Download presentation
Presentation is loading. Please wait.
Published byAsher Arnold Modified over 6 years ago
1
Patient-ventilator interaction: insight from waveforms
Dr Tang Kam Shing AC, ICU, TMH 29 April 2009
2
Waveforms on ventilators
Most modern ventilators have a graphical display on the panel of various mechanical parameters Mostly ignored in daily practices in ICU They actually give a lot of information regarding: The mechanical properties of the respiratory system The interaction between the patient and the ventilator
3
Waveforms on ventilators
However, there are several important points you have to bear in mind before you try to interpret waveforms Whether the patient is paralyzed or actively breathing The waveform analysis for both groups completely different The mode of the ventilator is in PCV or VCV (constant flow or decelerating flow)
4
Passive patients Either paralysed or passive patients with no significant triggering or ventilatory efforts General principle: the controlled variable waveform usually showed not much information (because it is solely controlled by ventilator in passive patients) The dependent variables waveforms showed much more information For VCV, look at the pressure time and flow time waveform For PCV, look at the flow time and volume time waveforms
5
Passive or paralyzed patient
Volume control mode Constant flow
6
Constant flow (VC) Idealized waveforms from constant flow volume control on left side Real life tracing from ventilator
7
Pressure-time waveform
At the beginning of flow, an almost vertical pressure increase occurs because of the frictional forces generated by gas flow which is necessary to overcome the resistance of the airways and the ET tube The curve shape then changes to a linear increase and follows a given slope to its maximum value normally linear Depends on the respiratory system compliance alone
8
Pressure-time in VCV Upper tracing measures at ventilator
Lower tracing meaures at ET tube tip 1st vertical part increased due to resistance of ET tube 2nd part same due to identical compliance of respiratory system resistive compliance
9
Resistance problem Compliance problem
10
Signs of over-stretching or decreasing compliance
11
End-inspiratory pause hold
P2 is the static pressure of the respiratory system Which in the absence of flow equals the alveolar pressure reflects the elastic retraction of the entire respiratory system pressure drop from PIP to P1 represents the pressure required to move the inspiratory flow along the airways Representing the pressure dissipated by the flow-dependent resistances
12
End-inspiratory pause hold
slow post-occlusion decay from P1 to P2 depends on the viscoelastic properties of the system and on the pendulum-like movement of the air (pendelluft) elastic rearrangement of lung volume Allows the different pressures in alveoli at different time constants to equalize Due to the inhomogeneity of the lung parenchyma rapid zones (low time constant) emptied into slow zones (high time constant)
13
End-inspiratory pause hold
Cstat = VT/P2 Cdyn = VT/PIP static compliance of the respiratory system mirrors the elastic features of the respiratory system dynamic compliance also includes the resistive (flow-dependent) component of the airways and the ET tube
14
Passive or paralyzed patient
Pressure control mode
15
Pressure control ventilation
Salient points Decelerating flow Decelerating lung expansion Constant pressure Whether end-inspiratory pause occur depend on inspiatory time and mechanical properties of respiratory system
16
PIP and plateau pressures in PCV
Flow stopped Plateau pressure
17
Resistance and compliance in PCV
Increasing resistance Reducing compliance
18
Rate of pressurization in PCV
Increasing rate of pressurization
19
Inspiratory artifacts common in daily practices
Spike due to pressurization of circuit especially for low pressure rise time
20
Expiratory artifacts common in obstructive lungs
COPD Spike from recoil of ventilator circiut Normal
21
Visualization of dynamic hyperinflation: auto PEEP
22
Measurement of auto-PEEP: Not so simple
24
Spontaneously breathing patients
A bit more complicated
25
Patient-ventilator interaction
In passive patients Basically only the ventilator and passive mechanical properties of respiratory system involved Any abnormal pattern is due to the ventilator or the respiratory system In actively breathing patients 4 different systems interacting: ventilator, respiratory system, respiratory pump and central ventilatory drive Patterns can be due to either one alone or in combinations
26
Now consider this… There are two pumps with two control acting on the same respiratory system Ventilator that is controlled by computer programs that is set by doctors (that commonly do not know what to do) Neuro respiratory centre controlling the respiratory muscles with its own “desirable setting” (that commonly do not agree with doctors)
27
Patient-ventilator asynchrony
Simply means the two pumps are not agreeing with each other A lot of times working against each other Ineffective ventilation Increase work of breathing prolonged weaning I am simple minded, so in essence there are two ways to due with this problem Stop the neuro drive if you believe the neuro drive is not appropriate by sedation +/- paralysis, may be needed in initial management of status asthmaticus Change your ventilator setting to agree as much as possible with neuro ventilatory drive (best to do but not easy) Change neuro drive to agree with your setting (impossible)
28
Why border?
29
Types of asynchrony Essentially 3 phases of mechanical ventilation can show asynchrony Trigger asynchrony Flow asynchrony Termination (cycling) asynchrony
30
Trigger asynchrony The ventilator failed to detect a patient trigger
Inspiratory effort completely wasted The start of inspiration is slower or later than neural inspiration Resulting in wasted efforts in the initial part of spontaneous breaths Double triggering Auto triggering
31
Failed triggering in COPD
32
PEEP = 0 PEEP = 10 Failed trigger and PEEP setting
33
Double triggering from too short inspiratory phase
34
Double triggering
35
Double triggering from far too low flow and volume setting
36
Auto-triggering from leak
37
Other causes of auto-triggering
Too sensitive trigger setting more prevalent in patients who had acquired valve disease and had more dynamic circulatory characteristics Larger heart size larger cardiac output higher ventricular filling pressure lower respiratory-system resistance). result in larger cardiogenic pressure oscillations, which when transmitted to the airway, causing auto-triggering.
38
Pressure or flow trigger??
2 common types of trigger mechanism available on commercial ventilators: pressure trigger and flow trigger initial clinical studies indicated that flow-triggering offered some advantage in reducing trigger asynchrony recent advances in the development of pressure transducers have resulted in nearly equivalent or comparable results
39
Flow asynchrony Occurs whenever the ventilator flow does not match the patient flow requirement a common problem, and the flow setting may be the most frequently incorrectly-set ventilator parameter divided into 2 sections, based on the 2 general methods for delivering gas: VCV with a fixed flow pattern PCV with variable flow
40
Flow asynchrony in VCV Remember the waveform of passive VCV breath with constant flow The shaded area represent the work done by the respiratory pump of the patient
41
Remember “dished out appearance of VCV waveform
43
Flow asynchrony in PCV or PSV
In essence should not be inadequate flow as pressure is fixed with flow as required by patient Still possible with too slow pressure rise time In initial phase of respiration Also result in a “dished out appearance in pressure waveform
44
Flow starvation from slow rise time
45
Too fast rise time flow asynchrony
Pressure-time Flow-time
46
Cycling asynchrony Basically two possibilities
Neuro expiration earlier than mechanical expiration, too late termination of support Neuro expiration later than mechanical expiration, too early termination of support
47
Too late termination of support
Problem is we usually do not have EEG of expiratory muscles Can we detect this problem with waveforms?
48
Too long inspiratory time
49
Premature termination of support: too early cycling
50
Too early cycling (cycling asynchrony)
Notch in expiratory flow
51
Conclusion Waveforms a good supplement to clinical examination as asynchrony can be subtle Always ask yourself 2 questions: Active or passive Mode of ventilation Familiarize with the appearance and usually it is spot diagnosis
52
Any questions? Otherwise go home and relax!
The End Any questions? Otherwise go home and relax!
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.