Introduction of NMIJ Gas Flow Standards Section
Toshihiro Morioka Gas Flow Standards Section Flow Standards Division NMIJ/AIST
2
Outline of GFSS Flow Rate Air Speed Low Gas Flow (various gases)
Medium Gas Flow (air only) Air Speed Low Air Speed Medium Air Speed
3
List of Facilities Flow Rate Low Gas Flow Medium Gas Flow Air Speed
・CVT System ・Weighing System ・Calibration Facility based on CFVNs Medium Gas Flow ・CVT System ・Closed-Loop Calibration Faccility Air Speed Low Air Speed ・Tow Carriage in Underground Tunnel Medium Air Speed ・Rotating Disk to Calibrate LDV ・Calibration Wind Tunnel
4
Flow Rate Ranges Low Gas Flow Medium Gas Flow (various gases)
(air only) 5 m3/h 200 m3/h CVT 10 μg/min 5 mg/min 180 g/min Weighing System CFNVs vs. Tank CVT CFVNs vs. CVT 1000 m3/h Flowmeters vs. CVT Closed-Loop Calibration Facility Calibration Facility Flowmeters vs. CFVNs Multi-Nozzle Chamber Flowmeters vs. CFVNs 5 MPa MPa 10 μg/min g/min 8 mm3/h 245 cm3/h 5 m3/h 9 m3/h 1000 m3/h 5000 m3/h Flow Rate converted into Atmospheric Condition
5
Rotating Disk, LDV, Wind Tunnel
Air Speed Ranges Low Air Speed Medium Air Speed 5 cm/s 1.5 m/s Tow Carriage in Underground Tunnel 1.3 m/s 40 m/s Rotating Disk, LDV, Wind Tunnel
6
Automatic Pressure Controller
Low Gas Flow : Weighing System 10 μg/min - 5 mg/min g/min Flow Rate converted into 1 bar : 245 cm3/h - 5 m3/h Uncertainty : 0.05 % % (k=2) Test Gases : Air, N2, H2, CH4, ... Vacuum Pump Filter Exhaust High-Flow Line Low-Flow Line Balance Systems Supply Gas Automatic Pressure Controller Three-Way Valve Cylinder Nozzle Holder Calibration Lines DUT : CFVNs Electric Balance Resolution : 0.01 mg Mechanical Balance Resolution : 0.1 mg
7
Automatic Pressure Controller
Low Gas Flow : CVT System 10 μg/min - 5 mg/min g/min Flow Rate converted into 1 bar : 8 mm3/h cm3/h Uncertainty : 0.42 % (k=2) Vacuum Chamber Vacuum Pump Vacuum Pump Exhaust Exhaust DUT Constant Volume Tank Three-Way Valve Supply Gas Test Gases : Air, N2, H2, CH4, ... Automatic Pressure Controller CVT in Vacuum Chamber Volumes : 0.1 L and 0.3 L Calibration Lines
8
Low Gas Flow : Calibration Facility
10 μg/min - 5 mg/min g/min Flow Rate converted into 1 bar : 245 cm3/h - 5 m3/h Uncertainty : 0.07 % % (k=2) Test Gases : Air, N2, H2, CH4, ... Reference Standard DUT Nozzle Holder Supply Gas Exhaust Automatic Pressure Controller Expand Flow Rate Range Ref. : CFVNs Multi-Nozzle Chamber Parallel Connection of 5 CFVNs Calibration Lines
9
Medium Gas Flow : CVT System
5 m3/h m3/h MPa Flow Rate converted into 1 bar : 5 m3/h m3/h Uncertainty : 0.17 % (k=2) Test Gases : Air Constant Volume Tank 11 m3 Double-Walled, Water Circulation for Quick-Stabilization of Temperature
10
Medium Gas Flow : Closed-Loop CF
5 m3/h m3/h MPa Flow Rate converted into 1 bar : 5 m3/h m3/h Uncertainty : 0.28 % (k=2) Test Gases : Air Closed-Loop Calibration Facility Ref. : Parallel Connection of 12 CFVNs
11
Low Air Speed : Tow Carriage
5 cm/s m/s Total Length : 45 m Uncertainty : 7 mm/s - 9 mm/s (k=2) Double-Walled Velocity Measurement by Laser Interferometer Tow Carriage Unmanned Space, Stable Temperature, No Wind Tow Carriage in Underground Tunnel
12
Ref. : Ultrasonic Anemometer
Medium Air Speed : Rotating Disk, etc. 1.3 m/s - 40 m/s Rotating Disk Rotating Disk LDV Uncertainty : 0.1 % % (k=2) Ref. : Ultrasonic Anemometer Uncertainty : 0.3 % % (k=2) Rotating Disk, LDV, and Wind Tunnel