PENTRA 60 C +
2
PENTRA 60
3
THROUGHPUT ANALYZES CBC Mode /h 30 µl CBC + 5Diff Mode /h 53 µl
4
PHYSICAL SPECIFICATIONS
Weight: 35 Kg Height: 516 mm Depth: 444 mm Width: 481 mm
5
PHYSICAL SPECIFICATIONS
Power supplies: from 100 Vac to 240 Vac 50 Hz to 60 Hz Power Consumption: Maximum 400 VA
6
Measurements and computation
Impedance for WBC, PLT, RBC, BASO Photometry for Hgb Impedance, light scattering and cytometry for : LYM, MON, NEU, EOS, ALY, LIC. Computation from stored data that was directly measured for Hct, MCV, MCH, MCHC, RDW, MPV, Pct, PDW.
7
TECHNOLOGY Multi-Distribution Sequential System MDSS (ABX Patent)
Double Hydrodynamic Sequential Syst. DHSS Counting Principle IMPEDANCE Permanent Rinse Flow PRF SOLENOID VALVES ABX PATENT
8
System technology MDSS Sample Dilution Process
9
Sample delivery - Aspiration of 53 µl of whole blood
(5 Diff cycle) through the sampling needle - Whole blood split in 3 aliquots injected successively in 3 tangential flows of reagent
10
Sample Volumes
11
Step 1 : Sample Aspiration
12
Step 2 : Initial discard and rinse
13
Step3 : First Dilution
14
Step 4 : WBC/BASO Dilution
15
Step 5 : DIFF Dilution
16
Step 6 : Probe Wash
17
Step 7 : DIFF - Stop Reaction
18
Step 8 : RBC/Plt Aspiration
19
Step 9 : RBC/Plt Dilution
20
Step 10 : Hgb Dilution
21
Final Dilutions First Dilution Diff chamber RBC/PLT chamber WBC/BASO chamber Hgb chamber 10µl + 1.7ml Dil µl = 1ml Eosinofix µL Dln µl + 2 ml Alphalyse - 42.5µl After 12 sec ml + 0.4 ml ml Dil DILUTION DILUTION DILUTION DILUTION 1/ / / /200
22
Differential Dilution
25 µL Blood µL Fix µL Diluent = 1/80 The EosinoFix reagent lyses the RBC, stabilizes the WBC in their native forms and stains the cells with a specific coloration, the Eosinophils are colored most intensely. After 12 seconds reaction the solution is stabilized with Diluent and is then transferred to the flowcell for analysis.
23
General Characteristics of P60 C+ User Interface
Pilots all the analysers’ cycles: analysis, regulation, maintenance and grinding cycles. Calculates a result from the analyser measurements, pathology generation. Follow-up system of the analyser reagent quantities. Full view of result in colour Management of a work list to predetermine analysis to be carried out.
24
General Characteristics of P60 C+ User Interface
Stat analysis function. Management of a barcode reader to identify samples. Stocking of results, with consultation and research carried out by identification Validation and re-run of the sample on request of the user. Maintenance and regulation functions of analyser. Parameters allowing the user to adapt the ASP01 to its needs. Managment of papers ensuring tracability of sensitive actions.
25
General Characteristics of P60 C+ User Interface
QC: Quality Control. Calibration of analyser XB: patient quality control. Repeatability. Connection with a host computer system for reception of files and emission of results. Printing.
26
CBC Result CBC
27
CBC/DIFF result