1 All Medical-Surgical and Critical Care
Clinical Communication System Global Programming All Medical-Surgical and Critical Care Inpatient Units

2 Click on a device to learn more.
Responder 5 Hardware Click on a device to learn more.

3 Nurse Console Please see the Nurse Console Overview Handout for further information.

4 Enhanced Pillow Speaker
Nurse Button Water Button Pain Button Toilet Button

5 Immediately call the Tech
Patient Call-Normal Immediately call the Tech If not answered after 1 minute, calls the Primary Nurse If not answered after 2 minutes, calls the Tech again If not answered after 3 minutes, calls the Console This call can be canceled by the following action: Pressing the cancel button Entering room with a Versus badge Upon Answer Room 404 Patient Room 404 Patient Room 404 Patient

6 Immediately call the Tech
Toilet Button Immediately call the Tech If not answered after 45 seconds, calls the Primary Nurse If not answered after 60 seconds, calls the Console If not answered after 90 seconds, loops back and calls the Tech -Upon answer, allows 2 minutes to cancel call in the room or system will repeat the above escalations This call can be canceled the following action: Pressing the cancel button Entering room with a Versus badge Room 404 Go To Toilet Room 404 Go To Toilet Room 404 Go To Toilet

7 Immediately calls the Tech
Water Button Immediately calls the Tech If not answered after 1 minute, calls the Primary Nurse If not answered after 2 minutes, calls the Tech again If not answered after 3 minutes, calls the Console This call can be canceled by the following action: Pressing the cancel button Entering room with a Versus badge Upon Answer Room 404 Water Room 404 Water Room 404 Water

8 Immediately calls the Primary Nurse
Pain Button Immediately calls the Primary Nurse If not answered after 45 seconds, calls the Charge Nurse If not answered after 60 seconds, calls the Console If not answered after 90 seconds, loops back and calls the Primary Nurse again -Upon answer, allows 10 minutes to cancel call in the room or system will repeat the above escalations This call can be canceled the following action: Pressing the cancel button Entering room with a Versus badge (Nurse Level Only) Room 404 In Pain Room 404 In Pain Room 404 In Pain

9 Enhanced Patient Station
Alternative Call Cord Staff Assist Code Trouble Codes

10 Staff Assist Notify all staff on duty for the unit, RRT group, and Console immediately. This call can be canceled by any of the following actions: Pressing the cancel button Room 404 Staff Assist Room 404 Staff Assist Room 404 Staff Assist Room 404 Staff Assist

11 Also, calls the Operator mini-console to page overhead.
Code Immediately notifies all staff on duty for the unit, the Code Group, and the Console. Also, calls the Operator mini-console to page overhead. Room 404 Code This call can be canceled by the following action: Pressing the cancel button Room 404 Code Room 404 Code Room 404 Code Room 404 Code

12 Patient Station ¼” Jack - AUX Call
Immediately calls the Tech If not answered after 1 minute, calls the Primary Nurse If not answered after 2 minutes, calls the Tech again If not answered after 3 minutes, calls the Console This call can be canceled by the following action: Pressing the cancel button Entering room with a Versus badge Upon Answer Room 404 AUX Room 404 AUX Room 404 AUX

13 Enhanced Patient Station
Cord out Plug Out Cord Out Plug Out

14 Immediately notifies the Tech.
Cord Out Immediately notifies the Tech. If not answered within 2 minutes, calls the Primary Nurse If not canceled within 3 minutes, calls the Console If not canceled within 4 minutes, loops back and calls the Tech again This call can be canceled by any of the following actions: Pressing the cancel button Room 404 Cord Out Room 404 Cord Out Room 404 Cord Out

15 Immediately notifies the Tech
Plug Out Immediately notifies the Tech If not answered within 2 minutes, calls the Primary Nurse If not answered within 3 minutes, calls the Console If not answered within 4 minutes, loops back and calls the Tech This call can be canceled by any of the following actions: Pressing the cancel button Room 404 Plug Out Room 404 Plug Out Room 404 Plug Out

16 Pull Cord Station with Audio
Bath Assist Bath Emergency

17 Immediately calls the Tech
Bath Assist Immediately calls the Tech If not answered after 45 seconds, calls the Primary Nurse If not answered after 1 minute, calls the Console If not answered after 90 seconds, loops back and calls the Tech again This call can be canceled by any of the following actions: Pressing the cancel button on the bathroom station Room 404 Bath Assist Room 404 Bath Assist Room 404 Bath Assist

18 Immediately calls the Tech and the Console
Bath Emergency Immediately calls the Tech and the Console If not answered after 45 seconds, calls the Primary Nurse If not answered after 90 seconds, loops back and calls the Tech again This call can be canceled by any of the following actions: Pressing the cancel button on the bathroom station Room 404 Bath Emergency Room 404 Bath Emergency Room 404 Bath Emergency

19 Immediately calls the Tech and the Console
Shower Emergency Immediately calls the Tech and the Console If not answered within 45 seconds, calls the Primary Nurse. If not answered within 90 seconds, loops back and calls the Tech again This call can be canceled by any of the following actions: Pressing the cancel button on the bathroom station Room 404 Shower Emergency Room 404 Shower Emergency Room 404 Shower Emergency

20 Feature Bed Connector Bed Out Bed Exit

21 If not canceled within 2 minutes, call the Primary Nurse
Bed Out Notify the Tech If not canceled within 2 minutes, call the Primary Nurse If not canceled within 3 minutes, call the console If not canceled within 4 minutes, loop back to the Tech This call can be canceled by any of the following actions: Pressing the cancel button Room 404 Bed Out Room 404 Bed Out Room 404 Bed Out

22 Immediately call the Tech phone and console
Bed Exit Immediately call the Tech phone and console If not canceled after 15 seconds, call the Primary Nurse If not canceled after 15 seconds, loop back to the Tech This call can be canceled by the following action: Pressing the cancel button Room 404 Bed Exit Room 404 Bed Exit Room 404 Bed Exit

23 Jack #1 Jack #2

24 Jack #1 – Ventilator Alarm
Calls all staff on-duty immediately This call can be canceled by the following action: Cancelling the alarm on the ventilator then pressing the cancel button Room 404 Ventilator Alarm Room 404 Ventilator

25 Rolling notification to the Console, Tech, and Primary Nurse
Jack #2 – Bed Alarm Rolling notification to the Console, Tech, and Primary Nurse This call can be canceled by the following action: Pressing the cancel button on the 2-Jack station Room 404 Bed Alarm Room 404 Bed Alarm Room 404 Bed Alarm

26 Staff Terminal-Layer 1 Codes (+) Patient Arrived Patient Out of Room
Call Tech Call RN Call Charge Call Desk Safety (+) Rounding 1 Hr Rounding 2 Hr Q 2 Turn

27 Staff Terminal-Layer 2 Code STEMI STROKE MET Toilet Tissue Paper
Towels Soap Trash Code Gray Active Shooter Call Night EVS Back

28 Staff Terminal-Layer 3 RT (+) RAD (+) EKG (+) UHS (+) Walk Pro- cedure
OFF Monitor Pro- cedure IV Placed Call EVS Call Dietary Back

29 Staff Terminal-Layer 4 Palliative Caution Falls Risk Back TURN PATIENT
NURSE BEDSIDE REPORT PCA BEDSIDE REPORT Back

30 Staff Terminal-Layer 5 PRN TX Resp Distress Trans-port Bi-Pap Needed
STAT ABG Vent Needed Vent Changes Intub-ation Call RT Charge Call RT Group Help Call RT Group Cover Back

31 Staff Terminal-Layer 6 STAT CXR ETT STAT CXR Line STAT KUB Tube STAT
Resp DST STAT KUB Tube STAT R/O PNTX READY FOR THERAP Back

32 Staff Terminal-Layer 7 STAT EKG CALL CN CP 911 CALL DESK CALL RE-
SOURCE CALL ANS CALL TELE CALL SAFETY Back

33 Staff Terminal-Layer 8 CRRT Air-2 Matt-ress Low Boy Bed Hepa Filter
SCD Pump Kanga Feeding Pump Alaris Module Fall Mats Back

34 Rolling 15 second notification to the Tech and Primary Nurse
Patient In Room Rolling 15 second notification to the Tech and Primary Nurse This call can be canceled by the following action: Self cancels after 45 seconds Room 404 Patient In Room Room 404 Patient In Room

35 Displays light pattern and on PC console.
Patient Out of Room Displays light pattern and on PC console. This call can be canceled by the following action: Toggle the Patient Out of Room button off Room 404 Patient Out of Room

36 Call Tech Calls the Tech only
This call can be canceled by the following action: Upon Answer Toggle the Call Tech button off Self cancels after 3 minutes Room 404 Call Tech

37 Calls the Primary Nurse only
Call RN Calls the Primary Nurse only This call can be canceled by the following action: Toggle off the Call RN button Upon Answer Self cancels after 3 minutes Room 404 Call RN

38 Calls the Charge Nurse only
Call Charge Calls the Charge Nurse only This call can be canceled by the following action: Toggle off the Call Charge Button Upon Answer Self cancels after 3 minutes Room 404 Call Charge

39 Call Desk Calls the Console only
This call can be canceled by the following action: Upon Answer Room 404 Call Desk

40 Pressing the button starts a timer.
Rounding 1 Hr Pressing the button starts a timer. At 50 minutes, it turns on a solid pink link outside the room. If not canceled after 70 minutes, the light changes to flashing and notifies the Tech and the Primary Nurse. The notification repeats every 5 minutes until canceled. This call can be canceled by the following action: Toggle off Rounding 1 Hr Button Room 404 Rounding 1 Hr Room 404 Rounding 1 Hr

41 Pressing the button starts a timer.
Rounding 2 Hr Pressing the button starts a timer. At 110 minutes, it turns on a solid pink link outside the room. If not canceled after 130 minutes, the light changes to flashing and notifies the Tech and the Primary Nurse. The notification repeats every 5 minutes until canceled. This call can be canceled by the following action: Toggle off Rounding 2 Hr Button Room 404 Rounding 2 Hr Room 404 Rounding 2 Hr

42 Pressing the button starts a timer.
Q 2 Turn Pressing the button starts a timer. At 1 hour and 45 minutes, two solid green lights turn on outside the room. If not canceled after 2 hours and 15 minutes, the lights change to flashing. This call can be canceled by the following action: Toggle off Q 2 Turn Button

43 Also, calls the Operator mini-console to page overhead.
Code Notifies all staff on duty for the unit, the Code Group, and the console immediately. Also, calls the Operator mini-console to page overhead. Room 404 Code This call can be canceled by the following action: Pressing the cancel button Toggle off the Code button Room 404 Code Room 404 Code Room 404 Code Room 404 Code

44 STEMI Notifies the Code STEMI Group and calls the Operator mini-console to page overhead. This call can be canceled by the following action: Toggle off the STEMI button Self cancels after 60 minutes Room 404 STEMI Room 404 STEMI

45 Stroke Notifies the Code Stroke Group and calls the Operator mini-console to page overhead. This call can be canceled by the following action: Toggle off the Stroke button Self cancels after 60 minutes Room 404 Stroke Room 404 Stroke

46 MET Notifies all staff on duty for the unit, the MET Group, and calls the Operator mini-console to page overhead. This call can be canceled by the following actions: Toggle off MET button Pressing the cancel button Room 404 MET Room 404 MET Room 404 MET

47 Notifies the Code Gray Group.
Calls the Console and Operator mini-console. This call can be canceled by the following actions: Toggle off Code Gray button Self cancels after 30 minutes Room 404 Code Gray Room 404 Code Gray Room 404 Code Gray

48 Notifies the Code Silver Group.
Active Shooter Notifies the Code Silver Group. Calls the Console and Operator mini-console. This call can be canceled by the following actions: Toggle off Active Shooter button Self cancels after 30 minutes Room 404 Active Shooter Room 404 Active Shooter Room 404 Active Shooter

49 At 2 hour and 45 minutes, the lights change to flashing.
Walk Pressing the button starts a timer and displays two solid green lights. At 2 hour and 45 minutes, the lights change to flashing. If not canceled after 3 hours and 45 minutes, the Tech is notified and the notification is resent every 30 minutes until canceled. This call can be canceled by the following action: Toggle off the Walk button Room 404 Walk

50 OFF Monitor Notifies the Monitor Tech Group and displays on the Monitor Tech’s PC console. This call can be canceled by the following action: Toggle off the OFF Monitor button Room 404 Off Monitor Room 404 Off Monitor

51 Displays the light pattern
Procedure Displays the light pattern This call can be canceled by the following action: Toggle off the Procedure button Self cancels after 2 hours

52 Notifies the Primary Nurse and the Console
IV Placed Notifies the Primary Nurse and the Console This call can be canceled by the following action: Self cancels after 45 seconds Room 404 IV Placed Room 404 IV Placed

53 Call EVS Call the Assigned EVS
This call can be canceled by the following action: Self cancels after 2 minutes Room 404 Call EVS

54 Call the Dietary phone only
Call Dietary Call the Dietary phone only This call can be canceled by the following action: Upon answer Self cancels after 3 minutes Room 404 Call Dietary

55 Displays the light pattern only
Falls Risk Displays the light pattern only This call can be canceled by the following action: Toggle off the Falls Risk button

56 Displays the light pattern only
Caution Displays the light pattern only This call can be canceled by the following action: Toggle off Caution button

57 Displays the light pattern only
Palliative Displays the light pattern only This call can be canceled by the following action: Toggle off the Palliative button

58 Notifies the Assigned RT phone
PRN TX Notifies the Assigned RT phone This call can be canceled by the following action: Self Cancel after 45 seconds Room 404 PRN TX

59 Notifies the Assigned RT phone
Resp Distress Notifies the Assigned RT phone This call can be canceled by the following action: Self Cancel after 45 seconds Room 404 Resp Distress

60 Notifies the Assigned RT phone
Transport Notifies the Assigned RT phone This call can be canceled by the following action: Self Cancel after 45 seconds Room 404 Transport

61 Notifies the Assigned RT phone
Bi-Pap Needed Notifies the Assigned RT phone This call can be canceled by the following action: Self Cancel after 45 seconds Room 404 Bi-Pap Needed

62 Notifies the Assigned RT phone
STAT ABG Notifies the Assigned RT phone This call can be canceled by the following action: Self Cancel after 45 seconds Room 404 STAT ABG

63 Notifies the Assigned RT phone
Vent Needed Notifies the Assigned RT phone This call can be canceled by the following action: Self Cancel after 45 seconds Room 404 Vent Needed

64 Notifies the Assigned RT phone
Vent Changes Notifies the Assigned RT phone This call can be canceled by the following action: Self Cancel after 30 minutes Toggle off the Vent Changes button Room 404 Vent Changes

65 Notifies the Assigned RT phone
Intubation Notifies the Assigned RT phone This call can be canceled by the following action: Self Cancel after 30 minutes Toggle off the Intubation button Room 404 Intubation

66 Calls the RT Charge Team
Call RT Charge Calls the RT Charge Team This call can be canceled by the following action: Toggle off Call RT Charge button Pressing the cancel button Upon answer Self cancels after 2 minutes Room 404 Call RT Charge

67 Call RT Group Help Calls the RT Group Team
This call can be canceled by the following action: Toggle off Call RT Charge button Pressing the cancel button Upon answer Self cancels after 2 minutes Room 404 Call RT Group Help Room 404 Call RT Group Help Room 404 Call RT Group Help

68 Call RT Group Cover Calls the RT Group Team.
This call can be canceled by following action: Self cancels after 2 minutes Room 404 Call RT Group Cover Room 404 Call RT Group Help Room 404 Call RT Group Help

69 Notifies the Portable X-ray Tech group phones
STAT CXR ETT Notifies the Portable X-ray Tech group phones This call can be canceled by the following action: Self cancels after 45 seconds Room 404 STAT CXR ETT

70 Notifies the Portable X-ray Tech group phones
STAT CXR Line Notifies the Portable X-ray Tech group phones This call can be canceled by the following action: Self cancels after 45 seconds Room 404 STAT CXR Line

71 Notifies the Portable X-ray Tech group phones
STAT CXR Resp DST Notifies the Portable X-ray Tech group phones This call can be canceled by the following action: Self cancels after 45 seconds Room 404 STAT CXR Resp DST

72 Notifies the Portable X-ray Tech group phones
STAT KUB Tube Notifies the Portable X-ray Tech group phones This call can be canceled by the following action: Self cancels after 45 seconds Room 404 STAT KUB Tube

73 Notifies the Portable X-ray Tech group phones
STAT R/O PNTX Notifies the Portable X-ray Tech group phones This call can be canceled by the following action: Self cancels after 45 seconds Room 404 STAT R/O PNTX

74 Notifies the EKG Tech Group phones
STAT EKG Notifies the EKG Tech Group phones This call can be canceled by the following action: Self cancels after 30 minutes Room 404 STAT EKG

75 Notifies the EKG Tech Group phones
CP 911 Notifies the EKG Tech Group phones This call can be canceled by the following action: Self cancels after 30 minutes Room 404 CP 911

76 Notifies the UHS Group pagers.
CRRT Notifies the UHS Group pagers. The call will stay active on PC Console until the equipment is delivered, then the Unit Secretary will cancel the call on PC Console and send a message to the Primary Nurse saying “Equipment Delivered.” This call can be canceled by the following action: Self cancels after 24 hours Toggle off the CRRT button PC Console Room 404 CRRT Room 404 Equipment Delivered Room 404 CRRT

77 Notifies the UHS Group pagers.
Air-2 Mattress Notifies the UHS Group pagers. The call will stay active on PC Console until the equipment is delivered, then the Unit Secretary will cancel the call on PC Console and send a message to the Primary Nurse saying “Equipment Delivered.” This call can be canceled by the following action: Self cancels after 24 hours Toggle off the CRRT button PC Console Room 404 Air-2 Mattress Room 404 Equipment Delivered Room 404 Air-2 Mattress

78 Notifies the UHS Group pagers.
Low Boy Bed Notifies the UHS Group pagers. The call will stay active on PC Console until the equipment is delivered, then the Unit Secretary will cancel the call on PC Console and send a message to the Primary Nurse saying “Equipment Delivered.” This call can be canceled by the following action: Self cancels after 24 hours Toggle off the CRRT button PC Console Room 404 Low Boy Bed Room 404 Equipment Delivered Room 404 Low Boy Bed

79 Notifies the UHS Group pagers.
Hepa Filter Notifies the UHS Group pagers. The call will stay active on PC Console until the equipment is delivered, then the Unit Secretary will cancel the call on PC Console and send a message to the Primary Nurse saying “Equipment Delivered.” This call can be canceled by the following action: Self cancels after 24 hours Toggle off the CRRT button PC Console Room 404 Hepa Filter Room 404 Equipment Delivered Room 404 Hepa Filter

80 Notifies the UHS Group pagers.
SCD Pump Notifies the UHS Group pagers. The call will stay active on PC Console until the equipment is delivered, then the Unit Secretary will cancel the call on PC Console and send a message to the Primary Nurse saying “Equipment Delivered.” This call can be canceled by the following action: Self cancels after 24 hours Toggle off the CRRT button PC Console Room 404 SCD Pump Room 404 Equipment Delivered Room 404 SCD Pump

81 Notifies the UHS Group pagers.
Kanga Feeding Pump Notifies the UHS Group pagers. The call will stay active on PC Console until the equipment is delivered, then the Unit Secretary will cancel the call on PC Console and send a message to the Primary Nurse saying “Equipment Delivered.” This call can be canceled by the following action: Self cancels after 24 hours Toggle off the CRRT button PC Console Room 404 Kanga Feeding Pump Room 404 Equipment Delivered Room 404 Kanga Feeding Pump

82 Notifies the UHS Group pagers.
Alaris Module Notifies the UHS Group pagers. The call will stay active on PC Console until the equipment is delivered, then the Unit Secretary will cancel the call on PC Console and send a message to the Primary Nurse saying “Equipment Delivered.” This call can be canceled by the following action: Self cancels after 24 hours Toggle off the CRRT button PC Console Room 404 Alaris Module Room 404 Equipment Delivered Room 404 Alaris Module

83 Notifies the UHS Group pagers.
Fall Mats Notifies the UHS Group pagers. The call will stay active on PC Console until the equipment is delivered, then the Unit Secretary will cancel the call on PC Console and send a message to the Primary Nurse saying “Equipment Delivered.” This call can be canceled by the following action: Self cancels after 24 hours Toggle off the CRRT button PC Console Room 404 Fall Mats Room 404 Equipment Delivered Room 404 Fall Mats

84 Versus RTLS Badges How does it work?
IR – Infrared light signals (every 3 seconds) RF – Radio frequency signals (every 2 minutes) Is it safe? Does not interfere with medical devices Tested in healthcare settings & FCC approved How do I wear it? High on the body Button facing outward Uncovered How do I clean it? Isopropyl Alcohol Solution (alcohol wipe) Water-Resistant What is the battery life? Up to 2 years with CR Lithium battery Check with badge tester or pressing button Call ___________ to replace battery

85 Versus RTLS Badges How do I store it when I am not at work?
Somewhere stationary to preserve battery life At rest, IR and RF signals every 2 minutes Not in employee lockers (metal) How will I get my badge? Assigned to you by your Department Director Will be required to sign a contract Where will my location be seen at? In the R5 Application in the PC Console module Staff Registration on Dome Lights Orange = Tech Green = Nurse