1 Pre-Test-stands at MPI Munich
Xiang Liu Introduction to Canberra REGe detector Test A: Single- & double-photon escape Test B: Single- & multi-site events through coincidence Test C: Co60 photons angular correlation

2 Canberra REGE detector
Crystal inside Dewar Be window Reversed Germanium Detector Gerda Collab. , Jun 27-29, 2005 Page 2

3 Resolution expected Xiang Liu, MPI München

4 REGe efficiencies 10% REGe 10-2000keV 1% 0.1% 5keV 1MeV
Xiang Liu, MPI München

5 Test A: setup & simulation
Co60 Co60 Study single- double-escape events And back-splash events Gerda Collab. , Jun 27-29, 2005 Page 2

6 Test A: Measured energy spectrum
Gerda Collab. , Jun 27-29, 2005 Page 2

7 Test A: Closed look at Measured
SE: single escape, DE: double escape Gerda Collab. , Jun 27-29, 2005 Page 2

8 Test A: Simulated An improvement of simulation is undergoing.
Gerda Collab. , Jun 27-29, 2005 Page 2

9 Test B: Single- & Multi-site event selection
photon Co60 photon with single-compton scattering at 90 has fixed energy. Gerda Collab. , Jun 27-29, 2005 Page 2

10 Test B: Co60 source Gerda Collab. , Jun 27-29, 2005 Page 2

11 Test B: Measured spectrum
Esum = E_REGE1+E_REGE2 E_REGe1 E_REGe2 Gerda Collab. , Jun 27-29, 2005 Page 2

12 TestB: Esum=1173 or 1332keV Esum=1173 Esum=1332 E_REGe1 E_LEGe~360keV
Single Compton scattering 90degree Single-site event E_REGe2 Xiang Liu, MPI München

13 Test B: MC simulation Data & MC comparison, pulse shape: Kevin’s talk
Gerda Collab. , Jun 27-29, 2005 Page 2

14 Test C: Two REGe facing Co60 with the same distance at different angles. Require coincidence trigger, measure the angular correlation of the two Co60 photons. Gerda Collab. , Jun 27-29, 2005 Page 2

15 Conclusion Many tests done with the REGe detector.
Single- & double-escape photon events. Single- & multi-site events. Compared with MaGe MC simulations for verification. These REGe detectors will help to study the naked Ge crystals. Gerda Collab. , Jun 27-29, 2005 Page 2