1 for pure liquid samples
IR sample preparation for pure liquid samples

2 What is a thin film cell? a thin film cell
is a drop of pure liquid sample sandwiched between salt (NaCl) discs

3 Thin film IR cells Salt (NaCl) discs are used To hold the sample
So the holder won’t absorb infrared light and only sample absorption is detected

4 Sample preparation so only dry (anhydrous) samples are used
because even traces of water etch the IR discs affecting the path length of light and the optical quality of discs

5 use clean gloves to work with salt discs
put on an extra pair of gloves to protect cells from anything aqueous on your regular gloves

6 Preparing salt discs Slide off The top of the Desiccator Remove

7 replace desiccator cover immediately remove discs from sleeves and rest on kim wipe

8 to clean discs add a few drops of chloroform to each disc rub gently
to clean off all traces of sample

9 adding sample add one or two drops of sample in the middle
of the 1st disc lower the 2nd disc on top as shown to spread sample evenly

10 placing discs in cell holder
use your finger to make a platform in the cell holder lower in the NaCl discs do not wedge them

11 gently place lid on top slide into position in light beam

12 close sample compartment remove gloves before using keyboard

13 to clean discs add a few drops of chloroform to each disc rub gently
to clean off all traces of sample

14 Keep discs dry replace in desiccator after cleaning