Sensistor ISH2000Training
Lennart Larsson
2
Sensistor ISH2000 Hydrogen Leak Detector
H2000 to ISH2000
3
Models and Specifications
Sensistor ISH2000 Sensistor ISH2000C Sensistor ISH2000P Operating time: 8 hours. Charging time: 6,5 hours. Charger: AC V 50/60Hz 24 VDC Current: max 3A AC V 50/60Hz All models: Sensitivity (Analysis Mode): 5x10-7 mbarl/s Range: 5x x10-2 mbarl/s Repeatability: Typ. ± (10% of measured value +3x10-7 mbarl/s)
4
Display Speaker Control buttons Earphone outlet Probe connector LED
LED indikeringar: Grön blinkande, uppvärmning pågar allt ok Grönt fast sken är lika med instrumentet är klart för att användas Rött fast sken tillsammans med Läck i displayen betyder en läcka är detekterad som är högre än den satta läckgränsen Rött blinkande, kolla displayen något är fel kontrollera i användare manualen Display Speaker Control buttons Earphone outlet Probe connector LED
5
Out and Inputs Printer port Active Probe Control 220 AC Connector
ISH2000 ISH2000P ISH2000C 24VDC Connector Printer port Active Probe Control Battery charger input Printer port Printer port Active Probe Control 220 AC Connector
6
Specification 500C 00C IP 64 10-90% RH ISH2000 3,9 kg ISH2000C 4,0 kg
ISH2000P 1,8 kg Bredd: 275mm Höjd: 155mm Djup: 170mm
7
P50 Hand Probe LED indications: P50 Flex
Green blinking: during warming up Green: instrument ready and under leak limit Red: instrument display indicates leak Red flashing: error indication, check user manual Both types of probes are fitted with dual LED’s that can be switched on to illuminate the area the probe tip is pointing at. P50 Flex P50
8
Sensor Change The sensor in the P50 probe can now easily be changed by the customers themselves. The sensor key (P/N ) will make the changing process easy. Note: when changing sensor the ISH2000 should be calibrated to ensure correct measuring data
9
Menu System New or changed functions: 1 Reference value
2 Reference unit 3 Correlation value 4 Multipoint Analyse 5 Multipoint Analyse Time 6 Audio Thresold 7 Show Reject Level 8Audio Ready Pulse 9 Reset Signal 10 Invert Colors 11 Probe Button 12 Probe Lamp 13 Audio Base Frequency 14 Info 15 System reset 16 Trigger Level 17 Number of Significant Digits 18 Service Mode
10
Menu System Reference value:
Should be equal to the value stated on the reference leak certificate. If the reference leak value is 4.2E-5 mbarl/s, then you should in the reference value field write 4.2E-05. Reference unit: Should be same as stated on reference leak certificate. If reference gas is used the unit should be PPM Correlation value: Conversion value between different units (needs to be calculated prior to entering the data in the Correlation value field) Multipoint Analysis: Summing up of the analysis results. Set a numeric value between 1 and 25 in this field to sum up the total leakage from all test points. The total value is used to determine if the object is leaking or not (above or below the reject point). Can only be used in Analysis or Combined Mode of the instrument. The APC function is not active when Multipoint is active. Multipoint Analysis Time: Min. time for Multipoint Analysis.
11
Menu System Audio Threshold:
The instrument can be set to be silent to a pre-defined % of the Leak Alarm Level Audio Ready Pulse: Sound is set to pulsating tone or no sound (silent) Show Reject Level: Displays the set Reject Level value in the display
12
Menu System Reset Signal:
Resets the sensor level in Analysis and Detection Mode. Invert Colors: Invert screen colors. Probe Button: Changes probe button functions, switch between Analysis and Detection mode Zero set detected signal in Analysis and Detection mode Initiate sample cycles (Multipoint measurement) Send values from the Probe to printer port Probe light on and off Probe Lamp:
13
Menu System Audio Base Frequency:
Adjusts the tone of the sound between hz in Analysis and Detection Mode Info: Information about program version, serial number and internet contact information System reset: resets all parameters to factory setting Trigger Level: Max level for ”Peak Hold” in Analysis mode (noise reduction) Number of Significant Digits: Number of digits after decimal point Service Mode: Used to identify sensor status and signal status in the instrument
14
Combox When the Sensistor ISH2000 is connected to other Probes than P50, an adapter ”Combox” must be connected in-between Combox