BioTector B3500ul technical presentation
November, 2017 Company Confidential © Hach 2017
2
B3500 Series – Low Cost of Ownership TOC Analyzer Uncompromised reliability for your less demanding applications New B3500e B3500c B3500dw B3500ul Typical Application Industrial WW effluent Condensate return Municipal drinking water Power Municipal WW effluent Cooling water Potable water Petrochemical Demineralized Water RO and Demineralized Water TOC Range 0-250 or 1000 ppm 0-25 or 100 ppm 0-25 ppm 0-5 ppm (0-5000ppb) TOC Mode TIC/TOC, TC TIC/TOC, TC, VOC TIC/TOC, % TOC Removal Two Stage Adv. Oxidation Yes Self Cleaning Internal sample tubing and reactor Reactor Sample characteristic < 100 µm particles <100 µm particles Some fat, oil, greases, sludge Clean Some pH swings Free of fat, oil, greases. Stable pH Free of fat, oil, greases. Stable pH Number of Streams 1 2 Company Confidential © Hach 2017
3
B3500ul - APPLICATION target: ultra pure water
Very low organic concentrations, 0-5ppm Clean sample, free of fat, oil, greases. Particulates are smaller than 100<µm Examples: Boiler feed water Condensate return monitoring RO monitoring Company Confidential © Hach 2017
4
Why B3500ul model was developed?
The “ul” version of BioTector enables detection of ppb level TOC changes ppb TOC level changes are an early indication of corrosion or inefficient operation in critical equipment such as turbines and (ultra) pure water RO systems Key markets include: (Petro)Chemical, Power, SemiCon. and other markets where TOC is measured The “ul” product enhances the portfolio of BioTector products by delivering resolution of ±2% or 10ppb detection capabilities Company Confidential © Hach 2017
5
B3500ul Analyser performance – accuracy and linearity
BioTector’s Product Leadership in the on-line analyser market is characterized by reliable and trusted TOC analysis +2% -2% B3500ul achieves excellent stability which can be demonstrated during long term tests where specified ±2% accuracy is maintained . BioTector B3500ul has an excellent linear response across the entire ppb range. Company Confidential © Hach 2017
6
B3500ul Analyser performance - carryover
B3500ul TOC analyzer has no sample-to-sample carryover. When a “dirty” 1000ppb KHP standard was analyzed no sample-to-sample carryover was recorded in the follow-up analysis of a once ultra-pure water sample. Company Confidential © Hach 2017
7
The main components inside the outer compartment of the
BioTector B3500ul Power PCB, with user terminals for Power, Relay Contacts and 4-20ma signals CO2 analyzer 24V DC Power Supply with voltage selector Ozone Generator Reagent Pumps Cooler On/Off switch Sample Pump Ozone Destructor Manual, Calibration or Multi-stream valve Reactor Valve Sample valve Exhaust Valve Reactor with integral mixer Company Confidential © Hach 2017
8
The main components inside the inner compartment of the
BioTector B3500ul Oxygen Concentrator Tank with regulator Oxygen Concentrator Sieve Beds Oxygen Concentrator HEPA filter Cooler Fan Oxygen Controller Board with pressure sensors, flow sensor and control valve (MFC) Instrument Air Inlet with Isolation valve Oxygen Concentrator Rotary valve Mixer Reactor Motor Company Confidential © Hach 2017
9
B3500 “c” and “ul” System Comparison
B3500ul B3500c B3500ul Enclosure 750mm x 500mm x 320mm 1000mm x 500mm x 320mm CO2 analyzer 10000ppm Hastelloy 2000ppm SS Reagents Acid 6N Base 4N Acid 1.8N Base 1.2N Sample injection volume 20p (~7.5ml) 30p (~11.25ml) © Hach 2017
10
Critical B3500ul Service Procedures.
System pressure settings Ozone line filter check Sample and reagent pumps pumping rates Filter mats CO2 analyzer service NF300 mixer reactor diaphragm replacement All detailed commissioning and service procedures can be found in below documents: B3500ul User Manual B3500ul Service check sheet Service check sheet and manual for BioTector Compressor (if installed with BioTector B3500ul). The BioTector B3500ul TOC analyzer Service and Maintenance presentation
11
To check the system pressure settings.
Check the instrument air supply pressure at the filter pack pressure regulator. The pressure should be 1.5 bar (1.2 bar if BioTector compressor is used) when the oxygen concentrator is not using any instrument air. This typically cycles from 1.5 bar to 0.9 bar when the oxygen concentrator is running. This pressure is also shown on the O2 Controller Status menu. Check the O2 PRESSURE SENSOR in O2 Controller Status menu. The pressure should be between 390 mbar and 400 mbar at the idle 1 l/h oxygen flow. At a 60 l/h MFC SETPOINT setting, the pressure should not be less than 320 mbar. The BioTector B3500ul contains a number of sensors to continuously monitor system performance, and report any issues which may occur. Details in the photograph. Using the appropriate Simulate menu, check the operation of the Mass Flow Controller (MFC) and confirm that the MFC is working at various flow set points. Regulator for adjusting oxygen pressure. Sensor for incoming instrument air pressure. This is monitored and shown on the LCD. Sensor for oxygen pressure. This is monitored and shown on the LCD. Flow sensor and control valve which together make the Mass Flow Controller (MFC).
12
The set point air pressure supplied from BioTector compressor should be 1.2 bar and the tube between compressor and the analyser should be as short as possible. BioTector Compressor. Compressor PCB Excess air discharge silencer and flow control valve Air Outlet to B3500ul. Outlet can be changed to opposite side of enclosure if required Compressor with air inlet filter / silencer Cooling fan with air filter Power requirement, 230V 50Hz or 115V 60Hz, 300W. Control cable from BioTector stops the compressor when the B3500ul is stopped (supplied with compressor). Pneumatic connection for air: 3/8” Nylon or PTFE tube.
13
To check the Ozone Line filter.
Salts and corrosion products can reduce ozone concentration and as a result decrease TOC recovery level. Remove the Ozone Line Filter and wash it with DIW (or tap water). Dry the filter well and reinstall in place.
14
Replace the filter mats, check the fan.
Replace the 100mm filter mats in fan and vent housing. The fan cover is levered out from the top with a screwdriver. Slot for screwdriver. Filter matt. The operation of the fan can easily be checked with the fan cover off. Note that the fan is thermostatically controlled, but can be activated from the appropriate simulation menu. To achieve optimal efficiency of the fan filter mats should be installed according to below diagram
15
To check the BioTector B3500ul CO2 Analyzer
The BioTector B3500ul CO2 analyzer is located in the top part of the enclosure and for service it needs to be removed from the enclosure. To remove the analyser: - disconnect the cable, - unscrew 4 nuts and remove the blue clamps, Below: The B3500ul CO2 analyzer. CAUTION: During service, do not remove the cover over the detector electronics, or disassemble the detector section of the CO2 analyzer. During service, do not remove any of the CO2 absorbers on the CO2 analyzer. Do not remove the cover over the source electronics.
16
Use gloves to reduce possible contamination
To change the diaphragm in the NF300 mixer reactor CAUTION: A torque screwdriver with TX20 bit calibrated to 1.4Nm, or 3mm Allen bit calibrated to 1.5Nm (for pumps built with metal cover plate) is a requirement for this service operation. Do not proceed unless you have this tool. Remove the screws on the front of the Mixer Reactor and remove the reactor motor at the back. Reactor and Mixer viewed from the side. Note that the reactor is independently bolted to the door panel with 4 screws from the back (2 screws visible above). These should not be removed. 4 screws located here Use gloves to reduce possible contamination
17
BioTector B3500ul typical site installation.
18
Initial contamination removal
Run ~10 analysis cycles to remove TOC contamination introduced during service of stoppage Shipping and long storage can introduce TOC contamination which will affect the zero offset. Connect DI Water or alternatively tap water to manual port and using Manual Program run 10 analysis cycles to remove the organic contamination. If the TOC results will not stabilize (within ±20ppb) run additional reactions using DIW. OPERATION MANUAL PROGRAM RETURN TO ON-LINE SAMPLING NO MANUAL 1 , RANGE 1 MANUAL - , RANGE - © Hach 2017
19
Zero Calibration Zero calibration carried using acid and base reagents only (default) Using the reagent configurations from the selected range (Range 1) a zero calibration is carried out by injecting the TIC acid into the reactor, followed by the base which is then oxidized, followed by the TOC acid. The zero offset is calculated from the result. To neutralize the fluid in the reactor, an appropriate volume of base is injected into the reactor which neutralizes the fluid in the reactor. Note, this is not required in the normal reaction, as the volume in the reactor is much larger due to the volume of liquid injected by the sample. The above procedure is repeated for the programmed number of zero reactions, and the zero offset calculated from the results. When this is complete, confirm that the zero readings are stable. © Hach 2017
20
Hach TOC standards for B3500ul system (acidified with H2SO4)
Span Calibration and Validation TOC Standard solution for B3500ul analyzer cannot be acidified with Phosphoric acid as this could affect the pH, reduce the oxidation and introduce significant error in TOC measurement. For calibration and validation use below Hach standard solutions: Connect the standard solution to the MANUAL/CALIBRATION port. If both of these ports are not available, use the SAMPLE 1 port, but first set the Sample Pump REVERSE time to 0 (zero) seconds in Sample Pump menu. It is recommended that the standard solution is located at the same height as the sample chamber or catch-pot. Run the Span Calibration (or Span Check for validation) cycle using the “RUN SPAN CALIBRATION (or RUN SPAN CHECK)” function in Span Calibration menu. A minimum of 5 span calibration reactions is recommended Hach TOC standards for B3500ul system (acidified with H2SO4) KHP 5mgC/l ( ) - for calibration KHP 1mgC/l ( ) – for validation © Hach 2017
21
Service Zero To ensure that the correct zero offset is applied and to eliminate any remaining TOC contamination from the analyzer, a Service Zero cycle should be initiated using the SET SERVICE ZERO function in the Service menu. When initiated, BioTector automatically runs a total of 10 zero calibration cycles during the following 500 online measurements and automatically adjusts the zero offset values to compensate for the removal of the contamination. It will therefore not be necessary to revisit the BioTector after the start-up or to repeat the Zero Calibration cycle manually. See Section Service in the User manual for further details. Inform the client that the analyser will automatically correct the zero offset during approximately 48h and elevated result of 20-50ppb as the system expels contaminants are expected. © Hach 2017
22
B3500ul analyser start-up sequence during commissioning and PM.
1 Sequence: 10 DIW reactions (point 1 on the chart) – removal of the main portion of TOC contamination, Zero calibration (point 2 on the chart) – initial zero offset adjustment to minimalize error during span calibration 5ppm Span calibration (point 3 on the chart, outside scale) – calibration factor adjustment Service Zero (points 4 on the chart) – final removal of remaining TOC contamination and background adjustment (total of 10 zero calibration cycles during the following 500 online measurements). Inform the client that the analyser will automatically correct the background during approximately 48h and elevated result of 20-50ppb as the system expels contaminants. © Hach 2017
23
BioTector’s Two Stage Advanced Oxidation (TSAO)
Advantages Strong and very powerful well proven (TSAO) oxidation Low maintenance (6 month prev. service) No calibration and/or cleaning required in between service intervals! High reliability (99.86% up time), nearly SIL-1 availability! Self cleaning reactor technology Self-diagnostic system/service feature No material built-up inside the reactor Company Confidential © Hach 2017
24
Materials available for service and technical support teams
HSO Biotector B3500ul Resource Library B3500ul Resource library Technical presentations: - BioTector B3500ul Introduction and Operation - BioTector B3500ul Commissioning - B3500ul Process Tests - B3500ul Service and Maintenance User Manual – 10-MAT-652 Drawings B3500ul Spares and Accessories Part List Consumable Wear Parts BioTector TOC B3500ul TOC standard solutions: , 5ppm calibration TOC standard solution , 1ppm validation TOC standard solution Accessories 19-COM-160 BioTector Compressor 115V / 60Hz SMC-001 B5C Instrument Air Filter Pack 19-COM-250 BioTector Compressor 230V / 50Hz BAS-031 Biotector Sample Overflow Chamber 19-KIT-126 Compressor Service Kit - 6 Month KIT-123 B5C 6 Month Service Kit 19-B5C-030 BioTector Network Control Unit. System-C Company Confidential © Hach 2017
25
B3500 Series typical Installation
Hach BioTector B3500 TOC analyzer For B3500ul analyzer 20l reagents drums are suitable for 6 months operation with 10 minutes cycle time. Company Confidential © Hach 2017
26
B3500ul Key Accessories Compressor Sampling chambers
19-COM-160 BioTector Compressor 115V / 60Hz 19-COM-250 BioTector Compressor 230V / 50Hz Sampling chambers “Sand Traps” can be used to reduce the sample pressure or separate corrosion products caused by rouging of SS in high purity hot water systems. 19-BAS-014 Flow Through Sampling Chamber with Sand Trap 19-PCS-054 Sampling Chamber with Sand Trap Required if plant air is not available or pressure is to low BioTector Sample Overflow Chamber 19-BAS-031 Optional chamber design to reduce the pressure of the sample. Imperial and metric fittings are available. Compatible with all BioTector analyzer models. Company Confidential © Hach 2017
27
B3500ul additional accessories – reduction panels
In steam/condensate return application temperature and pressure can exceed analyser specification and reduction panels needs to be installed. This equipment is manufactured by Sentry, can be modified by company Technopomiar and therefore complete solution could be offered to the client. EU ADM Power team leads this project and additional information can be found at below links: Hach STEAM Panel Solution PPT Sentry Sample Conditioning Panel Example of temperature and pressure reduction panel. Company Confidential © Hach 2017
28
Network Control Unit (19-B5C-030)
Company Confidential © Hach 2017
29
Company Confidential © Hach 2017
30
Biotector B3500 SERIES – WHY SERVICE
A Hach Service Agreement ensures maximized Plant/ Production Up-Time and Operational Process Safety In order to maximize measurement reliability & instrument uptime, B3500 requires that the wear parts are exchanged as required or with the typical replacement frequency Factory recommended maintenance interval twice per year A Hach Service Agreement protects your investment and helps ensure that you stay in compliance and on budget. You receive expert knowledge from the instrument supplier and in case of breakdown our contract customers get priority response time Hach experts have years of experience working on Hach equipment, and applying their certified expertise will ensure that your equipment stays as good as new!
31
Services Biotector B3500 Series (EU only)
FIELD Service S-FIEX2-BT3500 FIELD PLUS Service S-FIEP2-BT3500 PREMIUM Service S-PREX2-BT3500 PREMIUM PLUS Service S-PREP2-BT3500 Exclusive Priority Service S-EXLP2-BT3500 COMMISSIONING Service S-XCOMM-BT3500 B3500ul B3500c B3500s B3500e B3500dw For detailed Info about the different Service Programs, visit our Share Point Service * Preferred Service Offerings in bold
32
BIOTECTOR B3500 TOC Analyzer
Field Maintenance Central Maintenance Wear Parts Field Repairs Central Repairs Spare Parts Travel Shipping Response Time HACH BIOTECTOR B3500 systems are MCert. Certified with a 99.86% Up-Time and with patented two-stage advanced oxidation technology provides continuous, accurate TOC process information for ‘clean water’ and “effluent water” applications with high performance. WHY WHAT WEAR PARTS To ensure maximized Plant/ Production Up-Time and Operational Process Safety To ensure max. instrument reliability and accuracy ; which allows Plant Management/ Operators to take non-questionable decisions for Process Control and insight Professional instrument Commissioning with correct appl. and system setting as per needs/requirements Inspection of crucial parts like valves, pumps, O3 generator and NDIR CO2 analyzer Required exchange of mechanical parts, tubing's, seals and filters Cleaning NDIR, change of Acid-Base reagents, Zero-ing and Cal. Verification of the B3500 Maintenance interval twice per year 6 Month Service Kit 2.0 / Year Sample Pump 0.5 / Year Acid Pump 0.3 / Year Base Pump O2 Concentrator Valve
33
BioTector b3500ul is available for sale!
B3500 series delivers lowest total cost of ownership without compromising reliability for the less demanding applications! Full set of sales tools developed to support you through your sales effort! Questions ?
34
Q&A © Hach 2017