Download presentation
Presentation is loading. Please wait.
Published byBernard Summers Modified over 6 years ago
1
Rafael Ballabriga Suñé Xavier Xirgu Aleixandre
Towards the automation of the test of photomultiplier tubes in Barcelona Rafael Ballabriga Suñé Xavier Xirgu Aleixandre
2
Introduction Measures performed Instruments and Automation Summary
Future Plans Open Questions Punts a tractar: Laboratory equipment: Ensenyar amb fotografies el setup que tenim al laboratori de la UB (2xPMTs, ADCs de càrrega, Oscil.loscopi, Taules mòvils i el seu control, Caixa negre i polsador de llum) Measures results: Quines mesures hem realitzat amb el setup. Current Status: Que estem preparats a realitzar, Estat del VME, Problema amb la font de llum. Future plans Dates de l’evolució del projecte. Summary
3
Measures Performed
4
Measures performed Absolute Gain Linearity
Scintillation Pulses Analysis Crosstalk
5
Gain Measurement Two Methods have been tested
Fitting the spectrum of large number of photoelectrons chosen 2 measures: spectrum/pedestal Easy to compute: Two methods have been chosen : Fitting the spectrum of a large number of photoelectrons Fitting the spectrum of few phe Mathematical simplicity of 1st method 2 ped sig phe σ ) μ ( N - = 19 10 602 . 1 ) ( - = phe ped sig N R G m
6
Gain Measurement (Setup)
7
Gain Measurement (Results)
8
Linearity measurement
Double pulse method has been used. Deviation from linearity expressed as: 100 (%) h H D - = Ample de banda de l’oscil.loscopi!!!! Comentar el problema actual!!!
9
Linearity Measurement (setup)
10
Linearity Measurement (results)
11
Linearity as a function of High Voltage
Linearity is a parameter dependent on the High Voltage The aim of this measure is to study this dependence Relative linearity has been found: A small high frequency noise has been observed when the LED driver is pulsed (~400μV peak)
12
Linearity as a function of High Voltage
13
Linearity as a function of High Voltage
14
Scintillator Pulse Analysis
Objective: To study the performance of the system Scintillator-Fiber-PMT A 12x12 pad has been used with 101 cmts WLS fiber Aiming at: Mean number of photoelectrons/MIP Percentage of signal within 25 ns windows Podem tenir altres centellejadors per testejar??? Mascares???? Nomes tenim 1!!!!!
15
Scintillator Pulse Analysis (Setup I)
16
Finding the fiber’s optimal position (setup)
17
Scintillator Pulse Analysis (results)
Histogram of the number of photoelectrons per signal Time response waveform
18
Scintillator Pulse Analysis (results)
Mean time waveform in lineal axis (time vs voltage) Mean time waveform in logarithmical axis (time vs voltage) Slope= E-1 Statistics=9568signals
19
Scintillator Pulse Analysis (results)
20
Scintillator Pulse Analysis (results)
21
Crosstalk Measurement
A measure of the system Scintillator-Fiber-PMT A system to position automatically the fibers in the cathode has been designed The mask used in this test places diagonally the fibers in the fired pixel
22
Crosstalk Crosstalk results of two studied channels.
2 diagonal WLS fibers are placed in close contact with each channel. 2.51 100 3.35 2.01 2.09 29 28 36 27 20 35 19 37 21 2.35 100 2.60 2.43 2.26 45 44 52 43 36 51 35 53 37
23
Crosstalk 2.02 100 2.84 1.62 2.43 21 20 28 19 12 27 11 29 13 4.24 4.08 7.18 5.8 In the upper figure the fibers are placed in close contact with the cathode. In the bottom the fibers are placed 0.5 mm from the pixel.
24
Instruments & automation
25
Instruments and automation
2x64 channels PMT H7546 2x1 channel PMT XP2262 Light source (mimics scintillator pulses) 4 channel High Voltage Source 2x32 channel charge ADC (CAEN V792) PC with VME-MXI-2 interface (with LabVIEW software) 3 axis-motion stage ( X-Z motorized, Y manual) Oscilloscope TDS 380 Timers, discriminators, logic units Xp2262 used as a trigger PMT in test with cosmic rays. Amplada de banda oscilloscope The last ones are in a NIM standard Funcionament de la font de llum
26
Some pictures of the setup
27
Some pictures of the setup
28
Some pictures of the setup
29
Some pictures of the setup
The LED pulser
30
Summary
31
Summary The setup in Barcelona has been shown
It is still under construction aiming at a fully automated system Some measures have been made in order to test the performance of the PMT and calibrate the system
32
Future plans
33
Future plans Deadline: October 2003
Reduction of the noise of the LED pulse (1 month) in parallel with signals analysis Which tests are to be performed on each PMT? Absolute Gain of each channel Tests to be performed on some PMTs? Linearity Scintillator pulse analysis (Crosstalk) Deadline: means a massive test has to be designed and ready by October 2003 Fully Automated Criteria to decide which channels of which PMTs must be tested microVolt noise Can we get a complete set of pads in order toe reproduce the system in our Lab?
34
Open questions
35
Open questions Define the tests to be performed on each tube & criteria to decide which cannels Base study status Aging problem Performance under magnetic field? Possibility to test in the lab a complete set of Scintillators-fibres-PMT I have been abroad (Aging problem)
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.