SPAG Delivery through the Ventilator
David R. Gibson 2016
2
Gather Parts: Non-disposables
Complete SPAG-2 set-up. 50 PSI O2 hose Blender/hoses set-up
3
Gather Parts: disposables
3 filters 4-22mm ID (large) adaptors 2 – Briggs “T” adaptors One-way valve Corrugated tubing PEEP valve & adapter Scissors
4
SPAG Set-up Drying Flow OFF 1. Set up the SPAG as normal
The input should come from a blender, and Blender set same as the ventilators FIO2%. Drying Flow OFF 1. Set up the SPAG as normal See: PowerPoint on RTConnection.org explaining how, Wall outlets connections with the exception that the drying flow should be left OFF.
5
PEEP Relief Valve Set-up
2. Set-up PEEP valve and adapter: Set PEEP on valve to same level of PEEP on the ventilator Bacteria filter between briggs “T” and PEEP valve
6
Vent Circuit Set-up Patient wye 3. “T” into inspiratory limb about a tidal volume length away (two 50cc sections for each 100cc tidal volume) 4. Use a one-way valve to avoid loss of tidal volume during inspiration. Tidal Volume Length away One-way Valve goes this direction To SPAG
7
Other Considerations Assess for obstruction frequently!
Suction PRN Watch for sudden high pressure limits May need to changes expiratory filters Discard HME during SPAG delivery, and replace afterward.
8
Expiratory limb Set-up
Q2 ! Place two filters on the expiratory limb. Change Q2 hours…with each vent check
9
Summary view