1
May 2012 Project: IEEE P Working Group for Wireless Personal Area Networks (WPANs) Submission Title: TG4J MAC Features Date Submitted: May 15, 2012 Source: Raymond Krasinski, Philips 345 Scarborough Road, Briarcliff Manor, NY 10510 Voice: , FAX: , Abstract: A short outline of the 4 MAC features present in Letter Ballot 81 and how they relate to the FCC MBAN Rules Purpose: To outline the AMC features present in Letter Ballot 81. Notice: This document has been prepared to assist the IEEE P It is offered as a basis for discussion and is not binding on the contributing individual(s) or organization(s). The material in this document is subject to change in form and content after further study. The contributor(s) reserve(s) the right to add, amend or withdraw material contained herein. Release: The contributor acknowledges and accepts that this contribution becomes the property of IEEE and may be made publicly available by P Raymond Krasinski, Philips
2
TG4J MAC Feature Letter Ballot 81
May 2012 IEEE j Task Group TG4J MAC Feature Letter Ballot 81 Raymond Krasinski, Philips
3
Scope of MBAN Task Group
May 2012 Scope of MBAN Task Group Draft an amendment of for MBAN This amendment defines a physical layer for IEEE in the 2360 to 2400 MHz band which complies with Federal Communications Commission (FCC) MBAN rules. This amendment defines modifications to the MAC needed to support this new physical layer. Raymond Krasinski, Philips
4
May 2012 FCC MBAN Rules Allow use of the MHz band, on a secondary basis, for monitoring, diagnosing or treating patients (no voice). MHz MBANS operations limited to health care facilities (indoor use only), which are outside exclusion zones Must yield use of the band for the primary user In exclusion zones, usage of the band may be for a limited time, on a scheduled basis and allowing for use of only part of the band Healthcare facilities must file a Transition Plan that manages re-channelization out of the spectrum, to accommodate the primary user Raymond Krasinski, Philips
5
May 2012 FCC MBAN Rules (cont) An entity, called a FCC Band Coordinator, manages use of the band for Healthcare applications Manages relationship with primary user and hospitals to determine what spectrum is available and at what time Keeps track of which hospitals are within exclusion zones Raymond Krasinski, Philips
6
Letter Ballot 81 MAC Features
May 2012 Letter Ballot 81 MAC Features Channel Switch Notification Coordinator Switch Multi-Periodic GTS Association Proxy Raymond Krasinski, Philips
7
Channel Switch Notification
May 2012 Channel Switch Notification Description A method for switching channels at a scheduled time Use Case The primary band user is scheduled to use the part of the band a healthcare device is using at a specific time. Usage is normally scheduled by the primary user many days in advance. The device must vacate the portion of the band the primary user requires at the scheduled time. Raymond Krasinski, Philips
8
Coordinator Switch Description
May 2012 Coordinator Switch Description A method for switching a set of devices from one coordinator to another in a controlled manner. Use Case A patient with an array of sensors is located in a ward with a fixed coordinator near the patient. The patient must be moved to a different, known, room for a test. This room also has a fixed coordinator. The patient will be moved past multiple other coordinators that are inappropriate for association. Raymond Krasinski, Philips
9
Multi-Periodic GTS Description
May 2012 Multi-Periodic GTS Description A enhancement to GTS which allows for a defined period between time slots (greater than the beacon interval), resulting in more efficient resource allocation for long interval, periodic traffic applications. Use Case As wireless monitoring proliferates, sensors will remain on patients for longer periods of time, requiring more efficient use of power (maximizing battery life) for low duty cycle applications with periodic data. Raymond Krasinski, Philips
10
Association Proxy Description
May 2012 Association Proxy Description A method for associating a fixed set of sensors to a coordinator when not in close proximity or where there are multiple coordinators in proximity. Use Case A patient upon admission to the hospital is assigned a room. Before being moved to the room the patient is first taken to a central station where sensors are applied and are associated to the coordinator in the patient room using a proxy device. Raymond Krasinski, Philips