner tells the system to try for What you eventually get and treat with

61 General principles (5) Learn what “knobs” there are
DVH criteria for targets and structures Relative weights or tissue types Number of intensity levels Number and direction of beams Try each individually and systematically on idealized and actual patients

62 General principles (6) Dose uniformity vs Conformality
Target dose uniformity can be expected to decrease with increasing concavity increasing dose gradient decreasing number of beams

63 General principles (7) Don’t assume IMRT is the way to go
3D conformal, judged by the same criteria, may be better parallel-opposed beam pairs are often best IMRT system may have limitations 1x1 beamlets insufficient weight to dose uniformity

64 Building experience with artificial problems
Designed to illustrate performance for certain types of situations Observe how changing various planning parameters affects plan quality and delivery efficiency For each, decide on relevant measures of plan quality

65 Simple cylindrical geometry
Single target with 1 cm PTV expansion (PTV is 8 cm diam, 8 cm long) Goal: target dose uniformity PTV max/min PTV D2/D98 Compare to: 3 open fields evenly weighted

66 Target dose uniformity
As you vary the requested degree of target dose uniformity, how do the results change? Corvus V5 “Prescription Panel”

67 Goal: target uniformity (Corvus V5)
% vol below x10 10x12 Min Max Rx Max/Min Max/Min D2/D # segments # MU Forward plans Inverse plans

68 Simple cylinder with 4 structures PipesEasy
Goal: structure sparing vs target uniformity PTV D2/D98 Structure mean/PTV mean 15 fields equispaced Try different structure goals 50, 20, 10, 2% of target dose

69 PipesEasy: Effect of changing structure goals

70 PipesEasy: Effect of changing structure goals

71 PipesEasy: comments Asking for too much sparing degrades target uniformity with little improvement 90% and 20% isodoses Structure at 10% Structure at 2%

72 What is achievable? Limiting dose gradient

73 C Shape Goal: structure sparing vs target uniformity PTV D2/D98
Structure D5/PTV 98 15 fields equispaced Try different structure goals 60, 50, …, 10% of target dose

74 Corvus: C Shaped Target
Structure at 10% Structure at 60%

75 Corvus

76 Limiting dose gradient
6%/mm

77 Buildup Expand PTV by 2-10 mm Evaluate dose uniformity CTV max/min
Volume max/CTV mean

78 100% 110% 5 mm below 0 mm below

79 Comments on these artificial problems
Good for gaining some feel for the “knobs” for the limiting conditions (e.g. maximum dose gradients) for the consequences of being unrealistic in the problem statement

80 Dealing with real clinical plans
Determine method/conventions for defining structures and targets Determine margins (CTV and PTV) These do not change whether IMRT or 3D conformal

81 Dealing with real clinical plans
Decide on criteria for an acceptable plan e.g. PTV dose must be sufficiently uniform: PTV D2/D98  1.15 Decide on parameter to be optimized e.g. minimize mean parotid dose These will often be competing and cannot both be “optimized”

82 Dealing with real clinical plans
Start with a 3D conformal plan to get a sense of what is achievable Use these results as a starting point for DVH goals for the inverse planner Start with relaxed goals and gradually tighten them

83 Welcome to Arizona … and IMRT
… there is a way out