1 Irradiated SINTEF 3D sensor tests with PSI DAQ system
Enver Alagoz 18 March 2012

2 Irradiated 3D SINTEF sensors
Irradiated at Los Alamos LANCE facility in 2010 with 800 MeV protons Integrated fluence is 1,2, and 5E15 neq/cm2 Total 6 sensors Sensors seem fine but ROCs are dead except one 3D plaquettes placed in the environmental chamber Only working sensors is 4E_B2-10_1 Fluence = 1E15 neq/cm2 4 electrodes 200 µm thick

3 Leakage current

4 ROC pixel efficiency Pixel efficiency map Chamber @ -20 °C
Sensor: 4E_B2-10_1 ϕ = 1E15 neq/cm2 -20 °C Pixel efficiency map

5 Threshold vs calibration delay
Sensor: 4E_B2-10_1 ϕ = 1E15 neq/cm2 -20 °C Threshold vs calibration injection delay

6 ROC pixel address levels
Address level decoding -20 °C Address levels Sensor: 4E_B2-10_1 ϕ = 1E15 neq/cm2

7 ROC calibration summary
-20 °C Sensor: 4E_B2-10_1 ϕ = 1E15 neq/cm2

8 Noise scan -20 °C Sensor: 4E_B2-10_1 ϕ = 1E15 neq/cm2

9 Charge collection Sr-90 source: 1 mCi, Eβ = 0.546 ΜeV
Random trigger used -20 °C Sensor: 4E_B2-10_1 ϕ = 1E15 neq/cm2 Vbias = -90V Thickness = 200 µm MPV = 9600 electrons

10 Charge collection scan
-20 °C Sensor: 4E_B2-10_1 ϕ = 1E15 neq/cm2

11 ROC settings