Presentation is loading. Please wait.

Presentation is loading. Please wait.

AriaMx Real-Time PCR System

Similar presentations


Presentation on theme: "AriaMx Real-Time PCR System"— Presentation transcript:

1 AriaMx Real-Time PCR System
Better Systems | Brilliant Results Total Confidence Real-Time PCR for Today and Tomorrow

2 A COMPLETE REAL-TIME PCR PORTFOLIO

3 AriaMx AMPLIFIES YOUR PRODUCTIVTY
UPSTREAM EFFICIENCY | EASIER WORKFLOWS Detect bias in NGS runs before you run all of your samples Confirm alleles in gene expression and genotyping arrays Thermal precision and uniformity enable you to detect alleles better than other systems Easy reporting that exports in multiple formats Track experiments across multiple sites or collaborators ROBUST VALIDATION 100,000 samples, 14 assays and chemistries across

4 READY, GO! Factory calibrated:
Each instrument is calibrated as it leaves the factory so you have easier first time set up Blazing scan times: All channels, less than 2 seconds irrespective of plex number/optical channels used Ultra-fast reagents: Supports SYBR, Probe and HRM reagents STATS R2 = 0.999 Slope = E = 93.7 Run Time = 42 minutes Brilliant III SYBR Chemistry This assay shows 8 orders of magnitude change, but instrument can perform10 orders of magnitude measurements

5 CHOICES TODAY & TOMORROW
Next generation optical cartridges Each cartridge has 8 excitation and emission read points, one for each row minimizing edge and row effects 8 Photodiode detectors and dichromatic mirrors at each positions to reflect emitted light to photodiode Instrument always saves data for all wells and for all the optical cartridges that are inserted Lower instrument life-time costs On-board diagnostics mean less on-site field engineer visits – in-lab upgrades and in-lab solutions for instrument maintenance Intuitive interface Touch-screen makes the instrument easy to learn

6 SUPERIOR DATA, MULTIPLEXING
Standard Curve for all five channels Target information table Uniform performance across all channels Consistent performance provides you with confidence in routine testing No reference channel needed Maximize the productivity of your instrument Customize your instrument’s plex level With modular optical modules you can customize the instrument to your exact needs today and add modules when needed Ct difference between single-plex and multiplex assays is less than 0.7

7 SUPERIOR DATA, 2 FOLD DETECTION
Copy number detection AriaMx has the ability to detect down to a two fold change in copy number on various dye chemistries Relative or quantitative applications AriaMx can run a wide variety of applications – gene expression and miRNA detection in relative or quantitative assays STATS R2 = 0.999 Slope = Efficiency = 92.9 Chemistry: TaqMan Channel: FAM

8 SUPERIOR DATA, THERMAL PROFILE
Amplify your sensitivity AriaMx has little well to well variation. Our thermal uniformity assay standard of deviation is less than 0.2Cq Superior thermal accuracy at target AriaMx can reach target temperatures less than 0.2C discrimination. Temperature Uniformity Tight validation standards had well to well variability within 0.2C 97% of the time for thousands and thousands of samples run Six peltier elements 2 Ceramic plates with semi-conductor elements contribute to accuracy and robust instrumentation.

9 fastest in the industry
SUPERIOR DATA, HRM REAGENTS Class 3, 76bp amplicon Class 4, 142bp amplicon Class 1, 220bp amplicon Class 2, 114bp amplicon Coming in November EVA GREEN HRM fastest in the industry Early Access Available

10 AriaMx Specifications
Feature Description Excitation source 8 dye specific LEDs per optical module Detection sources 8 silicon photo-detectors per optical module Dyes (at launch) SYBR/FAM – 516.0nm Detectable Ct for 0.001ng/DNA per reaction (R2 > 0.980) HEX – 555.0nm  ROX – 610.0nm CY – 568.5nm CY nm 6 slots, swappable cartridges (Inquire about custom cartridges) Dye selection Excitation and emission Reaction volume 10ml to 30ml Chemistries Supported SYBR, Probe, HRM Thermal system Six peletiers made from two ceramic plates with semi-conductor elements, 96 well Thermal system temperature range 25.0 – 99.9 C; Heating: 6C/Sec; Cooling: 3C/sec Accuracy: ± 0.2°C or better of target Confidentiality Label October 14, 2018

11 AriaMx Specifications
Feature Description Dynamic Range 10 fold Comparative Quantitation 2x (Range 1.3x to 3x) 10x (Range 6.5x to 15x) Multiplex Difference in Ct between single-plex and multiplex assay wells is <0.7 Sensitivity (FAM) Discriminates between 2 fold populations ranging from 100k to 12 copies with 95% confidence Uniformity Ct St Dev (5s/10s) <0.20 5 colors, SYBR and TM assays Electrical power (input) VAC, 50/60Hz, 1100A Operating environment 18-35 °C (65-95°F), 85% non-condensing humidity 7500 feet, max altitude Weight 50 lbs. (23 kg) Dimensions 19.7” W x 18.1 “D x 16.5” H (50cm x 46cm x 42cm) Sample Containers (Consumables) 0.2ml tubes 96 well plates Strip tubes Confidentiality Label October 14, 2018

12 AriaMx Specifications
Feature Description Warranty 2 year warranty is standard 5 year warranty and service packages available Onboard analytics Thermal, Physical, Interactive (sensors) tests Extended: 125 performance points tested in 30 minutes Start-up: 59 performance points tested in ~1 minute Optional bypass of both features Services (upon request) Installation and Familiarization Standard and Enhanced Preventative Maintenance Additional yearly warranty (+1 increments) Return-to-Agilent Instrument Exchange Program Thermal block verification Confidentiality Label October 14, 2018

13 APPENDIX Additional AriaMx Data

14 SUPERIOR DATA, MULTIPLEXING
FAM ROX HEX CY5 CY3

15 Uniformity Assay: Standard Deviation is < 0
Uniformity Assay: Standard Deviation is < 0.2Cq 3 runs combined using multiple experiment analysis Chemistry = TaqMan Dye = ROX

16 High Thermal Uniformity and Discrimination: SNP Analysis using High Resolution Melt Curve Analysis
Class 1, 220bp amplicon Class 2, 114bp amplicon Class 3, 76bp amplicon Class 4, 142bp amplicon Class 4, 96bp amplicon

17 Consistent Performance with Multiple Dyes
FAM ROX HEX CY5 CY3 PR May 1, 2016 For Research Use Only. Not for use in diagnostic procedures.