Download presentation
Presentation is loading. Please wait.
1
Hadamard Transform Imaging
Paul Holcomb Tasha Nalywajko Melissa Walden
2
Problem Definition Current 3D imaging systems for brain surgery are too slow and possess too low of a resolution to be effective in an operating room setting *Cannot distinguish the tumor margin with MRI imaging
3
Tumor types Benign tumor Primary malignant Secondary malignant
Images:
4
Why is this important? 71% mortality rate for diagnosed brain tumors
Correlation between complete resectioning of tumors and improved prognosis Complete resectioning requires knowing the location of the tumor, especially tumor margins Imaging in a clinical setting should be fast Operating room billed by the quarter- or half hour
5
Cost/Benefit Analysis
Treatment costs: OR cost: $10K - $15K per surgery (depending on length) ICU: $1963/24 hrs Floor: $779/24 hrs Chemotherapy: $__/treatment Radiation therapy: $__/treatment Cost Reduction: Shorter surgeries Less time in hospital (ICU or floor) Less post-surgical treatment required
6
Design Criteria Must produce an image in real time
Must accurately reproduce area of interest in the brain Must distinguish healthy versus tumor tissue Must be small enough to be usable in an operating room setting Must interface with operating microscope
7
Design Objective Construct imaging system using digital micro-mirror device and Hadamard transform for use with operating microscope in a clinical setting
8
System Design Hadamard Transform Decreased imaging time Increased SNR
Hadamard Matrix Definition Inverse Hadamard Transform Digital Micro-mirror Device Allows use of Hadamard Transform
9
Fourier vs. Hadamard Imaging
SNR Increase with Hadamard: √n SNR Increase with S-Matrix: (√n)/2 Wuttig and Riesenburg, “Sensitive Hadamard Transform Imaging Spectrometer”
10
Illuminate sample with white light Collect reflected light
System Diagram Illuminate sample with white light Collect reflected light
11
Decrease image size to fit within 512 x 512 matrix
System Diagram Decrease image size to fit within 512 x 512 matrix Magnification:~0.4
12
Apply Hadamard matrix using DMD Compress image to 160um line
System Diagram Apply Hadamard matrix using DMD Compress image to 160um line 1 -1 1 -1
13
Apply inverse Hadamard transform using computer
System Diagram Disperse light spectrally using spectrograph and collect image using CCD camera Apply inverse Hadamard transform using computer X Y Spectrum
14
System Output
15
Design Timeline March: Finish compression system; insert, align, and test spectrograph; test system using reflectance standard to determine SNR April: Test system using normal and tumor tissue samples and present findings at Senior Design Day, resubmit NCIIA grant proposal
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.