Module building: plans & inventory
Admittedly not much progress the last fortnight. Grand plans nonetheless: Assemble first glass module very soon Populate remaining hybrids on panel with blank glass ASICs Exercise wirebonding on Liverpool training module Item Count Remarks glass populated hybrids 3 1x60µ, 2x80µ glue heights unpopulated hybrids 4 Note: 1 reject on panel dummy ASIC/mech sensor module 1 brown TM hybrids. not wirebonded blank glass ASICs 58 glass sensors 5 ATLAS12 mechanicals 2+49 ATLAS12 sensors 95 4 rejects tra-duct bipak 23+7 UBD Feb 2014, 9/2/2015 epolite 12 unknown UBD 18/6/2015 Bart Hommels - University of Cambridge
2
Hybrid 4 panel 1: metrology
Despite the £££ CMM, micrometer probably is the most accurate & reliable Reconstructed touchprobe heights overestimate glue layer thickness by 10µm Camera grabs through glass agree best with micrometer however this method cannot be used with real ASICs Lack of large gold surfaces outside ASICs make double touch method unreliable: touchdown on solder resist etc. etc. Unsurprisingly, the glue dot expansion ratio measurement is not reliable 18/6/2015 Bart Hommels - University of Cambridge