Intermediate and High Voltage Microscopy
د. عمر عبد القادر
2
د. عمر عبد القادر
3
The greater the accelerating voltage the shorter the l
Therefore, a 50,000 volt (50 kV) electron has a wavelength of nm and a 1MeV electron has a wavelength of nm! د. عمر عبد القادر
4
A million volt (MeV) TEM must maintain an accelerating voltage that is stable to within 0.5 volts.
د. عمر عبد القادر
5
A million volt (MeV) TEM must maintain an accelerating voltage that is stable to within 0.5 volts. With this coupled with a field emission source one can achieve resolutions that are in the range of 0.5Å د. عمر عبد القادر
6
Hitachi 3 MeV is the most powerful TEM ever made
د. عمر عبد القادر
7
Operator Hitachi 3 MeV is the most powerful TEM ever made
د. عمر عبد القادر
8
It requires its own separate building
د. عمر عبد القادر
9
Advantages of High Voltage EM
Increased Resolution: Although still a long way from a theoretical resolution of nm. the best actual resolution has been achieved with a 1 MeV field emission TEM. Nanocrystalline structure of silicon د. عمر عبد القادر
10
Advantages of High Voltage EM
Increased Resolution: Although still a long way from a theoretical resolution of nm. the best actual resolution has been achieved with a 1 MeV field emission TEM. 1 MeV kV kV د. عمر عبد القادر
11
Advantages of High Voltage EM
Increased Specimen Penetration: As accelerating voltage increases the ability of the beam to penetrate also increases. د. عمر عبد القادر
12
Dendrite in 3 mm thick section
Advantages of High Voltage EM Increased Depth of Information: The great depth of information allows an entire thick section to be viewed at once. Dendrite in 3 mm thick section د. عمر عبد القادر
13
Mitotic cell stereo pair
Advantages of High Voltage EM Increased Depth of Information: It also allows for stereo pairs to be created from thick sections. Mitotic cell stereo pair د. عمر عبد القادر
14
Advantages of High Voltage EM
Reduced Beam Damage: The increased speed of the electrons actually decreases the likelihood of an inelastic collision. The same section viewed with 1 MeV vs. 100 kV will have less contrast due to fewer scattered electrons د. عمر عبد القادر
15
The Boulder Laboratory for 3-Dimensional Electron Microscopy of Cells is located at the University of Colorado. It is a National Resource Center supported by the National Center for Research Resources (NCRR) at the National Institutes of Health (NIH). د. عمر عبد القادر
16
Intermediate Voltage (200-400 kV)
JEOL JEM-4000FX Hitachi H-9000NAR د. عمر عبد القادر